Dexcom G4 Platinum (Pedicatric) Replacement Receiver Kit STR-KD-PNKRecalls & reported problems

Dexcom, Inc. · G4 Platinum (Pediatric) Replacement Receiver Kit Pink · Class 3

All models of Dexcom G4 Platinum (41)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 30386270000102), received between Mar 22, 2017 and Oct 3, 2017 — classified as 6 malfunctions.

The most frequently reported specific device problem is “No Audible Alarm” (3 reports).

2 other recalls cover devices cleared under the same FDA submission (P120005); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • No Audible Alarm3
  • Device Displays Incorrect Message2
  • Defective Alarm1

Patient problems

  • No Consequences Or Impact To Patient6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 3, 2017Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR HWBBT CALL TECH SUPPORT . NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS RETURNED FOR EVALUATION. THE COMPLAINT COULD NOT BE CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6911030 ↗
  • MalfunctionSep 1, 2017No Audible Alarm; No Audible Alarm

    THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS EXTERNALLY VISUALLY INSPECTED AND NO DEFECT WAS FOUND. RECEIVER WAS CHARGED AND BOOTED UP. FUNCTIONAL TEST WAS PERFORMED AND FAILED. MANUAL "TRY IT" TEST WAS PERFORMED AND FAILED. PERFORMED MANUAL TEST BY SETTING SOUND LEVEL TO 55 AND NO AUDIO WAS HEARD. GLOBAL RECEIVER COMMUNICATION TOOL TEST WAS PERFORMED AND FAILED. THE REPORTED EVENT OF NO AUDIO OUTPUT WAS CONFIRMED. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE SPEAKER ASSEMBLY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6839159 ↗
  • MalfunctionJul 5, 2017Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED AN ERRHWRF. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT WAS RETURNED FOR EVALUATION. THE COMPLAINT COULD NOT BE CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6687819 ↗
  • MalfunctionMay 18, 2017No Audible Alarm; No Audible Alarm

    DEXCOM WAS MADE AWARE ON (B)(6) 2017 THAT ON (B)(6) 2016 THE RECEIVER HAD NO AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WERE PROVIDED FOR EVALUATION. THE REPORTED NO AUDIO OUTPUT COULD NOT BE CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6575959 ↗
  • MalfunctionApr 20, 2017No Audible Alarm; No Audible Alarm

    DEXCOM WAS MADE AWARE ON (B)(6) 2017 THAT ON (B)(6) 2016 THE RECEIVER HAD NO AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. THE RECEIVER WAS RETURNED FOR EVALUATION. AN EXTERIOR VISUAL INSPECTION WAS PERFORMED AND PASSED. A MANUAL TEST WAS PERFORMED AND PASSED. THE REPORTED EVENT OF NO AUDIO OUTPUT WAS NOT CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6508950 ↗
  • MalfunctionMar 22, 2017Defective Alarm; Defective Alarm

    DEXCOM WAS MADE AWARE ON (B)(6) 2017 THAT ON (B)(6) 2017 THE RECEIVER HAD INTERMITTENT AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE REPORTED EVENT COULD NOT BE CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6425657 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P120005. One submission often covers many models, so check the product and lot details in each record.

  • Z-1561-2020 · Feb 3, 2020 · Terminated
    Dexcom G5 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracki
  • Z-1560-2020 · Feb 3, 2020 · Terminated
    Dexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends a

Other Sensor, Glucose, Invasive devices

All Sensor, Glucose, Invasive devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page