Y-Knot, Hi-Fi YP1301BRecalls & reported problems

Conmed Corporation · Y-KNOT PRO FLEX 1.3 MM ANCHOR, NO. 2 HI-FI BLUE · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 20845854649012), received between Apr 3, 2025 and Mar 5, 2026 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Material Fragmentation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation3

Patient problems

  • Foreign Body In Patient2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 5, 2026Material Fragmentation

    CONMED JAPAN REPORTED ON BEHALF OF A CUSTOMER THAT THE YP1301B, Y-KNOT PRO FLEX 1.3 MM ANCHOR, NO. 2 HI-FI BLUE DEVICE WAS USED IN AN ARTHROSCOPIC SHOULDER LABRAL ARTHROPLASTY PROCEDURE ON (B)(6) 2026, AND ¿DURING BANKART REPAIR, AFTER DRILLING INTO THE GLENOID, THE ANCHOR WAS INSERTED, BUT THE TIP OF THE ANCHOR DID NOT FIT FIRMLY INTO THE BONE HOLE, OR THE BONE QUALITY WAS TOO GOOD AND HARD, SO IT DID NOT ADVANCE AT ALL, AND THE TIP OF THE INSERTER BENT AND BROKE. WE MANAGED TO REMOVE THE BROKEN PART FROM THE BODY USING A GRASPER. AFTER THAT, THE BONE HOLE WAS RE-DRILLED, THE ANCHOR WAS RE-INSERTED, AND THE SURGERY WAS COMPLETED SUCCESSFULLY. THE OPERATION TIME DUE TO THIS BREAKAGE WAS EXTENDED BY 5 TO 10 MINUTES.¿. THERE WAS NO REPORT OF INJURY, MEDICAL/SURGICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24521941 ↗
  • InjuryJan 12, 2026Material Fragmentation

    CONMED JAPAN REPORTED ON BEHALF OF A CUSTOMER THAT THE YP1301B, Y-KNOT PRO FLEX 1.3 MM ANCHOR, NO. 2 HI-FI BLUE DEVICE WAS USED IN AN ABR PROCEDURE ON (B)(6) 2025, AND "ONE SIDE OF THE U-SHAPED PORTION OF THE SHAFT AT THE TIP OF THE ANCHOR BROKE. THE BONE WAS NOT PARTICULARLY HARD, BUT IT WAS A LITTLE TOUGH WHEN TRYING TO REMOVE THE SHAFT. WHEN THE SHAFT WAS LATER REMOVED, IT WAS SLIGHTLY BENT AND THE TIP WAS BROKEN". THE PROCEDURE WAS COMPLETED WITH THE USE OF AN UNKNOWN ALTERNATE DEVICE. FURTHER ASSESSMENT FOUND "IN ORDER TO REMOVE THE FRAGMENTS FROM THE PATIENT, THE ENTIRE ANCHOR HAD TO BE REMOVED, AND CONSIDERING THE RISK TO THE PATIENT, IT WAS LEFT IN PLACE". IT WAS FURTHER CLARIFIED THAT "BOTH THE ANCHOR AND A FRAGMENT [OF THE DRIVER] WERE LEFT IN THE PATIENT". THERE WAS NO MEDICAL/SURGICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT. THIS REPORT IS BEING RAISED DUE TO THE REPORTED INJURY OF A RETAINED FRAGMENT OF THE DEVICE DRIVER.

    Patient problems: Foreign Body In Patient

    MAUDE report 24035502 ↗
  • InjuryApr 3, 2025Material Fragmentation; Material Fragmentation

    A DISTRIBUTOR REPORTED ON BEHALF OF A CUSTOMER THAT THE YP1301B, Y-KNOT PRO FLEX 1.3 MM ANCHOR, NO. 2 HI-FI BLUE DEVICE WAS USED IN A SHOULDER ARTHROSCOPY / INSTABILITY / BANKART PROCEDURE ON 26MAR25, AND ¿AN INSERTER TOOTH REMAINED IN THE BONE AFTER IMPACTION OF THE YP1301B ANCHOR.¿. FURTHER ASSESSMENT FOUND IT WAS "IMPOSSIBLE TO RETRIEVE" THE FRAGMENT. THE PROCEDURE WAS COMPLETED WITHOUT THE USE OF AN ALTERNATE DEVICE AND NO DELAY. THERE WAS NO REPORT OF MEDICAL/SURGICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT. THIS REPORT IS BEING RAISED DUE TO THE REPORTED INJURY OF RETAINED FOREIGN BODY.

    Patient problems: Foreign Body In Patient

    MAUDE report 21761142 ↗

Other Fastener, fixation, nondegradable, soft tissue devices

All Fastener, fixation, nondegradable, soft tissue devices →