GIA GIA10038SRecalls & reported problems
Covidien LP · Stapler with DST Series Technology · Class 2
All models of GIA (26)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Covidien GIA stapler is a surgical stapler used during operations to cut and staple tissue. Its 457 reports mostly describe the stapler failing to fire, misfiring, and staples that do not form properly, with some reports of parts breaking off. The reports include 1 death and 63 injuries, and fell to 11 in the last 12 months from 24 in the 12 months before. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 455
- Deaths reported
- 1
- Injuries reported
- 63
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 455 adverse event reports that list this exact device (DI 30884521031672), received between Sep 22, 2016 and May 21, 2026 — classified as 1 death, 63 injuries, 391 malfunctions.
The most frequently reported specific device problem is “Failure to Fire” (229 reports).
73 other recalls cover devices cleared under the same FDA submission (K111825, K221013); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Fire229
- Misfire61
- Failure to Form Staple54
- Adverse Event Without Identified Device or Use Problem34
- Detachment of Device or Device Component31
- Break24
- Mechanics Altered17
- Difficult to Open or Close11
Patient problems
- No Clinical Signs, Symptoms or Conditions162
- No Consequences Or Impact To Patient79
- No Known Impact Or Consequence To Patient77
- No Code Available42
- Insufficient Information37
- Unspecified Tissue Injury26
- No Information21
- Tissue Breakdown21
Recent adverse event reports
The 25 most recent of 455 reports, in the reporter's own words as published by the FDA.
InjuryMay 21, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING THE PROCEDURE, THE STAPLER BECAME STUCK WHILE FIRING AND CAUSES STAPLES DEFORMATION AND NOT A REGULAR SHAPE. SUTURING WAS PERFORMED AS A STAPLE LINE REPLACEMENT. ADDITIONALLY, POST-OPERATIVE HEMORRHAGE WAS REPORTED AND THE PATIENT WAS READMITTED TO THE OPERATING ROOM. THE PATIENT WAS ADMITTED TO THE INTENSIVE CARE UNIT, AND EXPERIENCED EXTENDED HOSPITALIZATION.
Patient problems: Laceration(s); Tissue Breakdown
MAUDE report 25244240 ↗MalfunctionMay 18, 2026Failure to Fire
IT WAS REPORTED THAT DURING PROCEDURE, THE INSTRUMENT DID NOT FIRE. TO ADDRESS THE ISSUE, THE RELOAD WAS REPLACED. THERE WAS NO CONSEQUENCES OR IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25203252 ↗MalfunctionApr 23, 2026Misfire
IT WAS REPORTED THAT DURING A PROCEDURE, THE STAPLER PARTIAL FIRED DISTALLY. THE SURGEON MANUALLY REDID THE ANASTOMOSIS, AND SUTURING WAS PERFORMED AS A STAPLE LINE REPLACEMENT.
Patient problems: Failure to Anastomose
MAUDE report 24969098 ↗MalfunctionApr 7, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING A PROCEDURE, THE STAPLE LINE WAS INCOMPLETE AFTER FIRING.
Patient problems: Tissue Breakdown
MAUDE report 24807659 ↗MalfunctionApr 3, 2026Mechanics Altered
IT WAS REPORTED THAT DURING A GASTRECTOMY PROCEDURE, THE DEVICE PERFORMED THE INCISION BUT FAILED TO DEPLOY ANY STAPLES. THIS ISSUE CAUSED THE PROCEDURE TO BE EXTENDED BY APPROXIMATELY 10 MINUTES. THE FAULTY DEVICE WAS REPLACED WITH ANOTHER OPEN STAPLER TO COMPLETE THE SUTURING, AND THE PROCEDURE WAS SUCCESSFULLY RESUMED.
Patient problems: Tissue Breakdown
MAUDE report 24775460 ↗MalfunctionJan 5, 2026Failure to Align; Packaging Problem
ACCORDING TO THE REPORTER, IN AN OPEN PANCREATICODUODENECTOMY, DURING GASTRIC BODY TRANSECTION, THE TWO HALVES OF THE STAPLER WOULD NOT JOIN AND LOCK INTO PLACE AS USUAL AT THE HINGE PIN. THE STAPLE JAWS COULD NOT BE CLOSED. THE PACKAGING OF THE DEVICE WAS ALSO DAMAGED. THE DEVICE WAS NOT USED ON THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23954682 ↗MalfunctionDec 12, 2025Misfire; Misfire
ACCORDING TO THE REPORTER DURING AN OPEN PANCREATICODUODENECTOMY, WHEN ON THE RESECTION OF THE STOMACH, THE STAPLER PARTIALLY FIRED (THE FIRING COULD NOT BE COMPLETED TO THE DISTAL END). THE RESECTION WAS CONTINUED AGAIN USING A NEW STAPLER. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23796842 ↗MalfunctionNov 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER DURING A PARTIAL GASTRECTOMY, WHILE STAPLING THE GASTRIC, THERE WAS A COMPLETE STAPLE LINE DEHISCENCE WITH INTENSE CONTAMINATION OF THE PERITONEAL CAVITY. THIS RESULTED IN INFECTION.
Patient problems: Tissue Breakdown; Unspecified Infection
MAUDE report 23566110 ↗MalfunctionNov 14, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE INSTRUMENT DID NOT FIRE AT ALL. THE SURGERY WAS SUCCESSFULLY COMPLETED USING ANOTHER DEVICE OF THE SAME TYPE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23554505 ↗MalfunctionOct 17, 2025Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE DEVICE WAS NOT FIRING. A NEW OPEN STAPLER WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23321038 ↗MalfunctionSep 16, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING THE PROCEDURE, THE STAPLER DID NOT FIRE AS THE BLADE DID NOT GO UP. THE TRIGGER WAS STIFF AND COULD NOT BE SQUEEZE.
Patient problems: Insufficient Information
MAUDE report 23066326 ↗MalfunctionAug 26, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT PRE-OPERATIVELY, THE BLACK BUTTON WAS BROKEN. THE BROKEN PART DISENGAGED, BUT NO PIECES FELL INTO A PATIENT CAVITY. THE DEVICE WAS REPLACED WITH A COMPETITOR DEVICE. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22900259 ↗MalfunctionAug 25, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE BLACK PUSH BUTTON FOR STAPLING AND CUTTING DETACHED FROM THE DEVICE, RENDERING THE PLIERS UNUSABLE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22889001 ↗MalfunctionAug 6, 2025Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING A PROCEDURE, AN ATTEMPT WAS MADE TO FIRE, BUT THE BLACK FIRING KNOB DID NOT ADVANCE. THE FIRING COULD NOT BE DONE. TO RESOLVE THE ISSUE, A NEW DEVICE WAS USED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22731150 ↗MalfunctionJul 18, 2025Misfire; Misfire
ACCORDING TO THE REPORTER, DURING THE FINAL ESOPHAGUS-STOMACH ANASTOMOSIS USING THE COLLARD MODIFICATION, THE BLACK FIRING KNOB STOP PED MIDWAY. IT WAS NOTED THAT IT WAS POSSIBLE TO PULL BACK THE BLACK FIRING KNOB. ANOTHER NEW CARTRIDGE WAS TAKEN OUT TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22542799 ↗MalfunctionJun 17, 2025Mechanics Altered; Mechanics Altered
IT WAS REPORTED THAT DURING A PROCEDURE, THE BLACK PUSHER THUMB PIECE COULD BE MOVED. THE DEVICE CUT THE TISSUE BUT DID NOT STAPLE. THE PRODUCT WAS REPLACED TO RESOLVE THE ISSUE.
Patient problems: Tissue Breakdown
MAUDE report 22236693 ↗MalfunctionJun 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING USE OF THE DEVICE IN A WHIPPLE OPERATION, THE STAPLE LINE DID NOT HOLD AND SUBSEQUENTLY OPENED UP AFTER FIRING. SUTURING WAS PERFORMED AS A REPLACEMENT FOR THE STAPLE LINE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Tissue Breakdown
MAUDE report 22236163 ↗InjuryJun 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING THE USE OF THE DEVICE, THERE WERE ISSUES WITH STAPLE LINE LEAKS AFTER FIRING. AN ADDITIONAL SURGERY WAS REQUIRED AS AN INTERVENTION.
Patient problems: Tissue Breakdown
MAUDE report 22141379 ↗MalfunctionJun 3, 2025Break; Break
ACCORDING TO THE REPORTER, DURING AN OPEN BLADDER COLON PROCEDURE, WHILE CUTTING THE TISSUE, THE CARTRIDGE BROKE. TO RESOLVE THE ISSUE, A NEW DEVICE WAS USED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22136967 ↗MalfunctionMay 21, 2025Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING AN OPEN GASTRIC CANCER SURGERY, DURING SEPARATION OF THE GASTRIC TISSUE, THE BLACK PUSHER ROD COULD NOT BE PUSHED. A NEW OPEN STAPLER WAS USED TO RESOLVE THE ISSUE. AFTER THE SURGERY, THE SALES REP TRIED TO PUSH THE BLACK PUSHER ROD, BUT IT WAS STUCK AT FIRST AND IT REQUIRED A LOT OF FORCE TO PUSH IT. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22062448 ↗MalfunctionMay 2, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING AN ILEOSTOMY TAKEDOWN WITH PARTIAL BOWEL RESECTION, THE STAPLES WERE POORLY FORMED AND THE STAPLE LINE FELL APART. THE SURGERY EXPERIENCED A DELAY WHILE OBTAINING AND OPENING A NEW DEVICE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21954831 ↗MalfunctionApr 30, 2025Misfire; Misfire
IT WAS REPORTED THAT DURING A COLECTOMY, THERE WERE FIRING ISSUES DURING A PROCEDURE INVOLVING THE TWO OPEN STAPLER. THE DEVICES EXPERIENCED PARTIAL FIRING, NOT COMPLETING TO THE DISTAL END. THE DEVICES WERE REPLACED TO RESOLVE THE ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21933430 ↗MalfunctionApr 24, 2025Failure to Align; Failure to Align
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE STAPLER WAS DIFFICULT TO CLOSE ON THE TARGET TISSUE AND COULD NOT BE LOCKED MULTIPLE TIMES, PREVENTING IT FROM FIRING. THE STAPLER WAS REPLACED WITH ANOTHER OF THE SAME MODEL, WHICH FUNCTIONED WITHOUT ISSUE, ALLOWING THE OPERATION TO CONTINUE AS NORMAL.
Patient problems: Insufficient Information
MAUDE report 21905384 ↗MalfunctionApr 21, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING AN ABDOMINAL PERINEAL RESECTION, THE HANDLE OF THE STAPLER CAME OFF BEFORE IT WAS USED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21878785 ↗MalfunctionMar 28, 2025Misfire; Misfire
ACCORDING TO THE REPORTER, DURING AN OCCLUSION REPAIR WITH RESECTION, THE OPEN STAPLER DID NOT FULLY EXTEND THE SHOT. THE FIRING STOPPED MIDWAY. A NEW STAPLER WAS USED TO RESOLVE THE ISSUE. THE CASE WAS AN OCCLUSION CURE WITH RESECTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21722360 ↗
FDA clearance and classification
Cleared through Special 510(k) K221013 on May 5, 2022 (decision: Substantially Equivalent). Submitted by Covidien as “GIA Auto Suture Stapler with DST Series Technology”. The FDA review took 30 days. FDA record ↗
Cleared through 510(k) K111825 on Oct 25, 2011 (decision: Substantially Equivalent). Submitted by Covidien, Formerly US Surgical A Divison of Tyco H as “Endo Gia(Tm) Staplers, Dst Series(Tm) Gia(Tm) Staplers, Dst Series(Tm) Ta(Tm) Staplers”. The FDA review took 119 days. FDA record ↗
The FDA regulates product code GDW (staple, implantable) under 21 CFR 878.4750 as a Class II device. 21 CFR 878.4750 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K111825, K221013. One submission often covers many models, so check the product and lot details in each record.
- Z-0747-2021 · Nov 23, 2020 · CompletedCovidien TA Auto Suture Vascular Stapler with DST Series Technology 30mm-V 3. Intended for abdominal, gynecological, pediatric and thoracic surgical procedures for resection or tra
- Z-0748-2021 · Nov 23, 2020 · CompletedCovidien TA Auto Suture Vascular Loading Unit with DST Series Technology 30 mm- V3. Intended for abdominal, gynecological, pediatric and thoracic surgical procedures for resection
- Z-0001-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Gray Universal Articulating Loading Unit 30 mm - 2.0 mm, Item Code 030450 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gyn
- Z-0006-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 45 mm - 3.5 mm, Item Code 030455 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-0004-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 45 mm - 2.0 mm, Item Code 030453 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-0005-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Loading Unit 45 mm - 2.5mm, Item Code 030454 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecologic, pediatric
- Z-0007-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Loading Unit, Item Code 030457 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecologic, pediatric and thoracic
- Z-0003-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 30 mm - 3.5 mm, Item Code 030452 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-0009-2021 · Aug 12, 2020 · Open, ClassifiedCovidien Best Practices Procedure Kits containing Auto Suture Endo GIA Universal Loading Units, Parent Codes 00Z2760 - Product Usage: The Endo GIA universal staplers have applicat
- Z-0002-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 30 mm - 2.5 mm, Item Code 030451 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-0008-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 60 mm - 3.5 mm, Item Code 030458 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-1264-2020 · Jan 17, 2020 · TerminatedCovidien Endo GIA Ultra Universal Stapler 12 mm-intended for resection, transection in general surgery Item Code: EGIAUSTND
- Z-1829-2019 · May 17, 2019 · Open, ClassifiedEGIA 60 ARTICULATING MED THICK SULU , EGIA60AMT
- Z-1833-2019 · May 17, 2019 · Open, ClassifiedSIGNIA TRI-STAPLE 2.0 BLACK 45MM EXTRA THICK RELOA, SIG45AXT
- Z-1831-2019 · May 17, 2019 · Open, ClassifiedEGIA45 CT AR MD THK REL, EGIA45CTAMT
…and 58 more.
Common questions about the GIA GIA10038S
Has the GIA GIA10038S been recalled?
No FDA recall record in our data names this exact device identifier (DI 30884521031672), based on the FDA recall file dated Nov 27, 2026. However, 73 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the GIA GIA10038S?
The FDA's MAUDE database lists 455 adverse event reports for this device, received between Sep 22, 2016 and May 21, 2026. In those reports, the most common specific device problems are “Failure to Fire” (229), “Misfire” (61) and “Failure to Form Staple” (54); the most common patient problems are “Unspecified Tissue Injury” (26) and “Tissue Breakdown” (21). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the GIA GIA10038S?
1 report is classified as a death, 63 as injuries and 391 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the GIA GIA10038S FDA approved?
It was cleared through 510(k) K111825 on Oct 25, 2011; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the GIA GIA10038S MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the GIA GIA10038S, and is it still sold?
The FDA's device database lists Covidien LP as the company responsible for it (brand GIA, model GIA10038S). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other STAPLE, IMPLANTABLE devices
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,660 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,727 reports
- Signia SIGPHANDLECovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths5,350 reports
- ECHELON FLEX PSEE60AEthicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE1 recall13 deaths3,279 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,006 reports
- Endo GIA EGIA45AMTCovidien LP · STAPLE, IMPLANTABLE10 deaths2,967 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,910 reports
- ENDOPATH ATS45Ethicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE2 deaths2,044 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record