Dexcom G6 Receiver STK-FM-001Recalls & reported problems

Dexcom, Inc. · G6 Falcon Receiver US Medicare · Class 2

All models of Dexcom G6 (13) · Dexcom G6 Receiver (5)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Dexcom G6 receiver is a display for the G6 glucose monitor. Its 483 adverse event reports mostly describe unexpected resets and shutdowns, with some mentioning missed alarms. There are no deaths and 2 injuries, and reports fell from 255 in the prior 12 months to 87 in the last 12. There is 1 open recall, started in May 2025, over a foam or assembly problem that may cause missed audible alerts for low or high blood sugar.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
479
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 479 adverse event reports that list this exact device (DI 00386270003232), received between Jun 23, 2023 and Sep 14, 2026 — classified as 2 injuries, 477 malfunctions.

The most frequently reported specific device problem is “Inappropriate or Unexpected Reset” (201 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 6 reportsJan 2024: 7 reportsFeb 2024: 6 reportsMar 2024: 8 reportsApr 2024: 14 reportsMay 2024: 12 reportsJun 2024: 16 reportsJul 2024: 13 reportsAug 2024: 19 reportsSep 2024: 14 reportsOct 2024: 16 reportsNov 2024: 24 reportsDec 2024: 18 reportsJan 2025: 22 reportsFeb 2025: 8 reportsMar 2025: 20 reportsApr 2025: 10 reportsMay 2025: 19 reportsJun 2025: 18 reportsJul 2025: 22 reportsAug 2025: 50 reportsSep 2025: 20 reportsOct 2025: 24 reportsNov 2025: 14 reportsDec 2025: 17 reportsJan 2026: 10 reportsFeb 2026: 18 reportsMar 2026: 5 reportsApr 2026: 4 reportsMay 2026: 4 reportsJun 2026: 5 reportsJul 2026: 5 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202650
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 12, 2025Class IOpen, Classified

    Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead to untreated hypo or hyperglycemia which can cause seizures, vomiting, loss of consciousness, or death.

    Z-1943-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Inappropriate or Unexpected Reset201
  • Unexpected Shutdown83
  • Defective Alarm52
  • No Audible Alarm40
  • Display or Visual Feedback Problem35
  • Wireless Communication Problem31
  • Low Audible Alarm15
  • Unable to Obtain Readings15

Patient problems

  • No Clinical Signs, Symptoms or Conditions476
  • Diaphoresis2
  • Hypoglycemia2
  • Convulsion/Seizure1
  • High Blood Pressure/ Hypertension1
  • Loss of consciousness1

Recent adverse event reports

The 25 most recent of 479 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 14, 2026Inappropriate or Unexpected Reset

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50765454 ↗
  • MalfunctionJul 27, 2026Inappropriate or Unexpected Reset

    It was reported that a receiver reinitialization occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50012777 ↗
  • MalfunctionJul 27, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50001257 ↗
  • MalfunctionJul 15, 2026Unexpected Shutdown

    IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25827788 ↗
  • MalfunctionJul 10, 2026Unexpected Shutdown

    IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25787447 ↗
  • MalfunctionJul 10, 2026Inappropriate or Unexpected Reset

    IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25782069 ↗
  • MalfunctionJun 25, 2026Inappropriate or Unexpected Reset

    IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25629750 ↗
  • MalfunctionJun 16, 2026Inappropriate or Unexpected Reset

    IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25521850 ↗
  • MalfunctionJun 5, 2026Inappropriate or Unexpected Reset

    IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25414988 ↗
  • MalfunctionJun 5, 2026Inappropriate or Unexpected Reset

    IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25413214 ↗
  • MalfunctionJun 5, 2026Inappropriate or Unexpected Reset

    IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25410714 ↗
  • MalfunctionMay 28, 2026Defective Alarm

    IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25319433 ↗
  • MalfunctionMay 28, 2026Inappropriate or Unexpected Reset

    IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25318598 ↗
  • MalfunctionMay 5, 2026Inappropriate or Unexpected Reset

    AFTER SUBMISSION OF THE INITIAL MDR, PRODUCT WAS RECEIVED ON 6/2/2026 AND IT WAS DETERMINED THIS COMPLAINT IS NOT REPORTABLE PER (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25088576 ↗
  • MalfunctionMay 1, 2026Inappropriate or Unexpected Reset

    IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25059037 ↗
  • MalfunctionApr 22, 2026Unexpected Shutdown

    IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24956358 ↗
  • MalfunctionApr 21, 2026Inappropriate or Unexpected Reset

    IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24942838 ↗
  • MalfunctionApr 13, 2026Inappropriate or Unexpected Reset

    IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24866300 ↗
  • MalfunctionApr 2, 2026Unexpected Shutdown

    AFTER SUBMISSION OF THE INITIAL MDR, PRODUCT WAS RETURNED ON 5/12/2026 AND IT WAS DETERMINED THIS COMPLAINT IS NOT REPORTABLE PER (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24769009 ↗
  • MalfunctionMar 26, 2026Inappropriate or Unexpected Reset

    AFTER SUBMISSION OF THE INITIAL MDR PRODUCT WAS RECEIVED ON 4/20/2026. IT WAS DETERMINED THIS COMPLAINT IS NOT REPORTABLE PER (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24702490 ↗
  • MalfunctionMar 12, 2026Unexpected Shutdown

    IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24581314 ↗
  • MalfunctionMar 11, 2026Inappropriate or Unexpected Reset

    IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24576272 ↗
  • MalfunctionMar 11, 2026Unexpected Shutdown

    IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24575923 ↗
  • MalfunctionMar 3, 2026Unexpected Shutdown

    IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24503948 ↗
  • MalfunctionFeb 26, 2026Defective Alarm

    IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. PRODUCT WAS PROVIDED FOR INVESTIGATION. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. RECEIVER CHARGE AND BOOT TESTS WERE PERFORMED AND PASSED. FUNCTIONAL TEST PERFORMED AND PASSED RELEVANT TESTING. AN ALARM/ALERT MANUAL TEST WAS PERFORMED AND PASSED. AN INTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. PICTURES WERE TAKEN. THE SPEAKER AND VIBRATOR RESISTANCES WERE MEASURED AND PASSED. PERFORMANCE DATA WAS REVIEWED. AN ALERT/NOTIFICATION SETTINGS ISSUE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24458809 ↗

FDA clearance and classification

The FDA regulates product code QBJ (integrated continuous glucose monitoring system, factory calibrated) under 21 CFR 862.1355 as a Class II device. 21 CFR 862.1355 ↗

Devices of this type normally need 510(k) clearance before sale.

FDA definition of this device type: An integrated continuous glucose monitoring system (iCGM) is intended to automatically measure glucose in bodily fluids continuously or frequently for a specified period of time. iCGM systems are designed to reliably and securely transmit glucose measurement data to digitally connected devices, including automated insulin dosing systems, and are intended to be used alone or in conjunction with these digitally connected medical devices for the purpose of managing a disease or condition related to glycemic control.

Common questions about the Dexcom G6 Receiver STK-FM-001

Has the Dexcom G6 Receiver STK-FM-001 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1943-2025 (Class I), began on May 12, 2025: Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Dexcom G6 Receiver STK-FM-001?

The FDA's MAUDE database lists 479 adverse event reports for this device, received between Jun 23, 2023 and Sep 14, 2026. In those reports, the most common specific device problems are “Inappropriate or Unexpected Reset” (201), “Unexpected Shutdown” (83) and “Defective Alarm” (52); the most common patient problems are “Diaphoresis” (2), “Hypoglycemia” (2) and “Convulsion/Seizure” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Dexcom G6 Receiver STK-FM-001?

None of the reports are classified as deaths, 2 as injuries and 477 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Dexcom G6 Receiver STK-FM-001, and is it still sold?

The FDA's device database lists Dexcom, Inc. as the company responsible for it (brand Dexcom G6 Receiver, model STK-FM-001). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Integrated Continuous Glucose Monitoring System, Factory Calibrated devices

All Integrated Continuous Glucose Monitoring System, Factory Calibrated devices →