AFFIXUSHIP FRACTURE NAIL - LEFT recallZ-0439-2018 · Zimmer Biomet, Inc. · initiated Aug 7, 2017
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
FDA root cause category: Process change control
What the company is doing
Customers were notified via FedEx and/or email on approximately 08/07/2017. Instructions include ensure appropriate personnel are notified of the recall, locate and quarantine affected product in inventory, complete and return the Certification of Acknowledgement, return all affected product, and provide the names of hospitals or customers that would have received this product if further distributed.
Recall details
- Recall number
- Z-0439-2018
- Status
- Terminated
- Initiated by the company
- Aug 7, 2017
- Recalling firm
- Zimmer Biomet, Inc.
- Quantity
- 638 units
- Product code
- HSB
- FDA submissions
- K100238
- Distribution
- US nationwide distribution. Also Portugal and the Netherlands.
Product description
Affixus Hip Fracture Nail, LH 130 DEG 11MM X 460MM
Code information (models, lots, serial numbers)
Item No. 814611460; Lot No. (Exp. Date) UDI: 555810 (05/27/2027) UDI (01)00887868039867 (17)270527 (10)555810; 788580 (07/17/2027) UDI (01)00887868039867 (17)270717 (10)788580
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-11-460Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
Product lines
Other devices under the same FDA submission
88 devices cleared under K100238 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- AFFIXUS®HIP FRACTURE NAIL 8145-11-180Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall2 reports
- AFFIXUS®HIP FRACTURE NAIL 8143-09-180Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8143-09-340Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL 8143-11-180Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8144-09-440Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8144-11-440Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL 8145-09-180Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-09-380Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-11-280Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-11-380Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-11-340Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-11-400Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 78763).
- Z-0420-2018 · TerminatedAffixus Hip Fracture Nail, 125 DEG 11MM X 180MM
- Z-0421-2018 · TerminatedAffixus Hip Fracture Nail, 125 DEG 9MM X 180MM
- Z-0422-2018 · TerminatedAffixus Hip Fracture Nail, 130 DEG 11MM X 180MM
- Z-0423-2018 · TerminatedAffixus Hip Fracture Nail, 130 DEG 9MM X 180MM
- Z-0424-2018 · TerminatedAffixus Hip Fracture Nail, LH 125 DEG 11MM X 320MM
- Z-0425-2018 · TerminatedAffixus Hip Fracture Nail, LH 125 DEG 11MM X 360MM
- Z-0426-2018 · TerminatedAffixus Hip Fracture Nail, LH 125 DEG 11MM X 440MM
- Z-0427-2018 · TerminatedAffixus Hip Fracture Nail, LH 125 DEG 9MM X 260MM
- Z-0428-2018 · TerminatedAffixus Hip Fracture Nail, LH 125 DEG 9MM X 340MM
- Z-0429-2018 · TerminatedAffixus Hip Fracture Nail, LH 125 DEG 9MM X 380MM
- Z-0430-2018 · TerminatedAffixus Hip Fracture Nail, LH 125 DEG 9MM X 440MM
- Z-0431-2018 · TerminatedAffixus Hip Fracture Nail, LH 130 DEG 11MM X 260MM
- Z-0432-2018 · TerminatedAffixus Hip Fracture Nail, LH 130 DEG 11MM X 280MM
- Z-0433-2018 · TerminatedAffixus Hip Fracture Nail, LH 130 DEG 11MM X 300MM
- Z-0434-2018 · TerminatedAffixus Hip Fracture Nail, LH 130 DEG 11MM X 340MM
- Z-0435-2018 · TerminatedAffixus Hip Fracture Nail, LH 130 DEG 11MM X 380MM
- Z-0436-2018 · TerminatedAffixus Hip Fracture Nail, LH 130 DEG 11MM X 400MM
- Z-0437-2018 · TerminatedAffixus Hip Fracture Nail, LH 130 DEG 11MM X 420MM
- Z-0438-2018 · TerminatedAffixus Hip Fracture Nail, LH 130 DEG 11MM X 440MM
- Z-0440-2018 · TerminatedAffixus Hip Fracture Nail, LH 130 DEG 13MM X 460MM
- Z-0441-2018 · TerminatedAffixus Hip Fracture Nail, LH 130 DEG 9MM X 280MM
- Z-0442-2018 · TerminatedAffixus Hip Fracture Nail, LH 130 DEG 9MM X 300MM
- Z-0443-2018 · TerminatedAffixus Hip Fracture Nail, LH 130 DEG 9MM X 320MM
- Z-0444-2018 · TerminatedAffixus Hip Fracture Nail, LH 130 DEG 9MM X 340MM
- Z-0445-2018 · TerminatedAffixus Hip Fracture Nail, LH 130 DEG 9MM X 400MM
- Z-0446-2018 · TerminatedAffixus Hip Fracture Nail, LH 130 DEG 9MM X 420MM
- Z-0447-2018 · TerminatedAffixus Hip Fracture Nail, LH 130 DEG 9MM X 440MM
- Z-0448-2018 · TerminatedAffixus Hip Fracture Nail, RH 125 DEG 11MM X 300MM
- Z-0449-2018 · TerminatedAffixus Hip Fracture Nail, RH 125 DEG 11MM X 380MM
- Z-0450-2018 · TerminatedAffixus Hip Fracture Nail, RH 125 DEG 11MM X 440MM
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page