WECK IPN028130Recalls & reported problems
Teleflex Incorporated · HEMOLOK XL 6 Clips/Cart · Class 2 · Implant
All models of WECK (95)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 34026704625216), received between Feb 9, 2021 and May 20, 2024 — classified as 2 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Device Damaged Prior to Receipt by User2
- Activation, Positioning or Separation Problem1
- Device Dislodged or Dislocated1
- Failure to Seal1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Anemia1
- Failure to Anastomose1
- Hypovolemia1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 20, 2024Break; Break
IT WAS REPORTED " CLIP WAS FOUND BROKEN DURING USAGE ON THE PATIENT". THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19354567 ↗InjuryMay 1, 2024Failure to Seal; Failure to Seal
IT WAS REPORTED THAT "FOLLOWING SURGICAL INTERVENTION WITH TERMINAL ILEOSTOMY, UPON AWAKENING THE PATIENT WAS IN HEMORRHAGIC SHOCK WITH ACUTE ANEMIA. WE PROCEED WITH THE EMERGENCY RELAPAROTOMY FINDING ACTIVE ARTERIAL BLEEDING OF THE ILEOCOLIC PEDUNCLE AT THE LEVEL OF THE WELL POSITIONED AND WELL CLOSED HEM-O-LOCK CLIPS. WE PROCEED WITH SUTURING WITH A TRANSFIXED STITCH OF "PROLENE" SUTURE, WITHOUT REMOVE THE HEM-O-LOCK, ACHIEVING COMPLETE HAEMOSTASIS. A REPORT IS SENT FOR A SUSPECTED DEFECT OR MALFUNCTION OF THE MEDICAL DEVICE". THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: Anemia; Failure to Anastomose; Hypovolemia
MAUDE report 19226120 ↗MalfunctionJan 9, 2024Device Damaged Prior to Receipt by User; Device Damaged Prior to Receipt by User
IT WAS REPORTED THAT "PACKAGE WAS FOUND BROKEN DURING INSPECTION". NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18477026 ↗MalfunctionJan 9, 2024Device Damaged Prior to Receipt by User; Device Damaged Prior to Receipt by User
IT WAS REPORTED THAT "PACKAGE WAS FOUND BROKEN DURING INSPECTION". NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18476989 ↗MalfunctionNov 21, 2023Break; Break
IT WAS REPORTED THAT "DURING OPEN NEPHRECTOMY SURGERY, THE CLIP BROKE AFTER IT WAS LOCKED". THE CLIP AND ANY FRAGMENTS WERE REMOVED IT THEIR ENTIRETY FROM THE PATIENT. NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18185849 ↗MalfunctionNov 21, 2023Break; Break
IT WAS REPORTED THAT "DURING OPEN SPLENECTOMY SURGERY, THE CLIP BROKE AFTER IT WAS LOCKED". THE BROKEN CLIP WAS REMOVED FROM THE PATIENT AND NO PARTS/FRAGMENTS WERE LEFT IN THE PATIENT. NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18182285 ↗InjuryOct 24, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
REPORTED IN A MEDICAL JOURNAL, THE REPORT STATES, "A 54-YEAR-OLD MAN UNDERWENT A ROBOT-ASSISTED RADICAL PROSTATECTOMY FOR PROSTATE CANCER ON JUNE 2014. HE HAD BEEN FINE AFTER THE OPERATION, BUT URINARY FREQUENCY AND HEMATURIA OCCURRED AFTER 3 MONTHS. A 7 MM STONE SHADOW WAS FOUND IN THE BLADDER BY ULTRASOUND AND CT SCAN, AND A NON-MOBILE STONE WAS FOUND AT THE 9 O'CLOCK POSITION AT THE VESICOURETHRAL ANASTOMOSIS BY CYSTOSCOPY. THE PATIENT WAS DIAGNOSED AS A BLADDER STONE AND TRANSURETHRAL BLADDER LITHOTRIPSY WAS PERFORMED. WHEN THE SURFACE OF THE STONE WAS CRUSHED, THE CLIP THAT HAD BEEN USED DURING THE RADICAL PROSTATECTOMY APPEARED FROM INSIDE. THE CLIP WAS REMOVED WITH FORCEPS, ALL STONES WERE CRUSHED AND REMOVED, AND THE SURGERY WAS COMPLETED. THE PATIENT HAS BEEN FINE AFTER THE SURGERY AND NO MIGRATION OF THE CLIP HAS BEEN OBSERVED AFTER THAT". NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17998020 ↗MalfunctionFeb 9, 2021Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
IT WAS REPORTED THAT DURING THE LOADING OF THE CLIPS ON THE APPLIER THE CLIP WAS NOT HOLDING PROPERLY ON THE APPLIER. IT IS UNKNOWN IF THE ISSUE CAME FROM THE CLIP OR THE APPLIER. THE ISSUE WAS OBSERVED BEFORE USING THE DEVICE ON THE PATIENT. ADDITIONAL INFORMATION: 2 CLIPS FROM THE CARTRIDGE WERE NOT PROPERLY HELD BY THE APPLIER. THESE ARE NOT AVAILABLE FOR INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11304911 ↗
FDA clearance and classification
Cleared through 510(k) K133202 on Dec 3, 2013 (decision: Substantially Equivalent). Submitted by Teleflex Medical as “Hem-O-Lok Ligating Clip”. The FDA review took 47 days. FDA record ↗
The FDA regulates product code FZP (clip, implantable) under 21 CFR 878.4300 as a Class II device. 21 CFR 878.4300 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the WECK IPN028130
Has the WECK IPN028130 been recalled?
No FDA recall record in our data names this exact device identifier (DI 34026704625216), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the WECK IPN028130?
The FDA's MAUDE database lists 8 adverse event reports for this device, received between Feb 9, 2021 and May 20, 2024. In those reports, the most common specific device problems are “Break” (3), “Device Damaged Prior to Receipt by User” (2) and “Activation, Positioning or Separation Problem” (1); the most common patient problems are “Anemia” (1), “Failure to Anastomose” (1) and “Hypovolemia” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the WECK IPN028130?
None of the reports are classified as deaths, 2 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the WECK IPN028130 FDA approved?
It was cleared through 510(k) K133202 on Dec 3, 2013. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the WECK IPN028130 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.
Who makes the WECK IPN028130, and is it still sold?
The FDA's device database lists Teleflex Incorporated as the company responsible for it (brand WECK, model IPN028130). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CLIP, IMPLANTABLE devices
- LIGAMAX EL5MLEthicon Endo-Surgery, LLC · CLIP, IMPLANTABLE3 deaths3,769 reports
- Endo Clip III 176630Covidien LP · CLIP, IMPLANTABLE1 death1,943 reports
- Epix® Universal Clip Applier CA500Applied Medical Resources Corporation · Clip, Implantable1 recall1 death1,176 reports
- Endo Clip II 176657Covidien LP · CLIP, IMPLANTABLE1,176 reports
- LIGACLIP ER320Ethicon Endo-Surgery, LLC · CLIP, IMPLANTABLE1,098 reports
- Direct Drive® Clip Applier CA090Applied Medical Resources Corporation · Clip, Implantable1 death806 reports
- LAPRA-TY XC200Ethicon Inc. · CLIP, IMPLANTABLE779 reports
- Endo Clip 176625Covidien LP · CLIP, IMPLANTABLE1 death767 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.