WECK IPN914926Recalls & reported problems
Teleflex Incorporated · HORIZON TI MED 24 Clips/Pouch · Class 2 · Implant
- Adverse event reports
- 6
- Deaths reported
- 1
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 34026704696315), received between Aug 26, 2021 and Oct 26, 2022 — classified as 1 death, 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Unintended Movement” (4 reports).
1 other recall cover devices cleared under the same FDA submission (K982313); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Unintended Movement4
- Difficult to Open or Close2
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Unspecified Heart Problem1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
DeathOct 26, 2022Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED "[ONE] FULLY DISLOCATED CLIP LEADING TO RETHORACOTOMY, AFTER HARVESTING THE VEIN GRAFT 1/3 OF THE CLIPS NOT FULLY OCCLUSIVE." ADDITIONAL INFORMATION RECEIVED STATED "THERE WERE 3 BYPASSES MADE WITH A VEIN FROM THE LEG. THERE WERE AROUND 10 CLIPS ON THE SIDE BRANCHES OF THOSE VEINS. ONE WAS TOTALLY OFF AND THE TWO OR THREE WERE STILL ON THE SIDE BRANCHES BUT SEEMED TO BE NOT FULLY CLOSED; STILL DRY BUT NOT TOTALLY CLOSED". IT WAS DESCRIBED THAT AFTER TWO HOURS AFTER WOUND CLOSURE, A REPEAT THORACOTOMY WAS PROMPTED DUE TO "BIG BLEEDING, DRAIN FULL OF BLOOD, AND VERY LOW BLOOD PRESSURE." PATIENT CONDITION WAS REPORTED AS EXPIRED.
Patient problems: Unspecified Heart Problem
MAUDE report 15672526 ↗MalfunctionFeb 23, 2022Difficult to Open or Close; Difficult to Open or Close
SURGEON WAS TRYING TO CLIP ABDOMINAL VESSELS IN PATIENT, CLIPS WERE APPLIED BUT NOTED TO BE SLIPPING ON THE VESSELS; NOT HOLDING AND SLIDING UP AND DOWN. APPLIERS CHECKED AND MET TELEFLEX APPLIER CARE STANDARD. CLIPS REMOVED FROM STERILE FIELD AND TESTED - THERE APPEARS TO BE SMALL GAP AT THE BOTTOM OF THE CLIP WHEN IT IS CLOSED AND THIS IS CAUSING THE CLIP TO SLIDE. SEVERAL CLIPS WERE TESTED AND OBSERVED TO BE DOING THE SAME THING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13584298 ↗MalfunctionAug 26, 2021Unintended Movement; Unintended Movement
MULTIPLE INCIDENCE OF ONE DOCTORS PATIENTS RETURNING AFTER CV PROCEDURES. HORIZON CLIPS HAVE SLIPPED OFF POST-OP. FOLLOW-UP REPORT INDICATES THAT 4 PATIENTS WERE AFFECTED OVER THE PAST YEAR WITH HORIZON CLIP OCCLUSION ISSUES. ALL PATIENTS WERE SUBSEQUENTLY LIGATED WITH AUTOMATIC METAL CLIP APPLIER. ONE PATIENT HAD TO RECEIVE SURGICAL INTERVENTION UPON READMITTANCE. THE FIRST INCIDENT DATE IS NOT KNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12371504 ↗MalfunctionAug 26, 2021Unintended Movement; Unintended Movement
MULTIPLE INCIDENCE OF ONE DOCTORS PATIENTS RETURNING AFTER CV PROCEDURES. HORIZON CLIPS HAVE SLIPPED OFF POST-OP. FOLLOW-UP REPORT INDICATES THAT 4 PATIENTS WERE AFFECTED OVER THE PAST YEAR WITH HORIZON CLIP OCCLUSION ISSUES. ALL PATIENTS WERE SUBSEQUENTLY LIGATED WITH AUTOMATIC METAL CLIP APPLIER. ONE (1) PATIENT HAD TO RECEIVE SURGICAL INTERVENTION UPON READMITTANCE. THE FIRST INCIDENT DATE IS NOT KNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12371383 ↗InjuryAug 26, 2021Unintended Movement; Unintended Movement
MULTIPLE INCIDENCE OF ONE DOCTORS PATIENTS RETURNING AFTER CV PROCEDURES. HORIZON CLIPS HAVE SLIPPED OFF POST-OP. FOLLOW-UP REPORT INDICATES THAT 4 PATIENTS WERE AFFECTED OVER THE PAST YEAR WITH HORIZON CLIP OCCLUSION ISSUES. ALL PATIENTS WERE SUBSEQUENTLY LIGATED WITH AUTOMATIC METAL CLIP APPLIER. ONE (1) PATIENT HAD TO RECEIVE SURGICAL INTERVENTION UPON READMITTANCE. THE FIRST INCIDENT DATE IS NOT KNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12371228 ↗MalfunctionAug 26, 2021Unintended Movement; Unintended Movement
MULTIPLE INCIDENCE OF ONE DOCTORS PATIENTS RETURNING AFTER CV PROCEDURES. HORIZON CLIPS HAVE SLIPPED OFF POST-OP. FOLLOW-UP REPORT INDICATES THAT 4 PATIENTS WERE AFFECTED OVER THE PAST YEAR WITH HORIZON CLIP OCCLUSION ISSUES. ALL PATIENTS WERE SUBSEQUENTLY LIGATED WITH AUTOMATIC METAL CLIP APPLIER. ONE (1) PATIENT HAD TO RECEIVE SURGICAL INTERVENTION UPON READMITTANCE. THE FIRST INCIDENT DATE IS NOT KNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12371101 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K982313. One submission often covers many models, so check the product and lot details in each record.
- Z-2327-2016 · Aug 8, 2013 · TerminatedHorizon 24 Large Titanium Clips with Adhesive Backing, Ref 004204, Rx Only, Sterile. The metal ligating clips are intended for use in procedures involving vessels or anatomic stru
Other CLIP, IMPLANTABLE devices
- LIGAMAX EL5MLEthicon Endo-Surgery, LLC · CLIP, IMPLANTABLE3 deaths3,808 reports
- Endo Clip III 176630Covidien LP · CLIP, IMPLANTABLE1 death1,959 reports
- Epix® Universal Clip Applier CA500Applied Medical Resources Corporation · Clip, Implantable1 recall1 death1,183 reports
- Endo Clip II 176657Covidien LP · CLIP, IMPLANTABLE1,183 reports
- LIGACLIP ER320Ethicon Endo-Surgery, LLC · CLIP, IMPLANTABLE1,115 reports
- Direct Drive® Clip Applier CA090Applied Medical Resources Corporation · Clip, Implantable1 death806 reports
- LAPRA-TY XC200Ethicon Inc. · CLIP, IMPLANTABLE791 reports
- Endo Clip 176625Covidien LP · CLIP, IMPLANTABLE1 death768 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.