Bralon 170013Recalls & reported problems
Covidien LP · Nylon Single Stitch Reload · Class 2
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 20884521126982), received between Jun 17, 2025 and Jun 25, 2025 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
2 other recalls cover devices cleared under the same FDA submission (K934738); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Insufficient Information2
- Foreign Body In Patient1
- Unintended Radiation Exposure1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJun 25, 2025Break; Break
ACCORDING TO THE REPORTER DURING LAPAROSCOPIC HERNIORRHAPHY PARAOESOPHAGEAL SURGERY, THE SURGEON INSERTED THE INSTRUMENT INTO THE TROCAR WITH AT LEAST THREE SUTURES WITHOUT ISSUE. USER NOTICED THAT THE NEEDLE WAS BROKEN ONLY ON THE FOURTH SUTURE. THE BROKEN NEEDLE REMAINED INSIDE THE PATIENT. MISSING NEEDLE PROTOCOL WAS FOLLOWED AND CONTACTED RADIOLOGY. THE RADIOLOGIST CONFIRMED THE PRESENCE OF THE NEEDLE INSIDE THE PATIENT. THE SURGEON RE-ENTERED THE ABDOMINAL CAVITY TO ATTEMPT TO RETRIEVE IT, BUT WITHOUT SUCCESS. THE SURGEON CONTINUED USING THE INSTRUMENT WITH OTHER NEEDLES, AND ALL SUBSEQUENT SUTURES CAME OUT INTACT. IT WAS ALSO REPORTED THAT, THE ASSISTANT NURSE AND SALES REP DID NOT NOTICE THAT THE NEEDLES HAD SHIFTED IN THE SMALL WHITE BOX DURING THE LOADING OF THE INSTRUMENT.
Patient problems: Foreign Body In Patient; Unintended Radiation Exposure
MAUDE report 22325292 ↗MalfunctionJun 23, 2025Break; Break
IT WAS REPORTED THAT DURING A LAPAROSCOPIC PARAESOPHAGEAL HERNIORRHAPHY, THE NURSE HAD ENSURED THE SUTURE WAS PROPERLY LOADED ON TO THE INSTRUMENT AND WHEN THE SURGEON INSERTED THE DEVICE INTO THE TROCAR, THE SURGEON IMMEDIATELY NOTICED THAT THE NEEDLE WAS BROKENIN TWO. ON THE SECOND SUTURE, THE NEEDLE WAS NOT PROPERLY PLACED IN THE PACKAGING WHEN OPENED PRIOR TO USE DURING THE PROCEDURE. IT WAS NOTED THAT THERE WAS NO NEEDLE LEFT INSIDE THE PATIENT. THERE WAS NO PATIENT INJURY.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 22305911 ↗MalfunctionJun 20, 2025Break; Break
ACCORDING TO THE REPORTER, THE NEEDLE WAS BROKEN. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22297407 ↗MalfunctionJun 17, 2025Break; Break
IT WAS REPORTED THAT THE NURSE HAD ENSURED THE SUTURE WAS PROPERLY LOADED ON TO THE INSTRUMENT AND WHEN THE SURGEON INSERTED THE DEVICE INTO THE TROCAR, THE SURGEON IMMEDIATELY NOTICED THAT THE NEEDLE WAS BROKEN IN TWO. ON THE SECOND SUTURE, THE NEEDLE WAS NOT PROPERLY PLACED IN THE PACKAGING WHEN OPENED PRIOR TO USE DURING A LAPAROSCOPIC PROCEDURE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 22231394 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K934738. One submission often covers many models, so check the product and lot details in each record.
- Z-2281-2026 · Apr 15, 2026 · Open, ClassifiedCovidien EndoStitch" Polysorb Single Use Loading Unit; Product numbers: 170053 and 170071;
- Z-2543-2017 · May 3, 2017 · TerminatedSutures. ENDO STITCH: ENDO STITCH POLYSORB 0 48" U/D DLU SU, Item Number: 170056; ENDO STITCH POLYSORB 0 48" VIO DLU SU, Item Number: 170052; ENDO STITCH POLYSORB 2/0 48" U/D DLU
Other Endoscopic tissue approximation device devices
- Endo Stitch 173016Covidien LP · Endoscopic tissue approximation device1 death1,333 reports
- ProTack 174006Covidien LP · Endoscopic tissue approximation device709 reports
- N/A ECS29AEthicon Endo-Surgery, LLC · Endoscopic tissue approximation device1 death519 reports
- N/A ECS25AEthicon Endo-Surgery, LLC · Endoscopic tissue approximation device233 reports
- OverStitch™ M00505530Boston Scientific Corporation · Endoscopic tissue approximation device1 recall64 reports
- Polysorb 170053Covidien LP · Endoscopic tissue approximation device1 recall63 reports
- Surgidac 173024Covidien LP · Endoscopic tissue approximation device70 reports
- NA M00505550Boston Scientific Corporation · Endoscopic tissue approximation device1 recall9 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.