Combined Spinal & EpiduralRecalls & reported problems across all models
B. Braun Medical Inc. · 3 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names at least one Combined Spinal & Epidural model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 3
We found 3 models of Combined Spinal & Epidural by B. Braun Medical Inc. in the FDA's device registry (GUDID). Together they are named in 5 adverse event reports in the FDA's MAUDE database, received between Nov 18, 2019 and Nov 12, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
3 FDA recall records name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Combined Spinal & Epidural recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Sep 18, 2025Open, Classified
Potential for the lid of the catheter connector to be in the incorrect position.
Z-0123-2026 · Root cause: Process control · Recall details · FDA record ↗
- Jul 10, 2019Terminated
The filter capacity is not achieved due to damage of the filter membrane.
Z-2467-2019 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jul 10, 2019Terminated
The filter capacity is not achieved due to damage of the filter membrane.
Z-2466-2019 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported Combined Spinal & Epidural problems
Device problems
- Break4
- Material Fragmentation1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Foreign Body In Patient2
- Discomfort1
- Needle Stick/Puncture1
- Pain1
Combined Spinal & Epidural models
3 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Combined Spinal & Epidural 333197Spinal epidural anesthesia kit1 recall3 reports
- Combined Spinal & Epidural 339114Epidural anesthesia kit1 recall2 reports
- Combined Spinal & Epidural 339115Epidural anesthesia kit1 recall0 reports
Recent Combined Spinal & Epidural adverse event reports
The 5 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionNov 12, 2025Break; Break
AS REPORTED BY THE USER FACILITY: REASON OF COMPLAINT: FX CATHETERS ARE KINKING, BEING STRETCHED TOO FAR AND SNAPPING OFF. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23534584 ↗MalfunctionSep 11, 2025Break; Break
AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION: CATHETER METAL COIL EXPOSED, AND SOME LEFT IN PATIENT CAUSING DISCOMFORT. DETAILED INQUIRY DESCRIPTION: DURAL PUNCTURE MADE WITH 27 GUAGE NEEDLE. PT FELT RIGHT SIDED DISCOMFORT/PAIN IN RIGHT HIP BUT DID NOT DESCRIBE AS PARESTHESIA. PAIN WENT AWAY WITH SPINAL NEEDLE REMOVAL. UPON THREADING THE EPIDURAL CATHETER (WENT IN SMOOTH) PT FELT THE SAME DISCOMFORT ON THE RIGHT SIDE. WENT THROUGH STANDARD PROCESS OF REMOVING EPIDURAL FOR REPLACEMENT. AFTER NEEDLE REMOVED BEGAN PULLING BACK CATHETER AND MET A LITTLE RESISTANCE. WITH GENTLE STEADY TRACTION, THE CATHETER WAS SLIDING BACK EASILY BUT FOUND TO BE SEVERED AT APPROXIMATELY THE 5CM MARK WITH THE MENTAL INTER COIL STILL INTACT. SLOW TRACTION APPLIED TO THE METAL COIL WAS BEING USED TO GENTLY REMOVE AS MUCH AS POSSIBLE. AFTER MOST OF THE MENTAL COIL WAS REMOVED IT DID BREAK, LEAVING AN UNKNOWN AMOUNT STILL IN THE BACK.
Patient problems: Discomfort; Foreign Body In Patient; Pain
MAUDE report 23026803 ↗MalfunctionMar 7, 2025Material Fragmentation; Material Fragmentation
ACCORDING TO THE EVENT DESCRIPTION: "CATHETER BROKE DOWN IN PATIENT BODY."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21549633 ↗MalfunctionJan 16, 2025Break; Break
ACCORDING EVENT DESCRIPTION: CATHETER SNAPPED DURING INSERTION OF EPIDURAL CATHETER PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21170736 ↗MalfunctionNov 18, 2019Break; Break
AS REPORTED BY THE USER FACILITY: IT WAS REPORTED THAT THE PATIENT WAS UNDERGOING THORACOTOMY FOR DIAPHRAGMATIC HERNIA REPAIR ON (B)(6) 2019. PATIENT RECEIVED AN THORACIC EPIDURAL AND WAS TAPED TO THE PATIENT'S BACK. PATIENT WAS MOVED FROM STRETCHER TO OR BED AND PROVIDER NOTICED THE CATHETER HAD COMPLETED SHEARED OFF IN HALF. THE CAUSE OF THE CATHETER SHEARING APPEARS TO HAVE BEEN SOME TENSION PLACED ON CATHETER AS THE PATIENT WAS MOVING. HALF OF THE CATHETER WAS STILL INSIDE THE PATIENT TAPED TO HER BACK. AS THIS IS A PAINFUL PROCEDURE, THE PATIENT NEEDED A PAIN CONTROL MODALITY BESIDE IV NARCOTICS AND AS PATIENT HAD JUST RECEIVE PREOPERATIVE HEPARIN, ANOTHER EPIDURAL WAS UNABLE TO BE PLACED. PATIENT RECEIVED POST OP TRUNCAL NERVE BLOCK (AN ADDITIONAL INVASIVE PROCEDURE).
Patient problems: Foreign Body In Patient; Needle Stick/Puncture
MAUDE report 9335004 ↗
Other B. Braun Medical Inc. products
- Infusomat54 models · 2,783 reports · 30 recalls
- IV Administration Set153 models · 281 reports · 68 recalls
- Hemodialysis Bloodlines9 models · 2,179 reports · 14 recalls
- Outlook25 models · 195 reports · 24 recalls
- Perifix57 models · 224 reports · 20 recalls
- Extension Set41 models · 101 reports · 16 recalls
- Design Options116 models · 132 reports · 14 recalls
- Blood Administration Set24 models · 177 reports · 13 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page