Dexcom G5 Mobile Replacement Receiver Kit STR-GF-001Recalls & reported problems
Dexcom, Inc. · G5 Mobile Replacement Receiver Kit Adult Black · Class 3
All models of Dexcom G5 (21) · Dexcom G5 Mobile Replacement Receiver Kit (3)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Dexcom G5 Mobile replacement receiver shows glucose readings and alerts from a sensor. Most reports describe incorrect messages, unexpected resets, alarm problems and wireless connection issues. Most reports record no patient impact. The device appears in 4,836 reports, including no deaths and 34 injuries. It has 1 closed recall from July 2019 for a software problem that could leave the receiver stuck on startup, blocking glucose values and alerts.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 4,836
- Deaths reported
- 0
- Injuries reported
- 34
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4,836 adverse event reports that list this exact device (DI 30386270000232), received between Aug 17, 2016 and Sep 10, 2020 — classified as 34 injuries, 4,802 malfunctions.
The most frequently reported specific device problem is “Device Displays Incorrect Message” (1,934 reports).
1 FDA recall record name this device's identifier.
2 other recalls cover devices cleared under the same FDA submission (P120005); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Jul 12, 2019Terminated
It has been reported that use of the mobile receiver with software version SW10617 rev 4.0.1.048 have reported: 1. the receiver becoming stuck on initialization screen when powering on. This will cause patients not to be able to receive glucose values or alerts; 2. Reinitialization of the receiver without user interaction. If this occurs, the receiver will either reboot and operate normally or require the user to press the Select button on the receiver in order to resume normal operation. In this situation, the user is provided an audio and vibratory alert every 5 minutes. Until the select button is pressed, patients will not receive glucose values or alerts.
Z-0929-2020 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Displays Incorrect Message1,934
- Inappropriate or Unexpected Reset664
- Defective Alarm493
- Wireless Communication Problem479
- Unable to Obtain Readings470
- Device Inoperable283
- Unexpected Shutdown227
- No Audible Alarm102
Patient problems
- No Consequences Or Impact To Patient4,802
- Hypoglycemia25
- Loss of consciousness10
- Seizures5
- Hyperglycemia4
- Fall3
- Cognitive Changes2
- Malaise2
Recent adverse event reports
The 25 most recent of 4,836 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 10, 2020Unable to Obtain Readings; Unable to Obtain Readings
THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. RECEIVER CHARGE AND BOOT TESTS WERE PERFORMED AND PASSED. FUNCTIONAL TESTS WERE PERFORMED AND PASSED ALL RELEVANT TESTING. AN INTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. THE RECEIVER LOG WAS DOWNLOADED AND REVIEWED FINDING ERROR RELATED TO THE COMPLAINT. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10515208 ↗MalfunctionSep 10, 2020Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10514337 ↗MalfunctionSep 3, 2020No Audible Alarm; No Audible Alarm
IT WAS REPORTED THAT THE RECEIVER HAD NO AUDIO OUTPUT OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A ROOT CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10491949 ↗MalfunctionAug 31, 2020No Audible Alarm; No Audible Alarm
SUBSEQUENT TO THE INITIAL MDR, CORRECTIONS AND ADDITIONAL INFORMATION IS REQUIRED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10473990 ↗MalfunctionAug 24, 2020Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10441722 ↗MalfunctionJul 22, 2020Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED THAT UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10307302 ↗MalfunctionJul 21, 2020Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED THAT UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10304122 ↗MalfunctionJul 21, 2020Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED THAT UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10303877 ↗MalfunctionJun 30, 2020Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10213450 ↗MalfunctionJun 11, 2020Unexpected Shutdown; Unexpected Shutdown
THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. RECEIVER CHARGE AND BOOT TESTS WERE PERFORMED AND FAILED DUE TO THE RECEIVER HAVING NO POWER. FUNCTIONAL TESTS WERE NOT PERFORMED DUE TO THE RECEIVER NOT POWERING ON. AN INTERNAL VISUAL INSPECTION WAS PERFORMED AND FAILED DUE TO A DAMAGED BATTERY. PICTURES WERE TAKEN. THE RECEIVER LOG WAS DOWNLOADED BY USING A GOOD KNOWN BATTERY. THE RECEIVER LOG WAS REVIEWED FINDING ERROR RELATED TO THE COMPLAINT. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE BATTERY DAMAGE. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10143735 ↗MalfunctionJun 3, 2020Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10114546 ↗MalfunctionJun 2, 2020Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED THAT UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10108710 ↗MalfunctionJun 1, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL/EXTERIOR VISUAL INSPECTION WAS PERFORMED AND PASSED. RECEIVER CHARGE AND BOOT. PASS. DOWNLOAD RECEIVER LOGS. PASS. A REVIEW OF THE RECEIVER LOGS WAS PERFORMED AND SIGNAL LOSS WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10106988 ↗MalfunctionJun 1, 2020Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10105107 ↗MalfunctionJun 1, 2020Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED THAT UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10104852 ↗MalfunctionMay 27, 2020Display or Visual Feedback Problem; Display or Visual Feedback Problem
SUBSEQUENT TO THE INITIAL MDR, A CORRECTION IS REQUIRED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10092320 ↗MalfunctionMay 19, 2020Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT THE RECEIVER DISPLAY WAS FROZEN. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10072601 ↗MalfunctionMay 12, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND A LOSS OF CONNECTION WAS FOUND WITHIN THE INVESTIGATION WINDOW. HOWEVER, THE DEVICE OPERATED WITHIN SPECIFICATION. THE ALLEGATION WAS NOT CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10049687 ↗MalfunctionMay 8, 2020Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10040489 ↗MalfunctionApr 22, 2020Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT THE RECEIVER DISPLAY WAS FROZEN. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9988662 ↗MalfunctionApr 15, 2020Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9964287 ↗MalfunctionApr 3, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. RECEIVER BOOT AND CHARGE WAS PERFORMED AND PASSED. DOWNLOAD RECEIVER LOG WAS PERFORMED AND PASSED. A REVIEW OF THE RECEIVER LOGS WAS PERFORMED AND SIGNAL LOSS WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9924691 ↗MalfunctionMar 30, 2020Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9896880 ↗MalfunctionMar 25, 2020Defective Alarm; Defective Alarm
THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. CHARGE AND BOOT TESTS WERE PERFORMED AND PASSED. FUNCTIONAL TESTS WERE PERFORMED AND PASSED ALL RELEVANT TESTING. MANUAL FUNCTIONAL "TRY IT" TESTS WERE PERFORMED AND PASSED. SPEAKER AUDIO WIGGLE TESTS WERE PERFORMED AND PASSED. AN INTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. THE SPEAKER RESISTANCE WAS MEASURED AND FOUND TO BE WITHIN SPECIFICATION. THE RECEIVER LOG WAS DOWNLOADED FOR REVIEW, FINDING NO ERROR RELATED TO THE COMPLAINT. THE ALLEGATION WAS NOT CONFIRMED. A PROBABLE CAUSE COULD NOT BE DETERMINED AS THE ALLEGATION WAS NOT FOUND. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9876943 ↗MalfunctionMar 24, 2020Display or Visual Feedback Problem; Display or Visual Feedback Problem
THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. CHARGE AND BOOT TESTS WERE PERFORMED AND PASSED. FUNCTIONAL TESTS WERE PERFORMED AND PASSED ALL RELEVANT TESTS. THE RECEIVER LOG WAS DOWNLOADED AND REVIEWED FINDING NO ERROR RELATED TO THE COMPLAINT. THE ALLEGATION WAS NOT CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9871726 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120005. One submission often covers many models, so check the product and lot details in each record.
- Z-1561-2020 · Feb 3, 2020 · TerminatedDexcom G5 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracki
- Z-1560-2020 · Feb 3, 2020 · TerminatedDexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends a
Other Sensor, Glucose, Invasive devices
- Dexcom G4 Platinum Sensor 9500-27Dexcom, Inc. · Sensor, Glucose, Invasive2 recalls2 deaths99,426 reports
- Dexcom G5 Transmitter Kit STT-GF-001Dexcom, Inc. · Sensor, Glucose, Invasive1 death58,932 reports
- Dexcom G5 Mobile IOS App SW10611Dexcom, Inc. · Sensor, Glucose, Invasive54,544 reports
- Dexcom G5 Transmitter 9438-06Dexcom, Inc. · Sensor, Glucose, Invasive15,360 reports
- Dexcom G5 Mobile Receiver MT22719Dexcom, Inc. · Sensor, Glucose, Invasive1 death12,594 reports
- Dexcom G4 SBT Transmitter 9438-05Dexcom, Inc. · Sensor, Glucose, Invasive1 death9,041 reports
- Dexcom G5 Mobile Receiver MT22719-BLUDexcom, Inc. · Sensor, Glucose, Invasive7,685 reports
- Dexcom G5 Mobile Receiver Kit STK-GF-001Dexcom, Inc. · Sensor, Glucose, Invasive1 recall6,345 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page