Skip to content

CONMED 270179Recalls & reported problems

Conmed Corporation · Infuser Compressor 110V 60 Hz · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 30653405048026), received between Jul 25, 2025 and Jul 25, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Unintended Electrical Shock” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unintended Electrical Shock1

Patient problems

  • Electric Shock1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 25, 2025Unintended Electrical Shock; Unintended Electrical Shock

    A SALES REPRESENTATIVE REPORTED ON BEHALF OF A CUSTOMER THAT THE 270179, INFUSER COMPRESSOR 110V 60 HZ DEVICE WAS USED IN AN UNNAMED PROCEDURE ON APPROXIMATELY ON (B)(6) 2025, AND ¿CAP TO FUSE BROKE OFF CAUSING IT TO LIGHTLY SHOCK PEOPLE (EVERYONE'S OKAY)¿. THROUGH FOLLOW-UP ASSESSMENT, THE CUSTOMER "CONFIRMED IT WAS ONLY ONE PERSON WHO GOT SHOCKED. THEY PULLED THE EQUIPMENT OUT OF THE OR RIGHT AFTER THE CASE AND IT WASN'T USED AGAIN.". THERE WAS NO REPORT OF INJURY, MEDICAL/SURGICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: Electric Shock

    MAUDE report 22609997 ↗

Other Pump, infusion devices

All Pump, infusion devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page