Skip to content

Dexcom G5 Mobile Receiver Kit STK-GF-PNKRecalls & reported problems

Dexcom, Inc. · G5 Mobile Receiver Kit Adult Pink · Class 3

Quick summary

This Dexcom G5 mobile receiver shows glucose readings sent from a sensor. Its 485 adverse event reports mostly describe the receiver showing incorrect messages, shutting down or resetting unexpectedly, and alarms failing to sound. There are no deaths and 4 injuries. There is 1 recall, now closed, from July 2019 about software that could leave the receiver stuck on its startup screen, so patients cannot receive glucose values or alerts.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
485
Deaths reported
0
Injuries reported
4
Recalls naming it
1

Summary

The FDA's MAUDE database contains 485 adverse event reports that list this exact device (DI 30386270000263), received between May 26, 2017 and Mar 23, 2024 — classified as 4 injuries, 481 malfunctions.

The most frequently reported specific device problem is “Device Displays Incorrect Message” (245 reports).

1 FDA recall record name this device's identifier.

2 other recalls cover devices cleared under the same FDA submission (P120005); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 12, 2019Terminated

    It has been reported that use of the mobile receiver with software version SW10617 rev 4.0.1.048 have reported: 1. the receiver becoming stuck on initialization screen when powering on. This will cause patients not to be able to receive glucose values or alerts; 2. Reinitialization of the receiver without user interaction. If this occurs, the receiver will either reboot and operate normally or require the user to press the Select button on the receiver in order to resume normal operation. In this situation, the user is provided an audio and vibratory alert every 5 minutes. Until the select button is pressed, patients will not receive glucose values or alerts.

    Z-0929-2020 · Root cause: Software design · FDA record ↗

Most reported problems

Device problems

  • Device Displays Incorrect Message245
  • Inappropriate or Unexpected Reset81
  • Device Inoperable61
  • Defective Alarm41
  • Wireless Communication Problem28
  • No Audible Alarm14
  • Use of Device Problem7
  • Display or Visual Feedback Problem5

Patient problems

  • No Consequences Or Impact To Patient481
  • Hypoglycemia2
  • Loss of consciousness2
  • Collapse1
  • Elevated Ketones/ Diabetic Ketoacidosis1
  • No Clinical Signs, Symptoms or Conditions1
  • Quadriplegia1
  • Spinal Column Injury1

Recent adverse event reports

The 25 most recent of 485 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 23, 2024Unable to Obtain Readings; Unable to Obtain Readings

    SUBSEQUENT TO THE INITIAL MDR, A CORRECTION IS REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18965984 ↗
  • MalfunctionAug 6, 2020Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT A UNEXPECTED RECEIVER SHUT-DOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A ROOT CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10376816 ↗
  • InjuryMay 28, 2020Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT UNEXPECTED RECEIVER SHUTDOWN OCCURRED. IT WAS REPORTED THAT THE PATIENT'S RECEIVER SCREEN WENT BLANK OFF AND ON, AND KEPT SHUTTING OFF. HER BLOOD GLUCOSE HAD BEEN VERY HIGH (NO SPECIFIC VALUES PROVIDED), AND HER INSULIN PUMP WAS ALSO FAILING. IT WAS STATED THAT THE PATIENT WAS LIKELY IN DIABETIC KETOACIDOSIS (DKA) AND SHE WAS INSTRUCTED TO GO TO THE EMERGENCY ROOM (ER). NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE.

    Patient problems: Elevated Ketones/ Diabetic Ketoacidosis

    MAUDE report 10097732 ↗
  • MalfunctionJun 29, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, THE RECEIVER DISPLAYED ERR 121 ICON. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR INVESTIGATION. CONFIRMATION OF THE PROBLEM AND ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7653534 ↗
  • MalfunctionJan 17, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR 68. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS RETURNED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7199480 ↗
  • MalfunctionJan 16, 2018Wireless Communication Problem; Wireless Communication Problem

    DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, A LOSS OF CONNECTION BETWEEN THE TRANSMITTER AND RECEIVER OCCURRED. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. DATA WAS RECEIVED BUT DOES NOT REFLECT THE FULL INVESTIGATION WINDOW. THE COMPLAINT CONFIRMATION OF A LOSS OF CONNECTION COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7198266 ↗
  • MalfunctionJan 16, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR121. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS RETURNED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7195709 ↗
  • MalfunctionJan 15, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR121. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS RETURNED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7190093 ↗
  • MalfunctionJan 15, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR121. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS RETURNED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7190089 ↗
  • MalfunctionJan 15, 2018Device Operates Differently Than Expected; Device Operates Differently Than Expected

    DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE PATIENT STATED THE RECEIVER STOPS SHOWING VALUES IF IT IS NOT CALIBRATED IMMEDIATELY AFTER BEING PROMPTED TO CALIBRATED. NO MEDICAL INTERVENTION WAS REPORTED. ADDITIONAL EVENT OR PATIENT INFORMATION IS NOT AVAILABLE. THE DEVICE HAS BEEN RECEIVED FOR EVALUATION.  A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE THE EVALUATION IS COMPLETE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7189998 ↗
  • MalfunctionJan 12, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR HWBBT. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS RETURNED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7188153 ↗
  • MalfunctionJan 12, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR121. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS RETURNED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7188150 ↗
  • MalfunctionJan 12, 2018Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset

    DEXCOM WAS MADE AWARE ON (B)(4) 2017 THAT ON (B)(6) 2017 THE RECEIVER INITIALIZED WITHOUT A MANUAL RESTART. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WERE PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE RECEIVER INITIALIZING WITHOUT MANUAL RESTART COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7187465 ↗
  • MalfunctionJan 12, 2018Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset

    THE COMPLAINT RECEIVER DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. THE RECEIVER WAS CHARGED AND WAS PERPETUALLY REBOOTING. THE RECEIVER WAS OPENED AND THE UI PCBA WAS DETACHED TO PERFORM A DOWNLOAD. THE RECEIVER LOG WAS REVIEWED, THERE WAS NO DATA WITHIN THE INVESTIGATION WINDOW. FUNCTIONAL TESTING WAS UNABLE TO BE PERFORMED. THE REPORTED EVENT OF INITIALIZING SCREEN WITHOUT MANUAL RESTART COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7187266 ↗
  • MalfunctionJan 12, 2018Device Inoperable; Device Inoperable

    THE COMPLAINT RECEIVER DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS EXTERNALLY VISUALLY INSPECTED AND NO DEFECTS WERE FOUND. THE RECEIVER FAILED TO CHARGE AND REBOOT. RECEIVER DOWNLOAD AND FUNCTIONAL TESTING CANNOT BE PERFORMED WITH RECEIVER IN THIS STATE. THE RECEIVER CASE WAS OPENED FOR FURTHER EVALUATION. THE MOISTURE INDICATOR WAS ACTIVATED. THE REPORTED EVENT OF RECEIVER CEASED TO FUNCTION WAS CONFIRMED. THE ROOT CAUSE WAS DETERMINED TO BE MOISTURE DAMAGE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7186915 ↗
  • MalfunctionJan 12, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR121. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS RETURNED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7186834 ↗
  • MalfunctionJan 12, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR HWBBT. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS RETURNED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7186818 ↗
  • MalfunctionJan 12, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR121. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS RETURNED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7186814 ↗
  • MalfunctionJan 11, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR121. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS RETURNED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7183136 ↗
  • MalfunctionJan 11, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR121. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. THE RECEIVER WAS RETURNED FOR EVALUATION. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. THE RECEIVER DATA LOG WAS DOWNLOADED AND REVIEWED AND FIRMWARE ERRORS WERE OBSERVED. THE REPORTED EVENT OF AN ERROR ICON DISPLAY WAS CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7182484 ↗
  • MalfunctionJan 10, 2018Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset

    DEXCOM WAS MADE AWARE ON (B)(4) 2017 THAT ON (B)(6) 2017 THE RECEIVER INITIALIZED WITHOUT A MANUAL RESTART. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WERE PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE RECEIVER INITIALIZING WITHOUT MANUAL RESTART COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7181054 ↗
  • MalfunctionJan 10, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR HWBAT. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS RETURNED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7180542 ↗
  • MalfunctionJan 10, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR121. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS RETURNED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7179428 ↗
  • MalfunctionJan 10, 2018Defective Alarm; Defective Alarm

    DEXCOM WAS MADE AWARE ON (B)(4) 2017 THAT ON (B)(6) 2017 THE RECEIVER HAD INTERMITTENT AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF INTERMITTENT AUDIO OUTPUT COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7179404 ↗
  • MalfunctionJan 9, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR 67. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS RETURNED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7177117 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P120005. One submission often covers many models, so check the product and lot details in each record.

  • Z-1560-2020 · Feb 3, 2020 · Terminated
    Dexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends a
  • Z-1561-2020 · Feb 3, 2020 · Terminated
    Dexcom G5 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracki

Other Sensor, Glucose, Invasive devices

All Sensor, Glucose, Invasive devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.