Skip to content

Omnipod 5 PodRecalls & reported problems across all models

Insulet Corporation · 3 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name at least one Omnipod 5 Pod model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
133,848
Deaths reported
86
Injuries reported
13,672
Recalls
2

We found 3 models of Omnipod 5 Pod by Insulet Corporation in the FDA's device registry (GUDID). Together they are named in 133,848 adverse event reports in the FDA's MAUDE database, received between Mar 8, 2022 and Sep 30, 2026 — classified as 86 deaths, 13,672 injuries, 120,087 malfunctions.

The most frequently reported specific device problem is “Unintended Movement” (64,009 reports).

2 FDA recall records name at least one of these models by its device identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 2,002 reportsDec 2023: 1,662 reportsJan 2024: 2,356 reportsFeb 2024: 716 reportsMar 2024: 2,502 reportsApr 2024: 3,284 reportsMay 2024: 2,869 reportsJun 2024: 3,012 reportsJul 2024: 3,126 reportsAug 2024: 3,156 reportsSep 2024: 3,405 reportsOct 2024: 3,428 reportsNov 2024: 2,794 reportsDec 2024: 1,594 reportsJan 2025: 3,290 reportsFeb 2025: 2,454 reportsMar 2025: 3,624 reportsApr 2025: 2,502 reportsMay 2025: 3,481 reportsJun 2025: 4,295 reportsJul 2025: 4,117 reportsAug 2025: 3,746 reportsSep 2025: 3,494 reportsOct 2025: 4,281 reportsNov 2025: 5,358 reportsDec 2025: 4,781 reportsJan 2026: 3,885 reportsFeb 2026: 5,004 reportsMar 2026: 5,011 reportsApr 2026: 6,159 reportsMay 2026: 5,253 reportsJun 2026: 4,614 reportsJul 2026: 1,914 reportsAug 2026: 1,306 reportsSep 2026: 1,091 reportsOct 2026: 0 reports2024202520266,159
Last 36 months. The most recent month may be incomplete.

Omnipod 5 Pod recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. May 20, 2026Open, Classified

    External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.

    Z-2603-2026 · Root cause: Process design · names 3 of these models · Recall details · FDA record ↗

  2. Mar 12, 2026Open, Classified

    Due to a manufacturing defect, certain Omnipod 5 Pods from 49 lots have an internal soft cannula tear that results in insulin leaking into the Pod instead of being delivered to the user regardless of basal or bolus delivery. This defect results from damage to the unexposed portion of the soft cannula during manufacturing, which would result in a compromised fluid path. The primary failure mode is pump under-delivery due to loss of insulin to an internal leak; in some cases, the defect may also lead to pump shutoff and cessation of insulin delivery when leaked insulin contacts Pod circuity in a manner that results in an electrical short. Under-delivery of insulin (both basal and bolus insulin) or cessation of insulin put users at risk of hyperglycemia, and complications from acute and chronic hyperglycemia, including dehydration, blurry vision, nausea, vomiting, altered mental status, diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death. Users may require hospitalization or medical intervention to treat severe adverse health consequences. Not all devices with the defect will issue an alarm or alert the user. If there is sufficient leakage of insulin to cause a short in the circuity, the Pod will issue a Hazard Alarm that stops all insulin delivery and alerts the user to replace their Pod. In addition, if a user s glucose is trending high and is not responding to insulin delivery, the system may reach the maximum amount of insulin microboluses allowed by the system and trigger the Automated Delivery Restriction (ADR) alert that tells users to check their blood glucose and take appropriate actions (i.e., ADR is a response to persistent hyperglycemia and maximum automated delivery constraints rather than a direct detection of the leak). The magnitude of under-delivery is unknown and based on multiple factors, including how much insulin is being delivered, whether an alarm and/or alert triggers, whether and when the user recognizes the device defect, the duration of Pod use, and the size of the tear.

    Z-1797-2026 · Root cause: Process design · Recall details · FDA record ↗

Most reported Omnipod 5 Pod problems

Device problems

  • Unintended Movement64,009
  • Loss of or Failure to Bond57,949
  • Material Twisted/Bent23,939
  • Insufficient Device Problem Information13,727
  • Fluid/Blood Leak13,647
  • Failure to Fire13,618
  • Difficult to Insert10,223
  • Activation, Positioning or Separation Problem6,043

Patient problems

  • Hyperglycemia104,265
  • No Clinical Signs, Symptoms or Conditions23,829
  • Elevated Ketones/ Diabetic Ketoacidosis4,277
  • Erythema3,393
  • Vomiting3,068
  • Pain2,812
  • Hemorrhage/Blood Loss/Bleeding2,666
  • Swelling/ Edema2,230

Omnipod 5 Pod models

3 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Omnipod 5 Pod adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Insulet Corporation products

All Insulet Corporation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page