Dexcom G6 Receiver STK-FE-001Recalls & reported problems
Dexcom, Inc. · G6 Falcon Receiver DME · Class 2
Quick summary
This Dexcom G6 receiver is a handheld display for readings from a body-worn glucose sensor. Many of its 188 reports describe unexpected resets or shutdowns, and some describe alarms that did not sound. Most reports say no patient was affected. There is 1 injury report and no deaths. No recalls name this device. Reports fell from 100 in the year before to 51 in the latest year.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 191
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 191 adverse event reports that list this exact device (DI 00386270002860), received between Jan 5, 2024 and Sep 24, 2026 — classified as 1 injury, 190 malfunctions.
The most frequently reported specific device problem is “Inappropriate or Unexpected Reset” (83 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Inappropriate or Unexpected Reset83
- Unexpected Shutdown50
- No Audible Alarm19
- Defective Alarm14
- Display or Visual Feedback Problem12
- Wireless Communication Problem10
- Appropriate Device Problem Term/Code Not Available1
- Low Audible Alarm1
Patient problems
- No Clinical Signs, Symptoms or Conditions190
- Hypoglycemia1
Recent adverse event reports
The 25 most recent of 191 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 24, 2026Inappropriate or Unexpected Reset
It was reported that a receiver reinitialization occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50885449 ↗MalfunctionSep 23, 2026Inappropriate or Unexpected Reset
It was reported that a receiver reinitialization occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50872872 ↗MalfunctionSep 23, 2026Wireless Communication Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50862941 ↗MalfunctionAug 25, 2026Inappropriate or Unexpected Reset
It was reported that a receiver reinitialization occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50436469 ↗MalfunctionAug 17, 2026Unexpected Shutdown
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50287599 ↗MalfunctionJul 28, 2026Inappropriate or Unexpected Reset
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50057620 ↗MalfunctionJul 16, 2026Inappropriate or Unexpected Reset
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25845676 ↗MalfunctionJul 13, 2026Wireless Communication Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25805841 ↗MalfunctionJul 7, 2026Defective Alarm
IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25738908 ↗MalfunctionJul 7, 2026Unexpected Shutdown
IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25738342 ↗MalfunctionJun 26, 2026Unexpected Shutdown
IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25636684 ↗MalfunctionMay 21, 2026Inappropriate or Unexpected Reset
IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25247998 ↗MalfunctionMay 7, 2026Inappropriate or Unexpected Reset
IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25117608 ↗MalfunctionMay 6, 2026Inappropriate or Unexpected Reset
IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25096860 ↗MalfunctionApr 29, 2026Inappropriate or Unexpected Reset
IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25034237 ↗MalfunctionApr 29, 2026Inappropriate or Unexpected Reset
IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25031108 ↗MalfunctionApr 22, 2026Wireless Communication Problem
IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED FREQUENTLY BETWEEN (B)(6) 2026. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24955986 ↗MalfunctionApr 13, 2026Defective Alarm
IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24866876 ↗MalfunctionApr 13, 2026Unexpected Shutdown
IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24863124 ↗MalfunctionApr 7, 2026No Audible Alarm
IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. PRODUCT WAS PROVIDED FOR INVESTIGATION. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND FAILED DUE TO USB CONNECTOR DAMAGE. RECEIVER CHARGE AND BOOT TESTS WERE PERFORMED AND FAILED. THE SPEAKER/VIBRATOR RESISTANCE WAS MEASURED AND PASSED. AN INTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. PERFORMANCE DATA WAS REVIEWED. AN ALERT/NOTIFICATION SETTINGS ISSUE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24805656 ↗MalfunctionMar 30, 2026Defective Alarm
IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24709101 ↗MalfunctionMar 26, 2026Inappropriate or Unexpected Reset
IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24702473 ↗MalfunctionMar 16, 2026Display or Visual Feedback Problem
IT WAS REPORTED THAT A DISPLAY ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24611744 ↗MalfunctionMar 9, 2026Inappropriate or Unexpected Reset
IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24551898 ↗MalfunctionMar 2, 2026Inappropriate or Unexpected Reset
IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24484317 ↗
Other Integrated Continuous Glucose Monitoring System, Factory Calibrated devices
- Dexcom G6 Transmitter 9445-24Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated240,475 reports
- Dexcom G6 Transmitter 9445-02Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated3 deaths229,493 reports
- Dexcom G6 Sensor Applicator Assembly 9500-45Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall4 deaths114,678 reports
- US G6 Transmitter (Retail, Receiver Optional) STT-OE-002Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated76,977 reports
- Dexcom G7 GSS (161) 9500-161Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated70,146 reports
- Dexcom G7 iOS App SW12300Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated3 recalls1 death55,219 reports
- Dexcom G7 GSS (161) STP-AT-012Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall5 deaths48,180 reports
- G6 Medicare Transmitter STT-OM-001Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated38,339 reports