Photodynamic Bone Stabilization Procedure Pack USSL-2213240Recalls & reported problems
Illuminoss Medical, Inc. · Photodynamic Bone Stabilization Procedure Pack · Class 2 · Implant
All models of Photodynamic Bone Stabilization Procedure Pack (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI M986USSL22132400), received between Dec 21, 2022 and Oct 29, 2025 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture3
- Adverse Event Without Identified Device or Use Problem1
- Gel Leak1
- Material Too Soft/Flexible1
Patient problems
- Bone Fracture(s)2
- Nonunion/Delayed-union Bone Fracture2
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 29, 2025Gel Leak; Material Too Soft/Flexible; Gel Leak; Material Too Soft/Flexible
AN EVENT WAS REPORTED IN WHICH THE MONOMER DID NOT FULLY CURE. THE LED LIGHT BOX COMPLETED A FULL CYCLE; HOWEVER, UPON REMOVAL OF THE LIGHT FIBER, UNCURED MONOMER LEAKED FROM THE PROXIMAL END WHERE THE CATHETER WAS CUT. THE CENTRAL PORTION OF THE BALLOON HAD CURED, AND THE PARTIALLY INFLATED BALLOON WAS LEFT IN PLACE. ADDITIONAL FIXATION WAS ACHIEVED USING A PLATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23417108 ↗InjuryMay 25, 2023Fracture; Fracture
AN 84 YEAR-OLD FEMALE PATIENT EXPERIENCED A BREAK THROUGH THE MYELOMA IN HER HUMERUS IN LATE DECEMBER THAT WENT ON TO HYPERTROPHIC NONUNION. ON (B)(6) 2023, HER HUMERUS WAS TREATED WITH AN ILLUMINOSS IMPLANT (SIZE 22/13X240MM) WITH 4 SCREWS. IN LATE APRIL, THE PATIENT PRESENTED WITH LOSS OF FIXATION IN THE ARM, AND THE IMPLANT WAS FOUND TO BE BROKEN AT THE SITE OF THE ORIGINAL BONE BREAK. THE PATIENT WAS NOT EXPRESSING BEING IN PAIN. THE USER TREATED THE PATIENT WITH REVISION SURGERY WITH A PLATE STABILIZED BY THE ILLUMINOSS IMPLANT WHICH REMAINED IN THE BONE. THE PATIENT CONDITION POST-REVISION SURGERY HAS BEEN REPORTED AS GOOD.
Patient problems: Bone Fracture(s)
MAUDE report 17002426 ↗InjuryApr 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AN ADVERSE EVENT WAS ENTERED INTO THE ILLUMINOSS CLINICAL STUDY REGISTRY ON (B)(6) 2023 FOR AN EVENT WHICH OCCURRED ON (B)(6) 2022. ON (B)(6) 2022 , A PATIENT WAS TREATED FOR A PERIPROSETHETIC FEMUR FRACTURE WITH TWO ILLUMINOSS IMPLANTS AND A PLATE AND SCREWS. THE PATIENT WAS DIAGNOSED WITH A NON-UNION OF THE PERIPROSTHETIC FEMUR FRACTURE ON (B)(6) 2022 , AND RECEIVED REVISION SURGERY - ORIF WITH BONE GRAFTING AND MEDIAL PLATING- ON (B)(6) 2022.
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 16820829 ↗MalfunctionDec 21, 2022Fracture; Fracture
ON (B)(6) 2022, A 77 YEAR OLD MALE WAS TREATED FOR AN IMPENDING PATHOLOGIC FRACTURE AND LESIONS IN THE HUMERUS WITH AN ILLUMINOSS IMPLANT [SIZE USSL-2213240]. ON (B)(6) 2022, THE BONE AND IMPLANT FRACTURED AT THE DISTAL THIRD. ALTHOUGH THE INVESTIGATION IS STILL ONGOING INTO THE CAUSE OF THE IMPLANT BREAK, AND WHETHER OR NOT THE PATIENT SUFFERED A SERIOUS INJURY, THIS INCIDENT MAY QUALIFY AS A DEVICE MALFUNCTION, AND THEREFORE THIS INITIAL MDR IS BEING SUBMITTED.
Patient problems: Bone Fracture(s)
MAUDE report 16029039 ↗MalfunctionDec 21, 2022Fracture; Fracture
A 52 YEAR OLD MALE WAS TREATED FOR A TRAUMATIC FRACTURE OF THE PROXIMAL HUMERUS WAS WITH AN ILLUMINOSS IMPLANT [SIZE USSL-2213240] AND PLATE AND SCREWS. ON (B)(6) 2022 THE PATIENT WAS SEEN BY THE SURGEON, AND THE SURGEON BELIEVES THE IMPLANT HAD FRACTURED. THE PATIENT HAS STARTED THE HEALING PROCESS, HOWEVER, AND THE TREATING PHYSICIAN DETERMINED NO ADDITIONAL TREATMENT WAS WARRANTED AT THIS TIME, AND THE SURGEON WOULD MONITOR PATIENT'S PROGRESS. ALTHOUGH THE INVESTIGATION IS STILL ONGOING INTO THE CAUSE OF USER'S REPORT THAT THE IMPLANT MAY BE BROKEN, THIS INCIDENT MAY QUALIFY AS A DEVICE MALFUNCTION, AND THEREFORE THIS INITIAL MDR IS BEING SUBMITTED.
Patient problems: Insufficient Information; Nonunion/Delayed-union Bone Fracture
MAUDE report 16023075 ↗
Other Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated devices
- Photodynamic Bone Stabilization Procedure Pack USSL-1700260Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated2 deaths8 reports
- Photodynamic Bone Stabilization Procedure Pack USSL-2213260Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated1 death6 reports
- Photodynamic Bone Stabilization Procedure Pack USSL-2213220Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated4 deaths5 reports
- Photodynamic Bone Stabilization Procedure Pack USSL-2213160Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated3 reports
- Photodynamic Bone Stabilization Procedure Pack USSL-1700220Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated3 reports
- Photodynamic Bone Stabilization Procedure Pack USSL-1700280Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated3 reports
- Photodynamic Bone Stabilization Procedure Pack USML-1310090Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated1 report
- Photodynamic Bone Stabilization Procedure Pack USSL-1500260Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page