Artis zeego recallZ-1598-2024 · Siemens Medical Solutions USA, Inc · initiated Mar 21, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message TUBE HOT, have a break . the system cannot be operated normally, may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system
FDA root cause category: Device Design
What the company is doing
Siemens issued Urgent Medical Device Correction letter (AX034/23/S) to US accounts Director of Diagnostic Imaging and/ or Radiology Laboratory to users of Artis zee, Artis zeego, Artis Q, Artis Q.zen and Artis Q zeego systems on 3/21/24. Letter states reason for recall, health risk and action to take: As the released corrective action to install a water level sensor is not applicable to your site due to the local building setup, Siemens Healthineers is advising users of Artis systems to check the cooling units and refill with water according to the provided Instruction for Use of the Artis system. As described in the Operator Manual, we recommend that the system operator checks the water level of the cooling circuit at least every three months and refills water In addition, a label will be affixed on the start-up console of the Artis system in the control room to remind users to check the cooling water levels. As a courtesy, Siemens service organization will check the water level once outside the regular maintenance interval. Prospectively, the check of the cooling level is part of each maintenance. Please ensure that all users of the affected products within your organization and others who may need to be informed will receive the relevant safety information provided within this notice and will comply with the recommendations therein. Acknowledge Receipt of this Safety Advisory Notice
Recall details
- Recall number
- Z-1598-2024
- Status
- Open, Classified
- Initiated by the company
- Mar 21, 2024
- Posted by the FDA
- Apr 18, 2024
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 583 units
- Product code
- OWB
- FDA submissions
- K181407
- Distribution
- US Nationwide distribution.
Product description
Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zee ceiling- 10094137 Artis zee multi-purpose- 10094139 Artis zee biplane- 10094141 Artis zeego- 10280959 Artis Q floor- 10848280 Artis Q ceiling- 10848281 Artis Q biplane- 10848282 Artis Q zeego- 10848283 Artis Q.zen floor- 10848353 Artis Q.zen ceiling- 10848354 Artis Q.zen biplane- 10848355 Artis zee floor MN- 10094142 Artis zee biplane MN- 10094143
Code information (models, lots, serial numbers)
Model/UDI-DI: Artis zee floor- 4056869010045 Artis zee ceiling- 4056869010052 Artis zee multi-purpose- 4056869010076 Artis zee biplane- 4056869010069 Artis zee floor MN- 4056869010090 Artis zee biplane MN- 4056869010106 Artis zeego- 4056869010083 Artis Q floor- 4056869009971 Artis Q ceiling- 4056869009988 Artis Q biplane- 4056869009995 Artis Q zeego- 4056869010007 Artis Q.zen floor- 4056869010014 Artis Q.zen ceiling- 4056869010021 Artis Q.zen biplane- 4056869010038
Devices named in this recall
14 devices in the FDA device registry whose identifier appears in the recall's code information.
- Artis zeego 10848283Siemens Healthcare GmbH · Interventional fluoroscopic x-ray system5 recalls1 death7 reports
- Artis Q ceiling 10848281Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls1 death49 reports
- Artis Q biplane 10848282Siemens Healthcare GmbH · System, x-ray, fluoroscopic, image-intensified4 recalls1 death46 reports
- Artis zee floor 10094135Siemens Healthcare GmbH · System, x-ray, tomography, computed3 recalls2 deaths94 reports
- Artis zee ceiling 10094137Siemens Healthcare GmbH · System, x-ray, tomography, computed3 recalls91 reports
- Artis zeego 10280959Siemens Healthcare GmbH · Interventional fluoroscopic x-ray system4 recalls1 death24 reports
- Artis Q floor 10848280Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls18 reports
- Artis zee biplane 10094141Siemens Healthcare GmbH · System, x-ray, angiographic3 recalls63 reports
- Artis Q.zen ceiling 10848354Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls10 reports
- Artis Q.zen floor 10848353Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls5 reports
- Artis Q.zen biplane 10848355Siemens Healthcare GmbH · Interventional fluoroscopic x-ray system4 recalls5 reports
- Artis zee multi-purpose 10094139Siemens Healthcare GmbH · System, x-ray, tomography, computed3 recalls26 reports
- Artis zee biplane MN 10094143Siemens Healthcare GmbH · System, x-ray, tomography, computed2 recalls3 reports
- Artis zee floor MN 10094142Siemens Healthcare GmbH · System, x-ray, angiographic2 recalls0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page