Linvatec SRS PopLok GKP-2801Recalls & reported problems

Conmed Corporation · POPLOK SUTURE ANCHOR WITH ONE #2 (5 METRIC) HI-FI SUTURE, 2.8 MM · Class 2 · Implant

All models of Linvatec SRS PopLok (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 20845854014117), received between May 29, 2024 and May 19, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (3 reports).

1 other recall cover devices cleared under the same FDA submission (K091549); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 19, 2026Material Fragmentation

    THE CUSTOMER REPORTED THAT THE DEVICE, GKP-2801, SUTURE ANCHOR WITH ONE #2 (5 METRIC) HI-FI SUTURE,2.8 MM, WAS USED DURING A RC REPAIR PROCEDURE ON (B)(6) 2026 WHEN IT WAS REPORTED, ¿DURING A SURGICAL PROCEDURE ONE POPLOK SUTURE 2.8MM ANCHOR WAS FOUND DAMAGED.¿. THE DRIVER TIP FRAGMENTATION HAD FALLEN IN THE PATIENT; HOWEVER, THE FRAGMENTATION WAS REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED AS PLANNED WITH AN ALTERNATE, UNKNOWN DEVICE. BOTH THE PATIENT AND DOCTOR WERE REPORTED AS ¿GOOD¿. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR HOSPITALIZATION. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25215064 ↗
  • MalfunctionMay 29, 2024Material Fragmentation; Material Fragmentation

    THE DISTRIBUTOR REPORTED ON BEHALF OF THE CUSTOMER THAT THE GKP-2801, 2.8 MM POPLOK SUTURE ANCHOR, DEVICE WAS BEING USED DURING A BANKART REPAIR PROCEDURE ON (B)(6) 2024 WHEN IT WAS REPORTED ¿SURGEON FOLLOWED THE IFU TO USE POPLOK 2.8MM. HE DRILLED PILOT HOLE AND INSERTED ANCHOR. WHILE TENSIONING SUTURE, THE SUTURE SLIPPED OUT FROM THE ANCHOR AND THE IMPLANT BROKE AT DISTAL END.¿. FURTHER ASSESSMENT QUESTIONING FOUND THAT THE DEVICE DID FRAGMENT AND WAS REMOVED WITH THE USE OF AN ARTHROSCOPIC GRASPER. THE PROCEDURE WAS COMPLETED WITH THE USE OF THE SAME ALTERNATE DEVICE. THE CURRENT STATUS OF THE PATIENT WAS REPORTED AS ¿RECOVERING WELL.¿. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19416483 ↗
  • MalfunctionMay 29, 2024Material Fragmentation; Material Fragmentation

    THE DISTRIBUTOR REPORTED ON BEHALF OF THE CUSTOMER THAT THE GKP-2801, 2.8 MM POPLOK SUTURE ANCHOR, DEVICE WAS BEING USED DURING A BANKART REPAIR PROCEDURE ON (B)(6) WHEN IT WAS REPORTED ¿SURGEON FOLLOWED THE IFU TO USE POPLOK 2.8MM. HE DRILLED PILOT HOLE AND INSERTED ANCHOR. WHILE TENSIONING SUTURE, THE SUTURE SLIPPED OUT FROM THE ANCHOR AND THE IMPLANT BROKE AT DISTAL END.¿. FURTHER ASSESSMENT QUESTIONING FOUND THAT THE DEVICE DID FRAGMENT AND WAS REMOVED WITH THE USE OF AN ARTHROSCOPIC GRASPER. THE PROCEDURE WAS COMPLETED WITH THE USE OF THE SAME ALTERNATE DEVICE. THE CURRENT STATUS OF THE PATIENT WAS REPORTED AS ¿RECOVERING WELL.¿. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19416480 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K091549. One submission often covers many models, so check the product and lot details in each record.

  • Z-0253-2018 · Apr 17, 2017 · Terminated
    Nonabsorbable suture anchor system intended to reattach soft tissue to bone in orthopedic surgical procedures may be used in either arthroscopic or open surgical procedures.

Other FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE devices

All FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page