Skip to content

Total Abscession 14000904Recalls & reported problems

Angiodynamics, Inc. · 14F x 40cm Total Abscession Biliary Drainage Catheter (Locking Pigtail) · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI H787140009040), received between Nov 15, 2018 and Jul 12, 2019 — classified as 2 injuries.

The most frequently reported specific device problem is “Crack” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Crack1
  • Detachment of Device or Device Component1

Patient problems

  • Device Embedded In Tissue or Plaque2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 12, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    THE SAME PATIENT HAS HAD MULTIPLE INTERNAL/EXTERNAL BILIARY DRAINAGE CATHETERS DISINTEGRATE. THEY WERE PLACED ON (B)(6) 2018, (B)(6) 2018, (B)(6) 2018, (B)(6) 2019, AND (B)(6) 2019. ALL WERE ANGIODYNAMICS 14FR 40CM BILIARY DRAIN TOTAL ABSCESSION CATHETERS. I ASKED IF THEY WERE USING ANYTHING CORROSIVE TO CLEAN. THE SURGEON SPOKE W/ THE PT. & HER FAMILY AND THEY STATE THAT THEY ARE FLUSHING OUT THE TUBE ONLY W/ SALINE. WHILE THE PT. WENT MORE THAN 4 WEEKS IN BETWEEN CHANGES THIS TIME AROUND, THE SURGEON SAID THAT AT TIMES HE HAS SEEN DETERIORATION IN THE TUBING IN AS LITTLE AS 3 WEEKS IN BETWEEN CHANGES, IN THE PAST. MD ATTEMPTED TO REMOVE AS MANY DRAIN PIECES AS HE COULD. PIGTAIL WAS LEFT IN THE PATIENT. IT WAS REPORTED THE DEFECTIVE DISPOSABLE DEVICE IS AVAILABLE FOR RETURN TO THE MANUFACTURER FOR EVALUATION.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 8784798 ↗
  • InjuryNov 15, 2018Crack; Crack

    CATHETER IMPLANTED (B)(6) 2018. CATHETER EXPLANTED (B)(6) 2018 DUE TO DRAINAGE CATHETER FRACTURING/BREAKING OFF INSIDE THE PATIENT. A FOREIGN BODY REMOVAL WAS THEN PERFORMED. THE DRAINAGE CATHETER WAS REPLACED. NO REPORTED HARM OR INJURY TO THE PATIENT DUE TO THIS EVENT. IT WAS REPORTED THE DEFECTIVE DISPOSABLE DEVICE IS AVAILABLE FOR RETURN TO THE MANUFACTURER FOR EVALUATION.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 8073415 ↗

Other CATHETER, BILIARY, DIAGNOSTIC devices

All CATHETER, BILIARY, DIAGNOSTIC devices →