Skip to content

Dexcom G4 PLATINUM (Pediatric) Replacement Receiver with Share Kit US STR-PR-BLURecalls & reported problems

Dexcom, Inc. · G4 PLATINUM (Pediatric) Replacement Receiver with Share Kit US Black · Class 3

Quick summary

This Dexcom G4 PLATINUM receiver is a handheld display for glucose sensor readings. Its 155 reports mostly describe the receiver showing wrong error messages, sounding no alarm, losing its wireless connection, or stopping working. The reports include no deaths and 1 injury, and 154 are malfunctions. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
155
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 155 adverse event reports that list this exact device (DI 30386270000201), received between Aug 18, 2016 and May 14, 2019 — classified as 1 injury, 154 malfunctions.

The most frequently reported specific device problem is “Device Displays Incorrect Message” (67 reports).

2 other recalls cover devices cleared under the same FDA submission (P120005); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Device Displays Incorrect Message67
  • No Audible Alarm27
  • Defective Alarm20
  • Wireless Communication Problem16
  • Inappropriate or Unexpected Reset10
  • Device Inoperable7
  • Unable to Obtain Readings5
  • Unexpected Shutdown2

Patient problems

  • No Consequences Or Impact To Patient155

Recent adverse event reports

The 25 most recent of 155 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 14, 2019Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT A "HWBBT" ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8607609 ↗
  • MalfunctionMar 12, 2019No Audible Alarm; No Audible Alarm

    DEXCOM WAS MADE AWARE ON (B)(4) 2019 THAT ON (B)(6) 2019, THE RECEIVER HAD NO AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE RECEIVER HAVING NO AUDIO OUTPUT COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8414243 ↗
  • MalfunctionNov 16, 2018Wireless Communication Problem; Wireless Communication Problem

    DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, A LOSS OF CONNECTION OCCURRED. NO DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ISSUE WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8079642 ↗
  • MalfunctionNov 16, 2018Unexpected Shutdown; Unexpected Shutdown

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER CEASED TO FUNCTION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE CONFIRMATION OF THE COMPLAINT IS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8078648 ↗
  • MalfunctionNov 9, 2018Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "ERR HWBBT". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8057054 ↗
  • MalfunctionNov 6, 2018Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, THE RECEIVER DISPLAYED "ERR HWBBT". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE RECEIVER HAS BEEN RECEIVED FOR DEVICE INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF DEVICE INVESTIGATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8043407 ↗
  • MalfunctionOct 12, 2018Unexpected Shutdown; Unexpected Shutdown

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER CEASED TO FUNCTION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. THE RECEIVER WAS RETURNED FOR EVALUATION. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. A RECEIVER CHARGE AND BOOT WAS PERFORMED AND FAILED DUE TO RECEIVER NOT TURNING ON. THE BATTERY WAS REPLACED WITH A KNOWN GOOD BATTERY TO RETRIEVE THE RECEIVER DOWNLOAD, BUT THE RECEIVER WOULD STILL NOT TURN ON. THE RECEIVER CASE WAS OPENED, AND AN INTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. CONFIRMATION OF THE COMPLAINT WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7961621 ↗
  • MalfunctionSep 27, 2018Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "ERR 281". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7916169 ↗
  • MalfunctionAug 24, 2018Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, AN ERROR ICON WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS NOT AVAILABLE FOR EVALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7814996 ↗
  • MalfunctionJul 16, 2018No Audible Alarm; No Audible Alarm

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER HAD NO AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE RECEIVER HAVING NO AUDIO OUTPUT COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7692625 ↗
  • MalfunctionJul 3, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "ERR HWRF". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7660342 ↗
  • MalfunctionJun 29, 2018Overheating of Device; Use of Device Problem; Overheating of Device; Use of Device Problem

    DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, THE RECEIVER WAS OVERHEATING. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE OVERHEATING OF THE RECEIVER COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED. IT WAS REPORTED THAT THE PATIENT DROPPED THEIR RECEIVER MULTIPLE TIMES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7652942 ↗
  • MalfunctionJun 20, 2018Device Inoperable; Device Inoperable

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER CEASED TO FUNCTION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE CONFIRMATION OF THE COMPLAINT IS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7623250 ↗
  • MalfunctionJun 20, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "CALL TECH SUPPORT AND ERR HWBAT". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7620469 ↗
  • MalfunctionJun 13, 2018No Audible Alarm; No Audible Alarm

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018, THE RECEIVER HAD NO AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE RECEIVER HAVING NO AUDIO OUTPUT COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7598554 ↗
  • MalfunctionJun 11, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "ERR HWBAT". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7587269 ↗
  • MalfunctionJun 8, 2018Defective Alarm; Defective Alarm

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018, THE RECEIVER HAD INTERMITTENT VIBRATION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE RECEIVER HAVING INTERMITTENT VIBRATION COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7583960 ↗
  • InjuryMay 11, 2018Wireless Communication Problem; Wireless Communication Problem

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018, A LOSS OF CONNECTION BETWEEN THE TRANSMITTER AND RECEIVER OCCURRED. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF A LOSS OF CONNECTION COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7507098 ↗
  • MalfunctionMar 23, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "ERR HWBBT". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7367073 ↗
  • MalfunctionMar 2, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, AN ERROR ICON WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS NOT AVAILABLE FOR EVALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7310448 ↗
  • MalfunctionMar 2, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, AN ERROR ICON WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS NOT AVAILABLE FOR EVALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7310215 ↗
  • MalfunctionFeb 27, 2018Wireless Communication Problem; Wireless Communication Problem

    DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, A LOSS OF CONNECTION OCCURRED. NO DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ISSUE WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7299413 ↗
  • MalfunctionFeb 21, 2018Wireless Communication Problem; Wireless Communication Problem

    PT MOTHER REPORTS LOSS OF CONNECTION. ALLEGIES THAT THE RECEIVER IS AT FAULT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7287454 ↗
  • MalfunctionFeb 14, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, AN ERROR ICON WAS DISPLAYED ON THE RECEIVER. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS NOT AVAILABLE FOR EVALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7270545 ↗
  • MalfunctionFeb 13, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, AN ERROR ICON WAS DISPLAYED ON THE RECEIVER. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. DATA WAS NOT AVAILABLE FOR EVALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7266781 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P120005. One submission often covers many models, so check the product and lot details in each record.

  • Z-1560-2020 · Feb 3, 2020 · Terminated
    Dexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends a
  • Z-1561-2020 · Feb 3, 2020 · Terminated
    Dexcom G5 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracki

Other Sensor, Glucose, Invasive devices

All Sensor, Glucose, Invasive devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.