Photodynamic Bone Stabilization Procedure Pack USSL-1700220Recalls & reported problems
Illuminoss Medical, Inc. · Photodynamic Bone Stabilization Procedure Pack · Class 2 · Implant
All models of Photodynamic Bone Stabilization Procedure Pack (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI M986USSL17002200), received between Oct 12, 2022 and Oct 11, 2024 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Fracture” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Fracture1
- Insufficient Device Problem Information1
- Material Too Soft/Flexible1
Patient problems
- Implant Pain1
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- Post Operative Wound Infection1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 11, 2024Insufficient Device Problem Information; Material Too Soft/Flexible; Insufficient Device Problem Information; Material Too Soft/Flexible
THE REP STATED THAT WHEN THE SURGEON TRIED TO INFUSE THE IMPLANT WITH MONOMER, HE WAS UNABLE TO APPLY PRESSURE TO THE SYRINGE HANDLE AND COULD NOT GET ANY MONOMER INTO THE BALLOON. ACCORDING TO THE REP, THE PREPPED IMPLANT HAD NOT BEEN EXPOSED TO LIGHT AT ANY POINT. THEY THEN REMOVED THE IMPLANT FROM THE CANAL AND USED A NEW ONE. THE SECOND IMPLANT WAS PREPPED USING THE MONOMER SYRINGE FROM THE FIRST ATTEMPT, BUT WHEN TRYING TO APPLY THE MONOMER, THE SURGEON STATED THAT THE SYRINGE WAS CURING AND AGAIN COULD NOT INFUSE THE BALLOON. THE SECOND IMPLANT WAS REMOVED AND DISCARDED. THEY THEN TRIED A THIRD BALLOON FROM A NEW IMPLANT KIT, WHICH WAS SUCCESSFUL WITH NO ISSUES.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 20428767 ↗InjuryApr 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AN ADVERSE EVENT WAS ENTERED INTO THE ILLUMINOSS CLINICAL STUDY REGISTRY ON MARCH 26, 2024 FOR AN EVENT WHICH STARTED ON MAY 31, 2022. A 75 YEAR OLD FEMALE EXPERIENCED A TRAUMATIC FALL AND LEFT FEMUR FRACTURE WHICH WAS TREATED WITH ONE 17X280MM ILLUMINOSS IMPLANT AND A PLATE WITH 8 SCREWS. ABOUT A MONTH AND HALF AFTER THE INDEX PROCEDURE, THE PATIENT EXPERIENCED AN INFECTION AND A DEHISCENCE WOUND ON THE LEFT MEDIAL WOUND, LIKELY WITH A DRAINING SINUS TUNNEL.
Patient problems: Post Operative Wound Infection
MAUDE report 19176939 ↗InjuryOct 12, 2022Fracture; Fracture
AN ADVERSE EVENT WAS ENTERED INTO THE ILLUMINOSS CLINICAL STUDY REGISTRY ON (B)(6) 2022 FOR AN EVENT WHICH OCCURRED ON (B)(6) 2019. A 59 Y/O FEMALE PATIENT WAS TREATED ON (B)(6) 2018 WITH ILLUMINOSS IMPLANT FOR A PATHOLOGICAL FRACTURE OF THE RIGHT HUMERUS. THE PATIENT EXPERIENCED PAIN AT THE TREATED SITE OVER THE COURSE OF FOLLOW-UP VISITS, AND 296 DAYS POST IMPLANTATION ON (B)(6) 2019 THE PATIENT RECEIVED SURGICAL INTERVENTION TO EXPLANT THE ILLUMINOSS IMPLANT, RESECT THE CENTRAL SEGMENT OF TUMOR FROM THE MID HUMERUS, AND RECONSTRUCT IT WITH A CEMENTED INTERCALARY METAL REPLACEMENT TO ACHIEVE STABILIZATION OF THE ORIGINAL BONE FRACTURE.
Patient problems: Implant Pain
MAUDE report 15591383 ↗
Other Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated devices
- Photodynamic Bone Stabilization Procedure Pack USSL-1700260Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated2 deaths8 reports
- Photodynamic Bone Stabilization Procedure Pack USSL-2213260Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated1 death6 reports
- Photodynamic Bone Stabilization Procedure Pack USSL-2213220Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated4 deaths5 reports
- Photodynamic Bone Stabilization Procedure Pack USSL-2213240Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated5 reports
- Photodynamic Bone Stabilization Procedure Pack USSL-2213160Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated3 reports
- Photodynamic Bone Stabilization Procedure Pack USSL-1700280Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated3 reports
- Photodynamic Bone Stabilization Procedure Pack USML-1310090Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated1 report
- Photodynamic Bone Stabilization Procedure Pack USSL-1500260Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page