samaritan PAD SAM 450PRecalls & reported problems
Heartsine Technologies Ltd · Semi-Automatic External Defibrillator · Class 3
All models of samaritan PAD (2)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This HeartSine Samaritan PAD is a semi-automatic external defibrillator. The 26 reports mostly describe the device failing to turn on, flashing red lights or beeping, and batteries draining early, generally without a patient involved, and mention 1 death and no injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 26
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 26 adverse event reports that list this exact device (DI M727SAM450P), received between Jun 7, 2018 and Mar 1, 2021 — classified as 1 death, 25 malfunctions.
The most frequently reported specific device problem is “Defective Device” (13 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Defective Device13
- Electrical /Electronic Property Problem5
- Failure to Power Up5
- Device Alarm System4
- Protective Measures Problem3
- Noise, Audible2
- Self-Activation or Keying2
- Therapeutic or Diagnostic Output Failure2
Patient problems
- No Patient Involvement15
- No Clinical Signs, Symptoms or Conditions12
- Death1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 25 most recent of 26 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 1, 2021Electrical /Electronic Property Problem; Electrical /Electronic Property Problem; Protective Measures Problem; Noise, Audible
DEVICE BEEPING, INTERMITTENT GREEN/RED LED. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11391892 ↗MalfunctionMar 1, 2021Electrical /Electronic Property Problem; Electrical /Electronic Property Problem; Protective Measures Problem
NO GREEN STATUS INDICATOR. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11391463 ↗MalfunctionJan 22, 2021Connection Problem
CONNECTION FAIL BETWEEN GATEWAY AND SAM 450P. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11213522 ↗MalfunctionJan 22, 2021Failure to Power Up; Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
DEVICE FAILS TO POWER ON. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11213516 ↗MalfunctionNov 23, 2020Defective Alarm; Failure to Power Up; Defective Alarm
THE STATUS INDICATOR FLASHED RED AND IT WAS FOUND THAT THE DEVICE COULD NOT TURN ON WHEN THE ON/OFF BUTTON WAS PRESSED. THERE WAS NO PATIENT INVOLVED IN THIS EVENT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10883103 ↗MalfunctionNov 23, 2020Premature Discharge of Battery
(HEARTSINE 450P) HAS DEPLETED ITS BATTERY WITHIN THE FIRST YEAR OF USE. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10883101 ↗MalfunctionNov 5, 2020Electrical /Electronic Property Problem; Electrical /Electronic Property Problem; Self-Activation or Keying
WHEN THE POWER IS TURNED ON IN THE SELF-TEST, ALL THE LAMPS LIGHT UP. THERE WAS NO PATIENT INVOLVED IN THIS EVENT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10792449 ↗MalfunctionOct 1, 2020Electrical /Electronic Property Problem; Defective Device; Electrical /Electronic Property Problem
NO STATUS INDICATOR. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10611784 ↗MalfunctionSep 28, 2020Device Alarm System; Activation Problem; Device Alarm System; Activation Problem
DEVICE SERVICE REQUIRED PROMPT. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 10592193 ↗MalfunctionSep 4, 2020Battery Problem
THERE WAS NO PATIENT INVOLVED IN THIS EVENT.. THERE WAS AN ALERT ON THE DEVICE THAT THE BATTERY NEEDED TO BE REPLACED. USE UNTIL (B)(6) 2022 PAD PAK LOT # J0389.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10492916 ↗MalfunctionAug 17, 2020Electrical /Electronic Property Problem; Incorrect, Inadequate or Imprecise Result or Readings; Electrical /Electronic Property Problem
ALL LAMPS LIGHT UP IN SELF-TEST. THERE IS NO PATIENT INVOVLED IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 10412350 ↗MalfunctionAug 17, 2020Device Alarm System; Noise, Audible; Device Alarm System
WE HEAR A SERIES OF 3 SHORT BEEPS COME FROM THE AED, WHILE IT IS SITTING IN OUR 70 DEGREE HOME. WHEN WE GET TO THE DEVICE, THE GREEN LIGHT IS BLINKING. THE CHECK IS ALWAYS NORMAL. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 10412348 ↗MalfunctionApr 22, 2020Defective Device
DEVICE FAILED TO SAVE AN EVENT ON (B)(6) 2020. PATIENT INVOLVED EVENT, NO IMPACT ON THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9985537 ↗MalfunctionMar 10, 2020Defective Device; Therapeutic or Diagnostic Output Failure; Defective Device
STATUS LED IS NOT BLINKING. (THE READINESS INDICATOR IS NOT BLINKING ON THIS DEVICE). NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 9811033 ↗DeathJan 31, 2020Failure to Deliver Shock/Stimulation; Failure to Analyze Signal; Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
THERE WAS A PATIENT INVOLVED IN THIS EVENT. A SAMARTIAN 450P DEVICE WAS USED DURING AN ALLEGED SCA IN THE USA. THE RESPONDER REMOVED PADS FROM DEVICE AND PLACED ON THE PATIENT. IT WAS ALLEGED THAT THE UNIT KEPT CYCLING THROUGH THE PROMPTS AND WOULD NOT PROGRESS TO ANALYZE FOR SHOCK THERAPY. THE RESPONDER USED AN ALTERNATIVE DEVICE TO PROVIDE THERAPY. CUSTOMER STATED THAT THE PATIENT DID NOT SURVIVE.
Patient problems: Death
MAUDE report 9652391 ↗MalfunctionNov 25, 2019Defective Device; Failure to Power Up; Defective Device
DEVICE DOESN¿T TURN ON. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 9368065 ↗MalfunctionAug 23, 2019Defective Device; Lack of Maintenance Documentation or Guidelines; Defective Device
DEVICE SERVICE REQUIRED PROMPT. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 8924684 ↗MalfunctionJul 11, 2019Defective Device; Device Alarm System; Defective Device
DEVICE BEEPING RED STATUS INDICATOR. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 8780092 ↗MalfunctionJul 11, 2019Defective Device; Self-Activation or Keying; Defective Device
UPON INSTALLATION, DEVICE SWITCHES ON AND FUNCTIONS BUT HAS NO GREEN STATUS INDICATOR. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 8780091 ↗MalfunctionJun 4, 2019Device Alarm System
NO STATUS INDICATOR PAD-PAK INSERTED CORRECTLY. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 8666087 ↗MalfunctionMay 8, 2019Failure to Power Up; Defective Device; Defective Device
DEVICE WILL NOT SWITCH ON WITH NEW PAD-PAK 01 A3144. NO STATUS INDICATOR OR LEDS. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 8590286 ↗MalfunctionMar 11, 2019Defective Device; Therapeutic or Diagnostic Output Failure; Defective Device
RED STATUS INDICATOR AND BEEP, PAD-PAK EXPIRY MAY 2022. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 8407580 ↗MalfunctionDec 6, 2018Defective Device; Failure to Power Up; Defective Device
NO STATUS INDICATOR DEVICE WILL NOT SWITCH ON EVEN WITH ANOTHER PAD-PAK. REPORTABLE US ONLY.
Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 8137020 ↗MalfunctionNov 25, 2018Defective Device
DEVICE SERVICE REQUIRED PROMPT. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 8100415 ↗MalfunctionOct 29, 2018Defective Device; Defective Device; Protective Measures Problem
DEVICE PROMPTS 'WARNING DEVICE SERVICE REQUIRED'. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 8013460 ↗
Other Automated External Defibrillators (Non-Wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗