Dexcom G6 Receiver STK-FE-001Recalls & reported problems

Dexcom, Inc. · G6 Falcon Receiver DME · Class 2

All models of Dexcom G6 (13) · Dexcom G6 Receiver (5)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Dexcom G6 receiver is a handheld display for readings from a body-worn glucose sensor. Many of its 188 reports describe unexpected resets or shutdowns, and some describe alarms that did not sound. Most reports say no patient was affected. There is 1 injury report and no deaths. No recalls name this device. Reports fell from 100 in the year before to 51 in the latest year.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
186
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 186 adverse event reports that list this exact device (DI 00386270002860), received between Jan 5, 2024 and Sep 23, 2026 — classified as 2 injuries, 184 malfunctions.

The most frequently reported specific device problem is “Inappropriate or Unexpected Reset” (79 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 4 reportsMay 2024: 3 reportsJun 2024: 4 reportsJul 2024: 2 reportsAug 2024: 4 reportsSep 2024: 9 reportsOct 2024: 9 reportsNov 2024: 8 reportsDec 2024: 7 reportsJan 2025: 5 reportsFeb 2025: 4 reportsMar 2025: 9 reportsApr 2025: 7 reportsMay 2025: 6 reportsJun 2025: 9 reportsJul 2025: 5 reportsAug 2025: 17 reportsSep 2025: 11 reportsOct 2025: 12 reportsNov 2025: 10 reportsDec 2025: 10 reportsJan 2026: 4 reportsFeb 2026: 5 reportsMar 2026: 5 reportsApr 2026: 5 reportsMay 2026: 2 reportsJun 2026: 1 reportsJul 2026: 5 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202617
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Inappropriate or Unexpected Reset79
  • Unexpected Shutdown49
  • No Audible Alarm19
  • Defective Alarm14
  • Display or Visual Feedback Problem12
  • Wireless Communication Problem10
  • Appropriate Device Problem Term/Code Not Available1
  • Low Audible Alarm1

Patient problems

  • No Clinical Signs, Symptoms or Conditions184
  • Hypoglycemia2
  • Loss of consciousness1

Recent adverse event reports

The 25 most recent of 186 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50862941 ↗
  • InjurySep 4, 2026Defective Alarm

    It was reported that an alert/notification settings issue occurred. According to the patient¿s spouse, on (b)(6) 2026, the patient woke up in the morning and went to the bathroom, where he eventually lost consciousness. The spouse found the patient unconscious minutes after and immediately administered Baqsimi intranasally. The patient regained consciousness within a few minutes and was given a glass of milk. No further treatment was reported to be needed, and the patient did not go to the hospital. The spouse reported that the Dexcom receiver displayed a glucose reading of 70 mg/dL, following the incident; however, no audible alert or alarm was issued to indicate a low glucose level, and neither the patient nor the spouse received any warning notification prior to the loss of consciousness. There was no documentation of further medical evaluation or intervention. At the time of the report the patient was stable. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No additional patient or event information is available.

    Patient problems: Hypoglycemia; Loss of consciousness

    MAUDE report 50621145 ↗
  • MalfunctionJul 28, 2026Inappropriate or Unexpected Reset

    It was reported that a receiver reinitialization occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50057620 ↗
  • MalfunctionJul 16, 2026Inappropriate or Unexpected Reset

    IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25845676 ↗
  • MalfunctionJul 13, 2026Wireless Communication Problem

    IT WAS REPORTED THAT \"SIGNAL LOSS\" OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR INVESTIGATION. THE ALLEGATION AND THE PROBABLE CAUSE CANNOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25805841 ↗
  • MalfunctionJul 7, 2026Defective Alarm

    IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25738908 ↗
  • MalfunctionJul 7, 2026Unexpected Shutdown

    IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25738342 ↗
  • MalfunctionJun 26, 2026Unexpected Shutdown

    IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25636684 ↗
  • MalfunctionMay 7, 2026Inappropriate or Unexpected Reset

    IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25117608 ↗
  • MalfunctionMay 6, 2026Inappropriate or Unexpected Reset

    IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25096860 ↗
  • MalfunctionApr 29, 2026Inappropriate or Unexpected Reset

    IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25034237 ↗
  • MalfunctionApr 29, 2026Inappropriate or Unexpected Reset

    IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25031108 ↗
  • MalfunctionApr 22, 2026Wireless Communication Problem

    IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED FREQUENTLY BETWEEN (B)(6) 2026. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24955986 ↗
  • MalfunctionApr 13, 2026Unexpected Shutdown

    IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24863124 ↗
  • MalfunctionApr 7, 2026No Audible Alarm

    IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24805656 ↗
  • MalfunctionMar 30, 2026Defective Alarm

    IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24709101 ↗
  • MalfunctionMar 26, 2026Inappropriate or Unexpected Reset

    IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24702473 ↗
  • MalfunctionMar 16, 2026Display or Visual Feedback Problem

    IT WAS REPORTED THAT A DISPLAY ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24611744 ↗
  • MalfunctionMar 9, 2026Inappropriate or Unexpected Reset

    IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24551898 ↗
  • MalfunctionMar 2, 2026Inappropriate or Unexpected Reset

    IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24484317 ↗
  • MalfunctionFeb 25, 2026Premature End-of-Life Indicator

    IT WAS REPORTED THAT EARLY SENSOR EXPIRATION OCCURRED. NO RELEVANT PRODUCT OR DATA RELATED TO THE ISSUE WAS PROVIDED FOR INVESTIGATION. CONFIRMATION OF THE COMPLAINT IS UNDETERMINED AND PROBABLE CAUSE CANNOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24445492 ↗
  • MalfunctionFeb 24, 2026Unexpected Shutdown

    IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24437194 ↗
  • MalfunctionFeb 13, 2026No Audible Alarm

    IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24338160 ↗
  • MalfunctionFeb 13, 2026Unexpected Shutdown

    IT WAS REPORTED THAT RECEIVER POWER ISSUE OCCURRED. PRODUCT WAS PROVIDED FOR INVESTIGATION. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. RECEIVER CHARGE WAS PERFORMED AND PASSED. RECEIVER BOOT WAS PERFORMED AND PASSED. FUNCTIONAL TEST WAS PERFORMED AND PASSED. THE RECEIVER LOG WAS DOWNLOADED. A RECEIVER CHARGE OR BATTERY ISSUE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. HOWEVER, AN UNEXPECTED RECEIVER SHUTDOWN WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE PROBABLE CAUSE OF THE UNEXPECTED RECEIVER SHUTDOWN WAS DETERMINED TO BE 170: ROOT CAUSE COULD NOT BE DETERMINED, PRODUCT/DATA RECEIVED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24336067 ↗
  • MalfunctionFeb 11, 2026Inappropriate or Unexpected Reset

    IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24319257 ↗

FDA clearance and classification

The FDA regulates product code QBJ (integrated continuous glucose monitoring system, factory calibrated) under 21 CFR 862.1355 as a Class II device. 21 CFR 862.1355 ↗

Devices of this type normally need 510(k) clearance before sale.

FDA definition of this device type: An integrated continuous glucose monitoring system (iCGM) is intended to automatically measure glucose in bodily fluids continuously or frequently for a specified period of time. iCGM systems are designed to reliably and securely transmit glucose measurement data to digitally connected devices, including automated insulin dosing systems, and are intended to be used alone or in conjunction with these digitally connected medical devices for the purpose of managing a disease or condition related to glycemic control.

Common questions about the Dexcom G6 Receiver STK-FE-001

Has the Dexcom G6 Receiver STK-FE-001 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00386270002860), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Dexcom G6 Receiver STK-FE-001?

The FDA's MAUDE database lists 186 adverse event reports for this device, received between Jan 5, 2024 and Sep 23, 2026. In those reports, the most common specific device problems are “Inappropriate or Unexpected Reset” (79), “Unexpected Shutdown” (49) and “No Audible Alarm” (19); the most common patient problems are “Hypoglycemia” (2) and “Loss of consciousness” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Dexcom G6 Receiver STK-FE-001?

None of the reports are classified as deaths, 2 as injuries and 184 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Dexcom G6 Receiver STK-FE-001, and is it still sold?

The FDA's device database lists Dexcom, Inc. as the company responsible for it (brand Dexcom G6 Receiver, model STK-FE-001). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.