Dexcom G4 Platinum (Pedicatric) Replacement Receiver Kit STR-KD-PNKRecalls & reported problems
Dexcom, Inc. · G4 Platinum (Pediatric) Replacement Receiver Kit Pink · Class 3
All models of Dexcom G4 Platinum (41)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 30386270000102), received between Mar 22, 2017 and Oct 3, 2017 — classified as 6 malfunctions.
The most frequently reported specific device problem is “No Audible Alarm” (3 reports).
2 other recalls cover devices cleared under the same FDA submission (P120005); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- No Audible Alarm3
- Device Displays Incorrect Message2
- Defective Alarm1
Patient problems
- No Consequences Or Impact To Patient6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 3, 2017Device Displays Incorrect Message; Device Displays Incorrect Message
DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR HWBBT CALL TECH SUPPORT . NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS RETURNED FOR EVALUATION. THE COMPLAINT COULD NOT BE CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6911030 ↗MalfunctionSep 1, 2017No Audible Alarm; No Audible Alarm
THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS EXTERNALLY VISUALLY INSPECTED AND NO DEFECT WAS FOUND. RECEIVER WAS CHARGED AND BOOTED UP. FUNCTIONAL TEST WAS PERFORMED AND FAILED. MANUAL "TRY IT" TEST WAS PERFORMED AND FAILED. PERFORMED MANUAL TEST BY SETTING SOUND LEVEL TO 55 AND NO AUDIO WAS HEARD. GLOBAL RECEIVER COMMUNICATION TOOL TEST WAS PERFORMED AND FAILED. THE REPORTED EVENT OF NO AUDIO OUTPUT WAS CONFIRMED. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE SPEAKER ASSEMBLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6839159 ↗MalfunctionJul 5, 2017Device Displays Incorrect Message; Device Displays Incorrect Message
DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED AN ERRHWRF. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT WAS RETURNED FOR EVALUATION. THE COMPLAINT COULD NOT BE CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6687819 ↗MalfunctionMay 18, 2017No Audible Alarm; No Audible Alarm
DEXCOM WAS MADE AWARE ON (B)(6) 2017 THAT ON (B)(6) 2016 THE RECEIVER HAD NO AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WERE PROVIDED FOR EVALUATION. THE REPORTED NO AUDIO OUTPUT COULD NOT BE CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6575959 ↗MalfunctionApr 20, 2017No Audible Alarm; No Audible Alarm
DEXCOM WAS MADE AWARE ON (B)(6) 2017 THAT ON (B)(6) 2016 THE RECEIVER HAD NO AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. THE RECEIVER WAS RETURNED FOR EVALUATION. AN EXTERIOR VISUAL INSPECTION WAS PERFORMED AND PASSED. A MANUAL TEST WAS PERFORMED AND PASSED. THE REPORTED EVENT OF NO AUDIO OUTPUT WAS NOT CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6508950 ↗MalfunctionMar 22, 2017Defective Alarm; Defective Alarm
DEXCOM WAS MADE AWARE ON (B)(6) 2017 THAT ON (B)(6) 2017 THE RECEIVER HAD INTERMITTENT AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE REPORTED EVENT COULD NOT BE CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6425657 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120005. One submission often covers many models, so check the product and lot details in each record.
- Z-1561-2020 · Feb 3, 2020 · TerminatedDexcom G5 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracki
- Z-1560-2020 · Feb 3, 2020 · TerminatedDexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends a
Other Sensor, Glucose, Invasive devices
- Dexcom G4 Platinum Sensor 9500-27Dexcom, Inc. · Sensor, Glucose, Invasive2 recalls2 deaths99,426 reports
- Dexcom G5 Transmitter Kit STT-GF-001Dexcom, Inc. · Sensor, Glucose, Invasive1 death58,932 reports
- Dexcom G5 Mobile IOS App SW10611Dexcom, Inc. · Sensor, Glucose, Invasive54,544 reports
- Dexcom G5 Transmitter 9438-06Dexcom, Inc. · Sensor, Glucose, Invasive15,360 reports
- Dexcom G5 Mobile Receiver MT22719Dexcom, Inc. · Sensor, Glucose, Invasive1 death12,594 reports
- Dexcom G4 SBT Transmitter 9438-05Dexcom, Inc. · Sensor, Glucose, Invasive1 death9,041 reports
- Dexcom G5 Mobile Receiver MT22719-BLUDexcom, Inc. · Sensor, Glucose, Invasive7,685 reports
- Dexcom G5 Mobile Receiver Kit STK-GF-001Dexcom, Inc. · Sensor, Glucose, Invasive1 recall6,345 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page