Dexcom G6 Receiver MT24078Recalls & reported problems

Dexcom, Inc. · G6 Receiver · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Dexcom G6 receiver has 15,817 adverse event reports in the FDA database, including 67 injuries and no deaths. Reports often describe wireless connection problems, unexpected shutdowns, and receivers that could not get readings. Reports in the last 12 months (57) are well below the 12 months before (324). There is 1 open recall, started in January 2025, because in rare cases the receiver may not sound high or low glucose alarms as designed after a software error.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
15,819
Deaths reported
0
Injuries reported
67
Recalls naming it
1

Summary

The FDA's MAUDE database contains 15,819 adverse event reports that list this exact device (DI 00386270000583), received between May 29, 2018 and Sep 21, 2026 — classified as 67 injuries, 15,752 malfunctions.

The most frequently reported specific device problem is “Wireless Communication Problem” (4,786 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 145 reportsDec 2023: 87 reportsJan 2024: 97 reportsFeb 2024: 45 reportsMar 2024: 41 reportsApr 2024: 79 reportsMay 2024: 90 reportsJun 2024: 93 reportsJul 2024: 80 reportsAug 2024: 63 reportsSep 2024: 50 reportsOct 2024: 58 reportsNov 2024: 31 reportsDec 2024: 0 reportsJan 2025: 34 reportsFeb 2025: 4 reportsMar 2025: 6 reportsApr 2025: 2 reportsMay 2025: 37 reportsJun 2025: 59 reportsJul 2025: 47 reportsAug 2025: 49 reportsSep 2025: 35 reportsOct 2025: 20 reportsNov 2025: 6 reportsDec 2025: 10 reportsJan 2026: 11 reportsFeb 2026: 7 reportsMar 2026: 9 reportsApr 2026: 3 reportsMay 2026: 3 reportsJun 2026: 3 reportsJul 2026: 2 reportsAug 2026: 3 reportsSep 2026: 2 reportsOct 2026: 0 reports202420252026145
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 27, 2025Open, Classified

    Under very rare situations, the Dexcom G6 touchscreen receiver may not provide high or low glucose alarms/alerts as designed, which can result in two different conditions. First, if a receiver operating system .net error occurs when an alarm/alert should be triggered, the initial alarm/alert will not be delivered until a subsequent alarm/alert is triggered. At that time, the initial alarm/alert is delivered, but not the second one. Alarm/alerts will continue to be delayed and be one alarm/alert behind, causing delayed alarm/alerts until the receiver is reset. This condition could result in the missed detection of a hyperglycemic or hypoglycemic event. Second, a single EGV reading may be delayed for 5 minutes after the initial .net operating system error. If the user receives a single EGV reading delayed by 5 minutes, the delay is not expected to cause user harm.

    Z-1202-2025 · Root cause: Software design · FDA record ↗

Most reported problems

Device problems

  • Wireless Communication Problem4,786
  • Unable to Obtain Readings3,182
  • Unexpected Shutdown2,349
  • Defective Alarm1,764
  • Display or Visual Feedback Problem1,391
  • No Audible Alarm969
  • Inappropriate or Unexpected Reset837
  • Low Audible Alarm240

Patient problems

  • No Clinical Signs, Symptoms or Conditions8,370
  • No Consequences Or Impact To Patient7,544
  • Hypoglycemia41
  • Loss of consciousness31
  • Hyperglycemia6
  • Diaphoresis4
  • Dizziness4
  • Seizures4

Recent adverse event reports

The 25 most recent of 15,819 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 21, 2026Inappropriate or Unexpected Reset

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50849086 ↗
  • MalfunctionSep 15, 2026Inappropriate or Unexpected Reset

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50780392 ↗
  • MalfunctionAug 28, 2026Unexpected Shutdown

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50501874 ↗
  • MalfunctionAug 28, 2026Unable to Obtain Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50500430 ↗
  • MalfunctionAug 26, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50460272 ↗
  • MalfunctionJul 10, 2026No Audible Alarm

    IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25779755 ↗
  • MalfunctionJul 7, 2026Display or Visual Feedback Problem

    IT WAS REPORTED THAT A DISPLAY ISSUE OCCURRED. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25739168 ↗
  • MalfunctionJun 9, 2026Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAY OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25451187 ↗
  • MalfunctionJun 5, 2026Wireless Communication Problem

    IT WAS REPORTED THAT"SIGNAL LOSS" OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR INVESTIGATION. THE ALLEGATION AND THE PROBABLE CAUSE CANNOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25403786 ↗
  • MalfunctionJun 3, 2026Wireless Communication Problem

    IT WAS REPORTED THAT"SIGNAL LOSS" OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR INVESTIGATION. THE ALLEGATION AND THE PROBABLE CAUSE CANNOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25370043 ↗
  • MalfunctionMay 27, 2026Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAY OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25311122 ↗
  • MalfunctionMay 19, 2026Wireless Communication Problem

    IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25219774 ↗
  • MalfunctionMay 15, 2026Wireless Communication Problem

    IT WAS REPORTED THAT "SIGNAL LOSS" OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR INVESTIGATION. THE ALLEGATION AND THE PROBABLE CAUSE CANNOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25194690 ↗
  • MalfunctionApr 30, 2026No Audible Alarm

    IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25046455 ↗
  • MalfunctionApr 14, 2026Unexpected Shutdown

    IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24877129 ↗
  • MalfunctionApr 14, 2026Defective Alarm

    IT WAS REPORTED THAT NO ALERT/NOTIFICATION OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERIOR VISUAL INSPECTION WAS PERFORMED AND PASSED. RECEIVER CHARGE AND BOOT TESTS WERE PERFORMED AND PASSED. FUNCTIONAL TESTS WERE PERFORMED AND PASSED. AN ALARM/ALERT MANUAL TEST WAS PERFORMED AND PASSED. THE SPEAKER/VIBRATOR RESISTANCE WAS MEASURED AND PASSED. AN INTERNAL VISUAL INSPECTION AND PASSED. PERFORMANCE DATA WAS REVIEWED FINDING ERRORS RELATED TO THE COMPLAINT. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE DELAYED ALERTS DUE TO DISPATCH ERROR. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24874222 ↗
  • MalfunctionMar 31, 2026Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAY OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24749021 ↗
  • MalfunctionMar 26, 2026No Audible Alarm

    IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. PRODUCT WAS PROVIDED FOR INVESTIGATION. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. RECEIVER CHARGE AND BOOT TESTS WERE PERFORMED AND PASSED. FUNCTIONAL TEST PERFORMED AND PASSED. AN ALARM/ALERT MANUAL TEST WAS PERFORMED AND PASSED. AN INTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. PICTURES WERE TAKEN. THE SPEAKER AND VIBRATOR RESISTANCES WERE MEASURED AND PASSED. PERFORMANCE DATA WAS REVIEWED. AN ALERT/NOTIFICATION SETTINGS ISSUE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24701674 ↗
  • MalfunctionMar 25, 2026Unexpected Shutdown

    IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24688740 ↗
  • MalfunctionMar 19, 2026Unexpected Shutdown

    IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24645911 ↗
  • MalfunctionMar 18, 2026Defective Alarm

    IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24633111 ↗
  • MalfunctionMar 18, 2026Defective Alarm

    IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24625859 ↗
  • MalfunctionMar 13, 2026Wireless Communication Problem

    IT WAS REPORTED THAT "SIGNAL LOSS" OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR INVESTIGATION. THE ALLEGATION AND THE PROBABLE CAUSE CANNOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24594767 ↗
  • MalfunctionMar 13, 2026Wireless Communication Problem

    IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24587903 ↗
  • MalfunctionMar 6, 2026Wireless Communication Problem

    IT WAS REPORTED THAT"SIGNAL LOSS" OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR INVESTIGATION. THE ALLEGATION AND THE PROBABLE CAUSE CANNOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24539199 ↗

Other Integrated Continuous Glucose Monitoring System, Factory Calibrated devices

All Integrated Continuous Glucose Monitoring System, Factory Calibrated devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.