3 open recalls name at least one Dexcom G6 Receiver model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 20,711
- Deaths reported
- 0
- Injuries reported
- 97
- Recalls
- 3
We found 5 models of Dexcom G6 Receiver by Dexcom, Inc. in the FDA's device registry (GUDID). Together they are named in 20,711 adverse event reports in the FDA's MAUDE database, received between May 29, 2018 and Sep 30, 2026 — classified as 97 injuries, 20,614 malfunctions.
The most frequently reported specific device problem is “Wireless Communication Problem” (5,377 reports).
3 FDA recall records name at least one of these models by its device identifier, 3 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Dexcom G6 Receiver recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- May 12, 2025Open, Classified
Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead to untreated hypo or hyperglycemia which can cause seizures, vomiting, loss of consciousness, or death.
Z-1943-2025 · Root cause: Nonconforming Material/Component · names 2 of these models · Recall details · FDA record ↗
- Jan 27, 2025Open, Classified
Under very rare situations, the Dexcom G6 touchscreen receiver may not provide high or low glucose alarms/alerts as designed, which can result in two different conditions. First, if a receiver operating system .net error occurs when an alarm/alert should be triggered, the initial alarm/alert will not be delivered until a subsequent alarm/alert is triggered. At that time, the initial alarm/alert is delivered, but not the second one. Alarm/alerts will continue to be delayed and be one alarm/alert behind, causing delayed alarm/alerts until the receiver is reset. This condition could result in the missed detection of a hyperglycemic or hypoglycemic event. Second, a single EGV reading may be delayed for 5 minutes after the initial .net operating system error. If the user receives a single EGV reading delayed by 5 minutes, the delay is not expected to cause user harm.
Z-1204-2025 · Root cause: Software design · Recall details · FDA record ↗
- Jan 27, 2025Open, Classified
Under very rare situations, the Dexcom G6 touchscreen receiver may not provide high or low glucose alarms/alerts as designed, which can result in two different conditions. First, if a receiver operating system .net error occurs when an alarm/alert should be triggered, the initial alarm/alert will not be delivered until a subsequent alarm/alert is triggered. At that time, the initial alarm/alert is delivered, but not the second one. Alarm/alerts will continue to be delayed and be one alarm/alert behind, causing delayed alarm/alerts until the receiver is reset. This condition could result in the missed detection of a hyperglycemic or hypoglycemic event. Second, a single EGV reading may be delayed for 5 minutes after the initial .net operating system error. If the user receives a single EGV reading delayed by 5 minutes, the delay is not expected to cause user harm.
Z-1202-2025 · Root cause: Software design · Recall details · FDA record ↗
Most reported Dexcom G6 Receiver problems
Device problems
- Wireless Communication Problem5,377
- Unable to Obtain Readings3,989
- Unexpected Shutdown3,174
- Defective Alarm2,176
- Inappropriate or Unexpected Reset2,010
- Display or Visual Feedback Problem1,862
- No Audible Alarm1,391
- Low Audible Alarm315
Patient problems
- No Clinical Signs, Symptoms or Conditions13,233
- No Consequences Or Impact To Patient7,544
- Hypoglycemia61
- Loss of consciousness42
- Diaphoresis10
- Hyperglycemia10
- Confusion/ Disorientation6
- Dizziness5
Dexcom G6 Receiver models
5 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Dexcom G6 Receiver MT24078Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall15,819 reports
- Dexcom G6 Receiver STK-OE-001Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall2,845 reports
- Dexcom G6 Receiver STK-FR-001Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall1,373 reports
- Dexcom G6 Receiver STK-FM-001Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall483 reports
- Dexcom G6 Receiver STK-FE-001Integrated Continuous Glucose Monitoring System, Factory Calibrated191 reports
Recent Dexcom G6 Receiver adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Wireless Communication Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954750 ↗MalfunctionSep 30, 2026Break
It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948827 ↗MalfunctionSep 29, 2026Unexpected Shutdown
It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50936265 ↗MalfunctionSep 29, 2026Inappropriate or Unexpected Reset
It was reported that a receiver reinitialization occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50935535 ↗MalfunctionSep 29, 2026Inappropriate or Unexpected Reset
It was reported that a receiver reinitialization occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50932477 ↗MalfunctionSep 25, 2026Appropriate Device Problem Term/Code Not Available
It was reported that an alert/notification settings issue occurred. Data was not investigated on an individual level. The issue will be reviewed at a strategic level on a regular basis. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50896014 ↗MalfunctionSep 24, 2026Inappropriate or Unexpected Reset
It was reported that a receiver reinitialization occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50885449 ↗MalfunctionSep 23, 2026Inappropriate or Unexpected Reset
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50874782 ↗MalfunctionSep 23, 2026Inappropriate or Unexpected Reset
It was reported that a receiver reinitialization occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50872872 ↗MalfunctionSep 23, 2026Wireless Communication Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50862941 ↗MalfunctionSep 22, 2026Inappropriate or Unexpected Reset
It was reported that a receiver reinitialization occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50860112 ↗MalfunctionSep 22, 2026Unexpected Shutdown
It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50858073 ↗MalfunctionSep 22, 2026Unable to Obtain Readings
It was reported that error icon display occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50856289 ↗MalfunctionSep 22, 2026Wireless Communication Problem
It was reported that a signal loss occurred frequently between 8/24/2026 and 8/26/2026. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50855824 ↗MalfunctionSep 21, 2026Inappropriate or Unexpected Reset
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50849086 ↗
Other Dexcom, Inc. products
- Dexcom G613 models · 637,302 reports · 7 recalls
- Dexcom G6 Transmitter2 models · 469,968 reports
- Dexcom G732 models · 325,909 reports · 8 recalls
- Dexcom G521 models · 206,314 reports · 1 recall
- Dexcom G7 GSS (161)4 models · 141,244 reports · 1 recall
- Dexcom G4 Platinum41 models · 119,308 reports · 2 recalls
- Dexcom G7 Receiver4 models · 33,732 reports · 1 recall
- Dexcom G7 GSS (160)4 models · 33,558 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page