McGRATH MACRecalls & reported problems across all models
Covidien LP · 11 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
5 open recalls name at least one McGRATH MAC model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 564
- Deaths reported
- 1
- Injuries reported
- 22
- Recalls
- 5
We found 11 models of McGRATH MAC by Covidien LP in the FDA's device registry (GUDID). Together they are named in 564 adverse event reports in the FDA's MAUDE database, received between Jun 21, 2019 and Sep 25, 2026 — classified as 1 death, 22 injuries, 541 malfunctions.
The most frequently reported specific device problem is “No Display/Image” (438 reports).
5 FDA recall records name at least one of these models by its device identifier, 5 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
McGRATH MAC recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Jul 9, 2024Open, Classified
Battery management system within Laryngoscope devices may deplete below the deign threshold which may result in thermal event followed by risk of explosion.
Z-2636-2024 · Root cause: Device Design · Recall details · FDA record ↗
- Jul 9, 2024Open, Classified
IFU update to address device's battery handling information.
Z-2717-2024 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗
- Sep 28, 2023Open, Classified
Defective video laryngoscopes that did not pass quality tests established for the product, that were designated for destruction, were stolen, and subsequently distributed for sale. Defective device use may result in failed laryngoscopy and/or intubation.
Z-0144-2024 · Root cause: Process control · names 2 of these models · Recall details · FDA record ↗
- Aug 10, 2023Open, Classified
Laryngoscope blades may contain an ineffective application of anti-fog agent which may result in condensation accumulating causing the screen to be blurred or obscured.
Z-0169-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Aug 10, 2023Open, Classified
Laryngoscope blades may contain an ineffective application of anti-fog agent which may result in condensation accumulating causing the screen to be blurred or obscured.
Z-0168-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported McGRATH MAC problems
Device problems
- No Display/Image438
- Electrical /Electronic Property Problem65
- Material Opacification31
- Failure to Power Up29
- Erratic or Intermittent Display20
- Failure to Run on Battery12
- Mechanical Problem11
- Device Damaged Prior to Use6
Patient problems
- No Clinical Signs, Symptoms or Conditions455
- Insufficient Information94
- Unspecified Tissue Injury14
- Foreign Body In Patient5
- Extubate3
- No Consequences Or Impact To Patient3
- Cardiac Arrest2
- Hemorrhage/Blood Loss/Bleeding2
McGRATH MAC models
11 models, most reported first. Each model has its own page with its reports, recalls and device details.
- McGRATH MAC 301-000-000LARYNGOSCOPE, RIGID2 recalls202 reports
- McGRATH MAC 300-000-000LARYNGOSCOPE, RIGID1 recall1 death205 reports
- McGRATH MAC 301-000-000LARYNGOSCOPE, RIGID2 recalls90 reports
- McGRATH MAC 350-017-000LARYNGOSCOPE, RIGID1 recall3 reports
- McGRATH MAC 350-084-000LARYNGOSCOPE, RIGID1 recall0 reports
- McGRATH MAC 350-013-000LARYNGOSCOPE, RIGID20 reports
- McGRATH MAC 350-005-000LARYNGOSCOPE, RIGID19 reports
- McGRATH MAC 350-005-000LARYNGOSCOPE, RIGID15 reports
- McGRATH MAC 350-017-000LARYNGOSCOPE, RIGID5 reports
- McGRATH MAC 350-013-000LARYNGOSCOPE, RIGID4 reports
- McGRATH MAC 350-086-000LARYNGOSCOPE, RIGID1 report
Recent McGRATH MAC adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 25, 2026No Display/Image
Patient problems: Insufficient Information
MAUDE report 50906171 ↗MalfunctionAug 6, 2026No Display/Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50168435 ↗MalfunctionJul 8, 2026No Display/Image
IT WAS REPORTED THAT DURING TESTING OR MAINTENANCE OF THE VIDEO LARYNGOSCOPE, THE LED LIGHT TURNED ON BUT THE SCREEN STAYED BLACK WITH NO DISPLAY OUTPUT. AFTER A NUMBER OF MONTHS OF NON-USE, WHEN THE DEVICE WAS INITIALLY POWERED ON, THE PICTURE WAS REPORTED AS VERY DULL WITH VERTICAL LINES RUNNING THROUGH THE DISPLAY IMAGE. ON THE NEXT USE, THERE WAS NO PICTURE AT ALL, DESCRIBED AS A COMPLETE LOSS OF IMAGE. WHEN AN IMAGE HAD BEEN PRESENT, THE BATTERY REPORTEDLY DISPLAYED ¿189¿ IN THE CORNER. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25747362 ↗MalfunctionJun 26, 2026No Display/Image
IT WAS REPORTED THAT DURING PRE-USAGE INSPECTION, THE VIDEO LARYNGOSCOPE HAD A FAULT DESCRIBED AS NO IMAGE ON SCREEN. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25641350 ↗InjuryMay 5, 2026Fitting Problem
IT WAS REPORTED THAT THE VIDEOLARYNGOSCOPE WITH SIZE 3 BLADE WAS USED AND THE PATIENT WAS SUCCESSFULLY INTUBATED IN THE FIELD . WHEN THE HOSPITAL EXTUBATED THE PATIENT AFTER 5 DAYS IT WAS DISCOVERED THAT THE BLADE WAS STILL IN THE PATIENT'S THROAT. IT SEEMS THAT THE BLADE DID NOT LATCH ONTO THE HANDLE AS EXPECTED AND CAME OFF THE HANDLE DURING INTUBATION. IT WAS MADE SURE THAT THE BLADE WAS STILL INTACT AND NOT MISSING ANY PIECES WHEN REMOVED. THE PATIENT WAS ALIVE BUT AN IDENTIFIED INJURY WAS REPORTED.
Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient; Unspecified Tissue Injury
MAUDE report 25088451 ↗InjuryMay 5, 2026Fitting Problem
IT WAS REPORTED THAT THE PATIENT WAS SUCCESSFULLY INTUBATED AND WHEN THE CREW WENT TO APPLY THE TUBE TAMER, IT WAS NOTICED THAT THE SIZE 3 BLADE HAD COME OFF THE HANDLE AND WAS STILL IN THE PATIENT'S THROAT. THE CREW REMOVED THE BLADE AND THEN APPLIED THE TUBE TAMER. A COUPLE OF DIFFERENT BLADES WAS TESTED TO THE HANDLE, AND IT LOOKS LIKE THE SIZE 3 BLADE WOULD NOT LATCH ONTO THE HANDLE CORRECTLY. THE OTHER SIZE BLADES LATCH ONTO THE HANDLE FINE. THE PATIENT WAS ALIVE BUT AN IDENTIFIED INJURY WAS REPORTED.
Patient problems: Foreign Body In Patient; Unspecified Tissue Injury
MAUDE report 25088445 ↗MalfunctionApr 30, 2026No Display/Image; Failure to Power Up
IT WAS REPORTED THAT A HEALTHCARE PROFESSIONAL USED THE VIDEO LARYNGOSCOPE DURING A DIFFICULT INTUBATION AND, WHEN THE HANDLE WAS TU RNED ON, A BLUE SCREEN DISPLAYED AND IT DID NOT PROGRESS TO THE VIDEO SCREEN. AFTER THE BATTERY WAS CHANGED, THE VIDEO SCREEN APPEARED BUT THEN WENT BLACK DURING THE INTUBATION. IT WAS FURTHER REPORTED THAT THE HANDLE WOULD NOT TURN ON. THE HANDLE WAS TESTED WITH KNOWN GOOD BATTERIES, BUT THE ISSUE WAS NOT RESOLVED. THE ANESTHETIST WAS ABLE TO INTUBATE WITHOUT IT IN THE END AS THE SCREEN WAS NOT RELIABLE. THERE WAS NO PATIENT INJURY.
Patient problems: Insufficient Information
MAUDE report 25043599 ↗MalfunctionApr 28, 2026No Display/Image
IT WAS REPORTED THAT THE LIGHT-EMITTING DIODE (LED) OF THE VIDEOLARYNGOSCOPE WAS ON BUT HAD NO DISPLAY EVEN THE BATTERY ICON WAS NOT DISPLAYED PRIOR TO USE. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25017250 ↗MalfunctionApr 27, 2026No Display/Image
IT WAS REPORTED THAT THE VIDEOLARYNGOSCOPE HAD NO DISPLAY, NOT EVEN THE BATTERY ICON DESPITE REPLACING THE BATTERY IN A HOSPITAL SETTING. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25009564 ↗MalfunctionApr 27, 2026Optical Problem
INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, HEALTHCARE PROVIDER) REGARDING A DEVICE USED FOR SPINAL THERAPY. IT WAS REPORTED THAT THE INSTRUMENT CENTER WOBBLES. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT. ADDITIONAL INFORMATION RECEIVED THAT IT WAS STATED THAT THE FOCUS COULD NOT BE PROPERLY ADJUSTED. PROCEDURE/THERAPY INVOLVED WAS MICROENDOSCOPIC DISCECTOMY FOR HERNIA TO L4/5. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24994998 ↗MalfunctionApr 24, 2026No Display/Image
IT WAS REPORTED THAT THE VIDEO LARYNGOSCOPE'S SCREEN DID NOT WORKED AND HAD A BLACK SCREEN. IT WAS UNKNOWN IF THERE WAS PATIENT INVOLVEMENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: Insufficient Information
MAUDE report 24987758 ↗MalfunctionApr 22, 2026No Display/Image
IT WAS REPORTED THAT THE VIDEOLARYNGOSCOPE TURNS ON, BUT THERE WAS NO DISPLAY AND THE DEVICE DID NOT FUNCTION PROPERLY. THE BATTERY WAS REPLACED, BUT THE ISSUE DID NOT IMPROVED PRIOR TO USE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24954109 ↗MalfunctionApr 20, 2026Material Opacification
IT WAS REPORTED THAT A VIDEOLARYNGOSCOPE BLADE HAD A BLURRED LENS PRIOR TO USE. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24927828 ↗MalfunctionApr 20, 2026Material Opacification
IT WAS REPORTED THAT A VIDEOLARYNGOSCOPE BLADE HAD A BLURRED LENS PRIOR TO USE. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24927790 ↗MalfunctionApr 6, 2026No Display/Image
IT WAS REPORTED THAT, WHILE PREPARING FOR USE, THE CUSTOMER ATTEMPTED TO POWER ON THE VIDEOLARYNGOSCOPE AND THE SCREEN STARTED UP BUT THEN WENT BLACK. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24794623 ↗
Other Covidien LP products
- Endo GIA48 models · 26,277 reports · 8 recalls
- Endo GIA Ultra3 models · 15,650 reports
- Shiley387 models · 9,392 reports · 5 recalls
- Signia16 models · 8,201 reports · 6 recalls
- LigaSure32 models · 7,283 reports · 1 recall
- Tri-Staple 2.037 models · 4,892 reports · 2 recalls
- Polysorb291 models · 4,564 reports · 2 recalls
- EEA39 models · 4,432 reports · 1 recall
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page