Dexcom G4 Platinum Professional Transmitter Kit US STT-PU-001Recalls & reported problems
Dexcom, Inc. · Transmitter Kit US (9438-01) · Class 3
All models of Dexcom G4 Platinum (41)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Dexcom G4 Platinum transmitter kit is a small wireless device that sends glucose readings from a sensor to a display. All 276 reports are malfunction reports, with no injuries or deaths. Most describe the transmitter losing its wireless signal or failing to communicate. Most also note no impact on the patient. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 276
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 276 adverse event reports that list this exact device (DI 50386270000045), received between Jan 11, 2016 and Jul 31, 2020 — classified as 276 malfunctions.
The most frequently reported specific device problem is “Wireless Communication Problem” (249 reports).
2 other recalls cover devices cleared under the same FDA submission (P120005); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Wireless Communication Problem249
- Device Displays Incorrect Message13
- No Device Output11
- Device Operates Differently Than Expected3
Patient problems
- No Consequences Or Impact To Patient264
- No Known Impact Or Consequence To Patient12
Recent adverse event reports
The 25 most recent of 276 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 31, 2020No Device Output; No Device Output
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10351511 ↗MalfunctionJul 8, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10244744 ↗MalfunctionJun 11, 2020No Device Output; No Device Output
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10146424 ↗MalfunctionMar 30, 2020No Device Output; No Device Output
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9901502 ↗MalfunctionFeb 24, 2020No Device Output; No Device Output
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A ROOT CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9747885 ↗MalfunctionFeb 6, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9677994 ↗MalfunctionDec 18, 2019Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9491731 ↗MalfunctionDec 10, 2019No Device Output; No Device Output
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9447542 ↗MalfunctionDec 4, 2019Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9414214 ↗MalfunctionNov 26, 2019Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9377760 ↗MalfunctionNov 6, 2019Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9286262 ↗MalfunctionOct 30, 2019Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT LOSS OF CONNECTION OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A ROOT CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9257771 ↗MalfunctionJun 13, 2019Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT A LOSS OF CONNECTION OCCURED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A ROOT CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8697100 ↗MalfunctionApr 10, 2019Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT LOSS OF CONNECTION OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A ROOT CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8500407 ↗MalfunctionMar 26, 2019No Device Output; No Device Output
DEXCOM WAS MADE AWARE ON (B)(4) 2019 THAT THE DISPLAY DEVICE SHOWED A TRANSMITTER FAILED ERROR ON (B)(6) 2019. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF A TRANSMITTER FAILED ERROR COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8453970 ↗MalfunctionMar 16, 2019Wireless Communication Problem; Wireless Communication Problem
DEXCOM WAS MADE AWARE ON (B)(4) 2019, THAT ON (B)(6) 2019, A LOSS OF CONNECTION OCCURRED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE COMPLAINT COULD NOT BE DETERMINED. THE ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8426795 ↗MalfunctionMar 6, 2019Wireless Communication Problem; Wireless Communication Problem
DEXCOM WAS MADE AWARE ON (B)(4) 2019, THAT ON (B)(6) 2019, A LOSS OF CONNECTION OCCURRED. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WERE PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF A LOSS OF CONNECTION COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8397421 ↗MalfunctionFeb 28, 2019Device Displays Incorrect Message; Device Displays Incorrect Message
DEXCOM WAS MADE AWARE ON (B)(4) 2019, THAT ON (B)(6) 2019, A TRANSMITTER FAILED ERROR OCCURED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE REPORTED TRANSMITTER FAILED ERROR COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8379381 ↗MalfunctionFeb 27, 2019Wireless Communication Problem; Wireless Communication Problem
DEXCOM WAS MADE AWARE ON (B)(4) 2019, THAT ON (B)(6) 2019, A LOSS OF CONNECTION OCCURRED. NO ADDITIONAL EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. LOSS OF CONNECTION COULD NOT BE DETERMINED. THE ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8375227 ↗MalfunctionFeb 19, 2019No Device Output; No Device Output
DEXCOM WAS MADE AWARE ON (B)(4) 2019, THAT ON (B)(6) 2019, PATIENT EXPERIENCED A FAILED TRANSMITTER ERROR. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WERE PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF A FAILED TRANSMITTER COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8352538 ↗MalfunctionFeb 19, 2019No Device Output; No Device Output
DEXCOM WAS MADE AWARE ON (B)(4) 2019 THAT ON (B)(6) 2019, PATIENT EXPERIENCED A FAILED TRANSMITTER ERROR. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WERE PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF A FAILED TRANSMITTER COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8352537 ↗MalfunctionFeb 19, 2019No Device Output; No Device Output
DEXCOM WAS MADE AWARE ON (B)(4) 2019 THAT THE DISPLAY DEVICE SHOWED A TRANSMITTER FAILED ERROR ON (B)(6) 2019. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF A TRANSMITTER FAILED ERROR COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8352073 ↗MalfunctionFeb 13, 2019No Device Output; No Device Output
DEXCOM WAS MADE AWARE ON (B)(4) 2019, THAT ON (B)(6) 2019, PATIENT EXPERIENCED A FAILED TRANSMITTER ERROR. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WERE PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF A FAILED TRANSMITTER COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8337730 ↗MalfunctionJan 17, 2019Wireless Communication Problem; Wireless Communication Problem
UPON FURTHER REVIEW, THIS RECORD HAS BEEN DETERMINED TO NOT BE REPORTABLE BASED ON CONFIRMATION THAT A LOSS OF CONNECTION DID NOT OCCUR. PLEASE DISREGARD INITIAL REPORTING UNDER MFR# 3004753838-2019-09410.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8259297 ↗MalfunctionJan 17, 2019Wireless Communication Problem; Wireless Communication Problem
UPON FURTHER REVIEW, THIS RECORD HAS BEEN DETERMINED TO NOT BE REPORTABLE BASED ON CONFIRMATION THAT A LOSS OF CONNECTION DID NOT OCCUR. PLEASE DISREGARD INITIAL REPORTING UNDER MFR# 3004753838-2019-09417.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8259295 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120005. One submission often covers many models, so check the product and lot details in each record.
- Z-1560-2020 · Feb 3, 2020 · TerminatedDexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends a
- Z-1561-2020 · Feb 3, 2020 · TerminatedDexcom G5 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracki
Other Sensor, Glucose, Invasive devices
- Dexcom G4 Platinum Sensor 9500-27Dexcom, Inc. · Sensor, Glucose, Invasive2 recalls2 deaths99,426 reports
- Dexcom G5 Transmitter Kit STT-GF-001Dexcom, Inc. · Sensor, Glucose, Invasive1 death58,932 reports
- Dexcom G5 Mobile IOS App SW10611Dexcom, Inc. · Sensor, Glucose, Invasive54,544 reports
- Dexcom G5 Transmitter 9438-06Dexcom, Inc. · Sensor, Glucose, Invasive15,360 reports
- Dexcom G5 Mobile Receiver MT22719Dexcom, Inc. · Sensor, Glucose, Invasive1 death12,594 reports
- Dexcom G4 SBT Transmitter 9438-05Dexcom, Inc. · Sensor, Glucose, Invasive1 death9,041 reports
- Dexcom G5 Mobile Receiver MT22719-BLUDexcom, Inc. · Sensor, Glucose, Invasive7,685 reports
- Dexcom G5 Mobile Receiver Kit STK-GF-001Dexcom, Inc. · Sensor, Glucose, Invasive1 recall6,345 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page