Y-Knot YPRC02RRecalls & reported problems

Conmed Corporation · Y-KNOT PRO RC ANCHOR, TWO RIBBONS · Class 2 · Implant

All models of Y-KNOT (14)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20845854648923), received between May 14, 2024 and Jun 19, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Material Fragmentation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation2

Patient problems

  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 19, 2025Material Fragmentation; Material Fragmentation

    CONMED JAPAN REPORTED ON BEHALF OF THEIR CUSTOMER THAT THE DEVICE YPRC02R, Y-KNOT PRO RC ANCHOR, TWO RIBBONS WAS BEING USED ON (B)(6) 2025 DURING AN ACROMIOCLAVICULAR DISLOCATION AND ¿A PILOT HOLE WAS MADE IN THE CORACOID PROCESS WITH A 2.4MM K-WIRE, AND THE ANCHOR WAS INSERTED, BUT PERHAPS BECAUSE THE INSERTION DEPTH WAS INSUFFICIENT, THE ANCHOR DID NOT ENTER THE BONE HALFWAY AND FELL OUT. DURING THIS PROCESS, THE TIP OF THE INSERTER BROKE AND REMAINED INSIDE THE BODY. THE SECOND ONE WAS INSERTED AGAIN FROM A DIFFERENT LOCATION, AND WAS SUCCESSFULLY INSERTED, COMPLETING THE OPERATION. THE TIP OF THE ITEM WAS LEFT INSIDE THE BODY. IT IS BELIEVED THAT THE DRILLED HOLE WAS NOT DEEP ENOUGH AND THE BONE WAS HARD, SO THE TIP WAS BENT AND HIT WITH A HAMMER, CAUSING IT TO BREAK.¿ THE PROCEDURE WAS COMPLETED. THIS REPORT IS BEING RAISED DUE TO THE REPORTED INJURY OF DEVICE FRAGMENTATION LEFT IN THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 22291944 ↗
  • MalfunctionMay 14, 2024Material Fragmentation; Material Fragmentation

    THE CONMED JAPAN REPORTED ON BEHALF OF THE CUSTOMER THAT THE YPRC02R Y-KNOT PRO RC ANCHOR, TWO RIBBONS DEVICE WAS BEING USED ON (B)(6) 2024 ON WHEN IT WAS REPORTED ¿THE FORK AT THE TIP OF THE INSERTER BROKE. THE FRAGMENT RETRIEVED FROM THE PATIENT BODY.¿. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT. THIS REPORT IS BEING RAISED ON THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19309340 ↗

Other Fastener, fixation, nondegradable, soft tissue devices

All Fastener, fixation, nondegradable, soft tissue devices →