Sunquest Blood Bank 10.0Recalls & reported problems

CliniSys, Inc · Sunquest Blood Bank is intended for use by trained health care professionals responsible for transfusion services. It is intended to: • Record testing results of patient specimens and blood products either manually or through instrument interfaces. • Maintain a historical record of the patient’s blood bank and transfusion related data. • Display data required to assist health care professionals when qualifying patients for electronic crossmatch. • Store records of manufactured and blood product component preparation. • Record the release of manufactured and blood products for infusion. • Maintain manufactured and blood product inventory for lookback. • Print ISBT 128 blood product labels for finished and further processed blood products and products intended for transfer. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI B236BB10000), received between Aug 19, 2022 and Aug 19, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Loss of Data” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Loss of Data1

Patient problems

  • Unspecified Blood or Lymphatic problem1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 19, 2022Loss of Data; Loss of Data

    WHEN NEW GUI BLOOD TYPE COMPATIBILITIES MAINTENANCE DEFINITIONS ARE SAVED TO THE DATABASE, IF THE NUMBER OF OVERALL COMPATIBLE AND SUITABLE DEFINITIONS EXCEEDS 497 LINES OF DATA, ALL DEFINITIONS ARE REMOVED. IF MAINTENANCE IS REMOVED, AND PROBLEM CODE DEFINITIONS WERE DEFINED, PATIENT/UNIT BLOOD TYPE COMPATIBILITIES BASED ON PROBLEM CODE DEFINITIONS WILL NOT BE GENERATED. THIS COULD RESULT IN A PATIENT RECEIVING INCOMPATIBLE BLOOD. MAINTENANCE FOR PATIENT/UNIT BLOOD TYPE COMPATIBILITY (BMA 10,2) ARE NOT AFFECTED BY THIS ISSUE.

    Patient problems: Unspecified Blood or Lymphatic problem

    MAUDE report 15258905 ↗

Other Blood Establishment Computer Software And Accessories devices

All Blood Establishment Computer Software And Accessories devices →