Dexcom G4 Platinum (Pedicatric) Replacement Receiver Kit STR-KD-BLURecalls & reported problems
Dexcom, Inc. · G4 Platinum (Pediatric) Replacement Receiver Kit Blue · Class 3
All models of Dexcom G4 Platinum (41)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 30386270000119), received between Feb 21, 2017 and Jan 22, 2018 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Device Displays Incorrect Message” (5 reports).
2 other recalls cover devices cleared under the same FDA submission (P120005); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Displays Incorrect Message5
- Inappropriate or Unexpected Reset1
- No Audible Alarm1
- Use of Device Problem1
- Wireless Communication Problem1
Patient problems
- No Consequences Or Impact To Patient8
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 22, 2018Device Displays Incorrect Message; Device Displays Incorrect Message
DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, AN ERRHWBAT WAS DISPLAYED ON THE RECEIVER. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS NOT AVAILABLE FOR EVALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7210543 ↗MalfunctionJan 12, 2018Device Displays Incorrect Message; Device Displays Incorrect Message
DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR HW121. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS RETURNED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7186832 ↗MalfunctionOct 16, 2017Device Displays Incorrect Message; Device Displays Incorrect Message
DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR HWBAT. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS RETURNED FOR EVALUATION. THE COMPLAINT COULD NOT BE CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6952895 ↗MalfunctionSep 19, 2017Device Displays Incorrect Message; Device Displays Incorrect Message
DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR HWRF. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS RETURNED FOR EVALUATION. THE COMPLAINT COULD NOT BE CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6875966 ↗MalfunctionApr 7, 2017Use of Device Problem; Device Displays Incorrect Message; Use of Device Problem; Device Displays Incorrect Message
DEXCOM WAS MADE AWARE ON (B)(6) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR241. IT WAS REPORTED THAT AN ADULT PATIENT WAS USING A PEDIATRIC RECEIVER. IT SHOULD BE NOTED THAT THE DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM USER'S GUIDE STATES: THE DEXCOM G4 PLATINUM (PEDIATRIC) CONTINUOUS GLUCOSE MONITORING SYSTEM IS A GLUCOSE MONITORING DEVICE INDICATED FOR DETECTING TRENDS AND TRACKING PATTERNS IN PERSONS AGES 2 TO 17 YEARS WITH DIABETES. NO PRODUCT OR DATA WERE RETURNED FOR EVALUATION. THE COMPLAINT COULD NOT BE CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6472962 ↗MalfunctionMar 29, 2017No Audible Alarm; No Audible Alarm
DEXCOM WAS MADE AWARE ON (B)(6) 2017 THAT ON (B)(6) 2017 THE RECEIVER HAD NO AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WERE PROVIDED FOR EVALUATION. THE REPORTED NO AUDIO OUTPUT COULD NOT BE CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6442885 ↗MalfunctionMar 2, 2017Wireless Communication Problem; Wireless Communication Problem
DEXCOM WAS MADE AWARE ON (B)(6) 2017 THAT ON (B)(6) 2017, PATIENT EXPERIENCED A LOSS OF CONNECTION BETWEEN THE TRANSMITTER AND RECEIVER. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6373467 ↗MalfunctionFeb 21, 2017Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset
DEXCOM WAS MADE AWARE ON (B)(6) 2017 THAT ON (B)(6) 2017 THE RECEIVER INITIALIZED WITHOUT A MANUAL RESTART. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WERE PROVIDED FOR EVALUATION. THE REPORT OF THE RECEIVER INITIALIZING WITHOUT MANUAL RESTART COULD NOT BE CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6348067 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120005. One submission often covers many models, so check the product and lot details in each record.
- Z-1561-2020 · Feb 3, 2020 · TerminatedDexcom G5 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracki
- Z-1560-2020 · Feb 3, 2020 · TerminatedDexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends a
Other Sensor, Glucose, Invasive devices
- Dexcom G4 Platinum Sensor 9500-27Dexcom, Inc. · Sensor, Glucose, Invasive2 recalls2 deaths99,426 reports
- Dexcom G5 Transmitter Kit STT-GF-001Dexcom, Inc. · Sensor, Glucose, Invasive1 death58,932 reports
- Dexcom G5 Mobile IOS App SW10611Dexcom, Inc. · Sensor, Glucose, Invasive54,544 reports
- Dexcom G5 Transmitter 9438-06Dexcom, Inc. · Sensor, Glucose, Invasive15,360 reports
- Dexcom G5 Mobile Receiver MT22719Dexcom, Inc. · Sensor, Glucose, Invasive1 death12,594 reports
- Dexcom G4 SBT Transmitter 9438-05Dexcom, Inc. · Sensor, Glucose, Invasive1 death9,041 reports
- Dexcom G5 Mobile Receiver MT22719-BLUDexcom, Inc. · Sensor, Glucose, Invasive7,685 reports
- Dexcom G5 Mobile Receiver Kit STK-GF-001Dexcom, Inc. · Sensor, Glucose, Invasive1 recall6,345 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page