Skip to content

samaritan PAD SAM 350PRecalls & reported problems

Heartsine Technologies Ltd · Semi-Automatic External Defibrillator · Class 3

All models of samaritan PAD (2)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Heartsine samaritan PAD SAM 350P is a semi-automatic external defibrillator, a device used to deliver a shock to restart the heart. Of its 278 reports, 270 are malfunction reports, often describing the unit failing to power on, switching itself on, or not delivering a shock, and 2 are injury reports, one describing voice prompts stopping mid-use. There were 6 deaths reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
278
Deaths reported
6
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 278 adverse event reports that list this exact device (DI M727SAM350P), received between Apr 11, 2017 and Sep 5, 2025 — classified as 6 deaths, 2 injuries, 270 malfunctions.

The most frequently reported specific device problem is “Defective Device” (141 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Defective Device141
  • Self-Activation or Keying39
  • Electrical /Electronic Property Problem38
  • Failure to Power Up33
  • Protective Measures Problem29
  • Device Operates Differently Than Expected22
  • Noise, Audible21
  • Battery Problem19

Patient problems

  • No Patient Involvement188
  • No Clinical Signs, Symptoms or Conditions93
  • No Known Impact Or Consequence To Patient28
  • Death7
  • Cardiac Arrest5
  • Insufficient Information3
  • Ventricular Fibrillation3
  • Asystole1

Recent adverse event reports

The 25 most recent of 278 reports, in the reporter's own words as published by the FDA.

  • InjurySep 5, 2025No Audible Prompt/Feedback; No Audible Prompt/Feedback

    THE CUSTOMER CONTACTED HEARTSINE TO REPORT THEIR DEVICE STOPPED GIVING VOICE PROMPTS WHILE IN USE. WITHOUT VOICE PROMPTS, A LAY USER WOULD NOT RECEIVE THE AUDIBLE INSTRUCTIONS ON HOW TO PROPERLY USE THE DEVICE ON A PATIENT. THIS ISSUE IS PATIENT RELATED AND THE PATIENT SURVIVED THE EVENT. THE CUSTOMER STATED THERE WAS A MEDICAL DELAY TO THERAPY.

    Patient problems: Cardiac Arrest

    MAUDE report 22982330 ↗
  • MalfunctionOct 11, 2021Failure to Power Up; Electrical Power Problem; Failure to Power Up; Electrical Power Problem

    THE DISTRIBUTOR CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE COULD NOT BE POWERED ON. INABILITY TO SWITCH ON A DEVICE MAY DELAY OR PREVENT DEFIBRILLATION THERAPY, WHICH COULD RESULT IN ADVERSE CONSEQUENCES. THE WAS NO PATIENT INVOLVEMENT REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12608401 ↗
  • MalfunctionSep 14, 2021Failure to Power Up; Electrical Power Problem; Failure to Power Up; Electrical Power Problem

    THE DISTRIBUTOR CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE WAS FAILING TO POWER ON. FAILURE TO POWER ON DEVICE MAY PREVENT OR DELAY DEFIBRILLATION, WHICH COULD CAUSE AN ADVERSE EVENT. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12466194 ↗
  • MalfunctionAug 13, 2021Defective Component; Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    THE DISTRIBUTOR CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE WAS SHOWING THE "SHOCK KEY STUCK" ERROR MESSAGE. INABILITY TO USE THE DEVICE SHOCK BUTTON, MAY PREVENT OR DELAY DEFIBRILLATION WHICH COULD RESULT IN ADVERSE CONSEQUENCES. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12317824 ↗
  • MalfunctionJul 20, 2021Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    REPORTED THAT THE DEVICE IS NOT DELIVERING A SHOCK DURING TESTING. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12194765 ↗
  • MalfunctionJul 2, 2021Power Problem; Failure to Power Up; Power Problem

    INABILITY TO SWITCH ON THE DEVICE COULD LEAD TO ADVERSE CONSEQUENCES FOR A PATIENT. NO PATIENT INVOLVED IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12106030 ↗
  • MalfunctionMay 14, 2021Device Displays Incorrect Message; Battery Problem

    LOW BATTERY WARNING AND BEEP WITH IN DATE PAD-PAK. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11824258 ↗
  • MalfunctionMay 6, 2021Defective Device; Device Displays Incorrect Message; Defective Device

    RED STATUS LED AND DEVICE SERVICE PROMPT. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11782085 ↗
  • MalfunctionMay 6, 2021Defective Device; Noise, Audible; Defective Device

    GREEN STATUS LED AND BEEP. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11782084 ↗
  • MalfunctionMay 6, 2021Electrical /Electronic Property Problem; Protective Measures Problem; Electrical /Electronic Property Problem

    DEVICE SERVICE REQUIRED PROMPT, BEEPING. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11782081 ↗
  • MalfunctionMay 6, 2021Device Alarm System; Difficult to Insert; Failure to Power Up; Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare

    DIFFICULTY INSERTING PAD PAK. WHEN INSERTED THE RED STATUS LED WAS ILLUMINATED AND THE DEVICE WOULD NOT POWER ON. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11782077 ↗
  • MalfunctionMay 6, 2021Failure to Shut Off

    DEVICE WON'T SWITCH OFF. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11782075 ↗
  • MalfunctionMay 6, 2021Defective Device

    RED STATUS LED WITH NEW PAD-PAK. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11782010 ↗
  • MalfunctionMay 4, 2021Failure to Power Up; Power Problem; Power Problem

    DEVICE WILL NOT POWER ON WITH PAD-PAK BUT WILL WHEN OTHERS ARE USED. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11768707 ↗
  • MalfunctionMay 4, 2021Defective Device; Output Problem; Defective Device

    GREEN STATUS LED AND BEEP. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11768704 ↗
  • MalfunctionMay 4, 2021Failure to Power Up; Failure to Shut Off; Failure to Shut Off

    DEVICE WON'T POWER OFF. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11768701 ↗
  • MalfunctionMay 4, 2021Battery Problem; Electrical Power Problem; Electrical Power Problem

    LOW BATTERY PROMPT. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11768695 ↗
  • MalfunctionApr 15, 2021Electrical /Electronic Property Problem; Appropriate Device Problem Term/Code Not Available; Electrical /Electronic Property Problem

    DEVICE SERVICE REQUIRED PROMPT. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11671595 ↗
  • MalfunctionApr 15, 2021Electrical /Electronic Property Problem; Output Problem; Electrical /Electronic Property Problem

    INTERMITTENT STATUS LED AND BEEP. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11671583 ↗
  • MalfunctionApr 7, 2021Defective Device; Protective Measures Problem; Defective Device

    GREEN STATUS LED AND BEEP. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11628492 ↗
  • MalfunctionApr 7, 2021Electrical /Electronic Property Problem; Self-Activation or Keying; Electrical /Electronic Property Problem

    DEVICE SWITCHES ON AUTOMATICALLY. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11628487 ↗
  • MalfunctionApr 7, 2021Electrical /Electronic Property Problem; Appropriate Device Problem Term/Code Not Available; Electrical /Electronic Property Problem

    RED STATUS LED AND BEEP. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11628486 ↗
  • MalfunctionApr 7, 2021Display or Visual Feedback Problem; Electrical /Electronic Property Problem; Defective Component; Display or Visual Feedback Problem; Electrical /Electronic Property Problem

    NO STATUS LED. NO PATIENT INVOLVEMENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11628483 ↗
  • MalfunctionApr 7, 2021Electrical /Electronic Property Problem; Appropriate Device Problem Term/Code Not Available; Electrical /Electronic Property Problem

    RED STATUS INDICATOR. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11628482 ↗
  • MalfunctionApr 7, 2021Self-Activation or Keying; Defective Device; Defective Device

    DEVICE SWITCHING ON AUTOMATICALLY. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11628476 ↗

Other Automated External Defibrillators (Non-Wearable) devices

All Automated External Defibrillators (Non-Wearable) devices →

Sources and updates