Y-KNOT Y1802Recalls & reported problems

Conmed Corporation · Y-KNOT FLEX 1.8 MM ALL-SUTURE ANCHOR WITH TWO #2 (5 METRIC) HI-FI SUTURE (1 WHITE/BLACK AND 1 WHITE/BLUE) · Class 2 · Implant

All models of Y-KNOT (14)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 20845854021672), received between Feb 5, 2024 and Dec 15, 2025 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation3

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 15, 2025Material Fragmentation; Material Fragmentation

    THE CUSTOMER REPORTED THAT THE DEVICE, Y1802, FLEX 1.8 MM ALL-SUTURE ANCHOR W/2 #2 (5 METRIC) HI-FI SUTURE WAS BEING USED ON (B)(6) 2025 AND "AFTER OPENING THE PACKAGE AND PREPARING TO ENTER THE BONE PASSAGE, WHEN THE HANDLE WAS TAPPED, THE SMALL TEETH AT THE FRONT END BROKE, MAKING THIS CONSUMABLE UNUSABLE. IT IS NOW SENT BACK TO THE AFTER-SALES CENTER IN BEIJING¿. THERE WAS NO IMPACT OR INJURY TO THE PATIENT. PER FURTHER ASSESSMENT IT WAS REPORTED THAT THE "SMALL TEETH" WERE "PART OF INSERTER". THE FRAGMENTS FELL INTO THE SURGICAL SITE AND WERE REMOVED. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23801830 ↗
  • MalfunctionJun 16, 2025Material Fragmentation; Material Fragmentation

    THE DISTRIBUTOR REPORTED ON BEHALF OF THEIR CUSTOMER THAT THE DEVICE, Y1802, FLEX 1.8 MM ALL-SUTURE ANCHOR W/2 #2 (5 METRIC) HI-FI SUTURE, WAS BEING USED DURING AN UNKNOWN PROCEDURE ON (B)(6) 2025 WHEN IT WAS REPORTED, ¿DURING THE GLENOID SURGERY, A SECTION OF THE FORKED END OF THE INSTRUMENT'S IMPLANTATION ROD BROKE OFF INSIDE THE BODY. IT WAS REMOVED USING FORCEPS, AND THE SURGERY WAS NOT DELAYED.¿. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR HOSPITALIZATION FOR THE PATIENT. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22210093 ↗
  • InjuryFeb 5, 2024Material Fragmentation; Material Fragmentation

    CONMED JAPAN REPORTED ON BEHALF OF THE CUSTOMER THAT THE Y1802, Y-KNOT FLEX 1.8MM ALL-SUTURE ANCHOR DEVICE WAS BEING USED DURING A KNEE MENISCUS PRESERVATION SURGERY PROCEDURE ON (B)(6) 2024 WHEN IT WAS REPORTED ¿Y1802 WAS USED FOR KNEE CENTRALIZATION UNDER DOCTORS¿ SURGERY. AT THAT TIME, THE FORK PART AT THE TIP OF THE INSERTER WAS DAMAGED. DAMAGED METAL REMAINS IN THE BODY.¿ FURTHER ASSESSMENT QUESTIONING FOUND THAT THERE WAS NO DELAY REPORTED AND THERE WAS NO MEDICAL/SURGICAL INTERVENTION. CURRENTLY, IT IS UNKNOWN IF THERE ARE ANY PLANS TO REMOVE THE FRAGMENT FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT THE USE OF AN ALTERNATE DEVICE. THE CURRENT STATUS IF THE PATIENT WAS REPORTED AS UNKNOWN. THIS REPORT IS BEING RAISED ON THE REPORTED INJURY DUE TO THE PATIENT RETAINING A FOREIGN BODY.

    Patient problems: Foreign Body In Patient

    MAUDE report 18639316 ↗