- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI M986USSL22131600), received between Mar 18, 2023 and Dec 14, 2023 — classified as 3 injuries.
The most frequently reported specific device problem is “Fracture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture2
- Gel Leak1
- Insufficient Device Problem Information1
Patient problems
- Bone Fracture(s)1
- Fever1
- Implant Pain1
- Purulent Discharge1
- Swelling/ Edema1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryDec 14, 2023Gel Leak; Insufficient Device Problem Information; Gel Leak; Insufficient Device Problem Information
AN ADVERSE EVENT WAS ENTERED INTO THE ILLUMINOSS CLINICAL STUDY REGISTRY ON NOVEMBER 14, 2023 FOR AN EVENT WHICH OCCURRED ON (B)(6) 2022. A PATIENT UNDERWENT A PERCUTANEOUS STABILIZATION DUE TO EXTENSIVE PATHOLOGIC FRACTURES AND EXTENSIVE OSTEOLYSIS THROUGHOUT HER LEFT ACETABULUM AND HEMIPELVIS SECONDARY TO BREAST CANCER. THIS WAS FOLLOWED BY A STAGED TOTAL HIP ARTHROPLASTY. POST OPERATIVELY THE PATIENT WAS ADMITTED FOR SURGICAL DEBRIDEMENT OF THE HIP AND EXCISIONAL DEBRIDEMENT OF CHRONIC DRAINING SINUS TRACT FROM THE LEFT ILIUM, AND SEPARATE SURGICAL SITE WITH DEBRIDEMENT OF ILIOPSOAS MUSCLE IN EXCISIONAL MANNER.
Patient problems: Fever; Purulent Discharge; Swelling/ Edema
MAUDE report 18332046 ↗InjuryMay 24, 2023Fracture; Fracture
A VERY SICK PATIENT FELL AND EXPERIENCED A PERIPROSTHETIC PROXIMAL HUMERUS FRACTURE. THE PATIENT HAD EXPERIENCED A FRACTURE OF THE SAME HUMERUS 10 YEARS PREVIOUSLY WHICH HAD BEEN TREATED WITH A PLATE. 2 MONTHS AFTER THE PERIPROSTHETIC FRACTURE, THE PATIENT WAS TREATED WITH AN ILLUMINOSS IMPLANT [SIZE 22/13X160MM]. TWO WEEKS POST IMPLANTATION, THE ILLUMINOSS IMPLANT BROKE THROUGH THE FRACTURE SITE AT THE EDGE OF THE HUMERUS PLATE.
Patient problems: Bone Fracture(s)
MAUDE report 16997022 ↗InjuryMar 18, 2023Fracture; Fracture
AN ADVERSE EVENT WAS ENTERED INTO THE ILLUMINOSS CLINICAL STUDY REGISTRY ON (B)(6) 2023 FOR AN EVENT WHICH OCCURRED (B)(6), 2023. ON (B)(6) 2022, AN ILLUMINOSS IMPLANT (SIZE 22/13X160MM) WITH 3 SCREWS WAS USED TO TREAT A PATIENT WITH A TRAUMATIC FRACTURE OF THE RIGHT HUMERUS IN A 68 YEAR-OLD FEMALE PATIENT WITH HYPERTENSION AND A HISTORY OF HEMORRHAGIC STROKE. APPROXIMATELY 6 WEEKS AFTER HER INITIAL PROCEDURE THE PATIENT PRESENTED WITH WORSENING PAIN IN THE RIGHT SHOULDER. ON (B)(6) 2023 SHE UNDERWENT EXPLANT OF THE ILUMINOSS DEVICE AND IT WAS FOUND TO BE BROKEN. SHE WAS REFERRED FOR SURGICAL INTERVENTION TO ADDRESS THIS WITH REVERSE TOTAL SHOULDER ARTHROPLASTY.
Patient problems: Implant Pain
MAUDE report 16569473 ↗
Other Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated devices
- Photodynamic Bone Stabilization Procedure Pack USSL-1700260Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated2 deaths8 reports
- Photodynamic Bone Stabilization Procedure Pack USSL-2213260Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated1 death6 reports
- Photodynamic Bone Stabilization Procedure Pack USSL-2213220Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated4 deaths5 reports
- Photodynamic Bone Stabilization Procedure Pack USSL-2213240Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated5 reports
- Photodynamic Bone Stabilization Procedure Pack USSL-1700220Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated3 reports
- Photodynamic Bone Stabilization Procedure Pack USSL-1700280Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated3 reports
- Photodynamic Bone Stabilization Procedure Pack USML-1310090Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated1 report
- Photodynamic Bone Stabilization Procedure Pack USSL-1500260Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗