Kendall 40000006 (UDI 20884527008435)Recalls & reported problems
Cardinal Health, Inc. · 1710H Multi-Function Defibrillation Electrodes · Class 3
- Adverse event reports
- 3
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 20884527008435), received between Aug 9, 2019 and Feb 5, 2020 — classified as 1 death, 2 malfunctions.
The most frequently reported specific device problem is “Difficult or Delayed Separation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Difficult or Delayed Separation2
- Product Quality Problem2
- Off-Label Use1
Patient problems
- No Consequences Or Impact To Patient2
- Death1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 5, 2020Product Quality Problem; Difficult or Delayed Separation; Product Quality Problem; Difficult or Delayed Separation
THE CUSTOMER REPORTED THAT THERE IS NOT ENOUGH ADHESIVE PROPERTIES ON THE PAD LEAVING THE ADHESIVE RING TO BE INSUFFICIENT. THEY ALSO STATED THAT THE CONNECTION IS EXTREMELY SECURE SO IT IS DIFFICULT TO DETACH THE PATIENT PADS FROM THE DEFIBRILLATOR CABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9671216 ↗MalfunctionFeb 5, 2020Product Quality Problem; Difficult or Delayed Separation; Product Quality Problem; Difficult or Delayed Separation
THE CUSTOMER REPORTED THAT THERE IS NOT ENOUGH ADHESIVE PROPERTIES ON THE PAD LEAVING THE ADHESIVE RING TO BE INSUFFICIENT. THEY ALSO STATED THAT THE CONNECTION IS EXTREMELY SECURE SO IT IS DIFFICULT TO DETACH THE PATIENT PADS FROM THE DEFIBRILLATOR CABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9671100 ↗DeathAug 9, 2019Off-Label Use; Off-Label Use
THE CUSTOMER REPORTED, A HEALTH CARE PROVIDER ATTEMPTED TO USE A DEFIBRILLATOR PAD WITH THE WRONG AUTOMATIC EXTERNAL DEFIBRILLATOR (AED) MACHINE AND THE PATIENT EXPIRED AS A RESULT. ADDITIONALLY, THE CUSTOMER STATED THE ISSUE WAS USER ERROR AND NOT THE PRODUCT. THE NURSE CONNECTED THE INCORRECT DEFIBRILLATOR PAD TO THE DEFIBRILLATOR. THE PAD WAS CLEARLY MARKED ON THE PACKAGE AND THE NURSE DID NOT SEE IT.
Patient problems: Death
MAUDE report 8878277 ↗
Other Automated External Defibrillators (Non-Wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.