Skip to content

Photodynamic Bone Stabilization Procedure Pack USSL-1700280Recalls & reported problems

Illuminoss Medical, Inc. · Photodynamic Bone Stabilization Procedure Pack · Class 2 · Implant

All models of Photodynamic Bone Stabilization Procedure Pack (11)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI M986USSL17002800), received between Nov 22, 2022 and Oct 17, 2025 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Gel Leak” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Gel Leak1
  • Material Too Soft/Flexible1

Patient problems

  • Bone Fracture(s)1
  • Low Blood Pressure/ Hypotension1
  • Low Oxygen Saturation1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 17, 2025Gel Leak; Material Too Soft/Flexible; Gel Leak; Material Too Soft/Flexible

    AN EVENT WAS REPORTED IN WHICH THE IMPLANT DID NOT FULLY CURE. THE PROXIMAL PORTION OF THE BALLOON CURED, WHILE THE DISTAL PORTION REMAINED UNCURED. THE PHYSICIAN PLACED A SECOND BALLOON RETROGRADE TO COMPLETE FILLING OF THE LESION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23324676 ↗
  • InjuryDec 8, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ON (B)(6) 2022, AN 82 YEAR OLD FEMALE PATIENT WITH DEMENTIA AND MULTIPLE CO-MORBIDITIES WAS TREATED FOR A TRAUMATIC DISTAL FEMUR FRACTURE WITH AN ILLUMINOSS IMPLANT [SIZE USSL-1700280]. AFTER THE MONOMER WAS INFUSED INTO THE BALLOON AND THE IMPLANT CURING PROCESS WAS STARTED, THE PATIENT BECAME EXTREMELY HYPOTENSIVE, HYPOXIC, AND CODED. THE PATIENT WAS STABILIZED USING CHEMICAL CPR, CHEST COMPRESSIONS, AND BLOOD INFUSION. THE PROCEDURE PROGRESSED AND WAS COMPLETED. AS OF (B)(6) THE PATIENT OUTCOME WAS GOOD.

    Patient problems: Low Blood Pressure/ Hypotension; Low Oxygen Saturation

    MAUDE report 15942255 ↗
  • InjuryNov 22, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AN ADVERSE EVENT WAS ENTERED INTO THE ILLUMINOSS CLINICAL STUDY REGISTRY ON (B)(6) 2022 FOR AN EVENT WHICH OCCURRED ON (B)(6) 2021. ON (B)(6) 2021, AN ILLUMINOSS IMPLANT (SIZE 17MMX280MM) WAS USED TO TREAT A PATIENT WITH A NON-TRAUMATIC IMPENDING PATHOLOGIC FRACTURE OF THE LEFT HUMERUS IN A 63 YEAR-OLD MALE PATIENT WITH MULTIPLE MYELOMA. DIAPHYSEAL, SIMPLE, TRANSVERSE, CLOSED FRACTURE. SIGNIFICANT BONE LOSS WAS NOTED BY THE TREATING PHYSICIAN AT THE TIME OF FINAL REDUCTION AND IMPLANTATION. NO DEVICES WERE USED IN CONJUNCTION WITH THE ILLUMINOSS IMPLANT. A SOFT CAST/SPLIT WAS USED POST IMPLANTATION FOR FRACTURE SUPPORT. THE ADVERSE EVENT "LOSS OF OR INADEQUATE FIXATION IN BONE" WAS OBSERVED BY THE TREATING PHYSICIAN ON (B)(6) 2022. SURGICAL INTERVENTION WAS PERFORMED ON (B)(6) 2021 TO ADDRESS THIS WITH ADDITIONAL FIXATION PROCEDURE IM NAILING. ALTHOUGH THE INVESTIGATION IS STILL ONGOING INTO THE CAUSE OF THE LOSS OF FIXATION, THIS INCIDENT NECESSITATED SURGICAL INTERVENTION TO ADD ADDITIONAL FIXATION (IM NAILING) TO ACHIEVE STABILIZATION OF THE ORIGINAL BONE FRACTURE, AND THEREFORE THIS INITIAL MDR IS BEING SUBMITTED.

    Patient problems: Bone Fracture(s)

    MAUDE report 15847264 ↗

Other Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated devices

All Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page