Omnipod 5 Pod POD-OMNI-I1-6720Recalls & reported problems

Insulet Corporation · Omnipod 5 Pod · Class 2

All models of Omnipod 5 (8) · Omnipod 5 Pod (3)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This is an Omnipod 5 pod. Its reports mostly describe pods coming loose or leaking and cannulas that do not deploy. There are 63,471 reports, including 9 deaths and 4,206 injuries. Two recalls are open: cannulas damaged during manufacturing can leak insulin around the pod, and certain pods have cannula tears from a manufacturing problem. Reports rose to 38,902 in the last 12 months, from 23,754 before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
60,668
Deaths reported
11
Injuries reported
4,386
Recalls naming it
2

Summary

The FDA's MAUDE database contains 60,668 adverse event reports that list this exact device (DI 10385083000527), received between Mar 7, 2024 and Sep 30, 2026 — classified as 11 deaths, 4,386 injuries, 56,267 malfunctions.

The most frequently reported specific device problem is “Unintended Movement” (30,476 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 7 reportsApr 2024: 6 reportsMay 2024: 14 reportsJun 2024: 8 reportsJul 2024: 19 reportsAug 2024: 82 reportsSep 2024: 165 reportsOct 2024: 514 reportsNov 2024: 935 reportsDec 2024: 710 reportsJan 2025: 1,765 reportsFeb 2025: 1,340 reportsMar 2025: 2,052 reportsApr 2025: 1,633 reportsMay 2025: 2,093 reportsJun 2025: 2,726 reportsJul 2025: 2,452 reportsAug 2025: 2,636 reportsSep 2025: 2,370 reportsOct 2025: 3,042 reportsNov 2025: 4,232 reportsDec 2025: 3,912 reportsJan 2026: 3,255 reportsFeb 2026: 4,103 reportsMar 2026: 4,302 reportsApr 2026: 5,316 reportsMay 2026: 4,311 reportsJun 2026: 3,695 reportsJul 2026: 1,438 reportsAug 2026: 1,101 reportsSep 2026: 434 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520265,316
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 20, 2026Class IOpen, Classified

    External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.

    Z-2603-2026 · Root cause: Process design · Recall details · FDA record ↗

  2. Mar 12, 2026Class IOpen, Classified

    Due to a manufacturing defect, certain Omnipod 5 Pods from 49 lots have an internal soft cannula tear that results in insulin leaking into the Pod instead of being delivered to the user regardless of basal or bolus delivery. This defect results from damage to the unexposed portion of the soft cannula during manufacturing, which would result in a compromised fluid path. The primary failure mode is pump under-delivery due to loss of insulin to an internal leak; in some cases, the defect may also lead to pump shutoff and cessation of insulin delivery when leaked insulin contacts Pod circuity in a manner that results in an electrical short. Under-delivery of insulin (both basal and bolus insulin) or cessation of insulin put users at risk of hyperglycemia, and complications from acute and chronic hyperglycemia, including dehydration, blurry vision, nausea, vomiting, altered mental status, diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death. Users may require hospitalization or medical intervention to treat severe adverse health consequences. Not all devices with the defect will issue an alarm or alert the user. If there is sufficient leakage of insulin to cause a short in the circuity, the Pod will issue a Hazard Alarm that stops all insulin delivery and alerts the user to replace their Pod. In addition, if a user s glucose is trending high and is not responding to insulin delivery, the system may reach the maximum amount of insulin microboluses allowed by the system and trigger the Automated Delivery Restriction (ADR) alert that tells users to check their blood glucose and take appropriate actions (i.e., ADR is a response to persistent hyperglycemia and maximum automated delivery constraints rather than a direct detection of the leak). The magnitude of under-delivery is unknown and based on multiple factors, including how much insulin is being delivered, whether an alarm and/or alert triggers, whether and when the user recognizes the device defect, the duration of Pod use, and the size of the tear.

    Z-1797-2026 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Unintended Movement30,476
  • Loss of or Failure to Bond28,215
  • Material Twisted/Bent10,986
  • Fluid/Blood Leak7,216
  • Failure to Fire5,569
  • Insufficient Device Problem Information5,340
  • Difficult to Insert5,095
  • Activation, Positioning or Separation Problem2,661

Patient problems

  • Hyperglycemia49,122
  • No Clinical Signs, Symptoms or Conditions10,011
  • Erythema1,859
  • Hemorrhage/Blood Loss/Bleeding1,560
  • Elevated Ketones/ Diabetic Ketoacidosis1,446
  • Pain1,297
  • Vomiting1,273
  • Swelling/ Edema1,085

Recent adverse event reports

The 25 most recent of 60,668 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Loss of or Failure to Bond; Insufficient Device Problem Information

    It was reported by the patient's legal guardian that, after approximately 37-48 hours of Pod wear, the Pod's adhesive failed to adhere to the skin. Reportedly, the adhesion issue was attributed to the patient's prescribed skin cream used to treat eczema. It was not specified whether the eczema developed as a result of Pod wear, and no further details regarding the healthcare professional consultation or the prescribed treatment were reported. Insulet is in contact with the initial reporter to obtain clarification regarding the reported event. Should additional information become available, a supplemental report will be submitted.

    Patient problems: Eczema

    MAUDE report 50955170 ↗
  • InjurySep 30, 2026Insufficient Device Problem Information

    It was reported that, on (b)(6) 2026, the patient was hospitalized with hyperglycemia. Reportedly, the patient experienced persistent hyperglycemia while wearing two different Pods. While wearing the first Pod, blood glucose reportedly increased to approximately 26.2 mmol/L (~ 472 mg/dL). An ambulance was called and, prior to its arrival, the first Pod was removed and replaced at home (documented and reported under internal Insulet reference number (b)(4)). While wearing the second Pod, blood glucose was reportedly measured at 21.5 mmol/L (~387 mg/dL), with ketone levels of 4.1 mmol/L upon evaluation in the emergency department. Reported symptoms included headache, fatigue, malaise, and difficulty staying awake. The patient was admitted overnight to (b)(6) Hospital, where they received intravenous fluids and insulin and were monitored at regular intervals before a new Pod was applied. The patient was discharged approximately 19 hours later. The second Pod was removed and discarded at the hospital.

    Patient problems: Extreme Exhaustion; Headache; Hyperglycemia; Malaise

    MAUDE report 50955140 ↗
  • InjurySep 30, 2026Insufficient Device Problem Information

    It was reported that, on (b)(6) 2026, after approximately 1-4 hours of Pod wear on the arm, the patient was hospitalized with diabetic ketoacidosis (DKA) at Hôpital Pourtalès. It was further reported that the patient experienced persistent hyperglycemia, with blood glucose increasing to greater than 21 mmol/L (378 mg/dL) following breakfast and a bolus dose, with readings displayed as ¿HIGH¿ on the Dexcom G7 continuous glucose monitoring system. Reportedly, blood glucose did not decrease despite multiple bolus doses totaling approximately 16 units of insulin. The Omnipod system prompted a switch from Automated Mode to Manual Mode due to an insulin restriction message, indicating that the maximum deliverable dose had been reached. The patient subsequently replaced the Pod and delivered an additional 6 units of insulin using the new Pod; however, blood glucose continued to display as ¿HIGH¿ after two hours. The patient developed symptoms including urinary frequency, polydipsia, nausea, and vomiting, prompting them to seek medical attention. The patient was subsequently hospitalized with a diagnosis of severe hyperglycemia and DKA. Upon admission, blood glucose measured 34 mmol/L (612

    Patient problems: Elevated Ketones/ Diabetic Ketoacidosis; Hyperglycemia; Nausea; Polydipsia; Urinary Frequency / Polyuria; Vomiting

    MAUDE report 50955137 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Fluid/Blood Leak; Material Twisted/Bent

    Patient problems: Hyperglycemia

    MAUDE report 50954897 ↗
  • MalfunctionSep 30, 2026Retraction Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954819 ↗
  • InjurySep 30, 2026Insufficient Device Problem Information

    It was reported that the patient developed an infection at the Pod¿s insertion site (abdomen) while wearing the device longer than 48 hours. Patient reported the infusion site to have skin irritation, redness, and pain. Due to worsening symptoms, the patient sought medical attention from a physician on (b)(6), 2026. Following evaluation, the patient was diagnosed with an infection at the Pod site, which the patient believed was related to the Pod being worn on the abdomen. As treatment, the physician prescribed an antibiotic; however, the patient was unable to recall the name of the medication. The medical visit was completed on the same day. The faulty pod was discarded.

    Patient problems: Erythema; Pain; Skin Inflammation/ Irritation; Unspecified Infection

    MAUDE report 50954578 ↗
  • MalfunctionSep 30, 2026Fluid/Blood Leak; Unintended Movement

    Patient problems: Hyperglycemia

    MAUDE report 50954567 ↗
  • MalfunctionSep 30, 2026Material Twisted/Bent

    Patient problems: Hyperglycemia

    MAUDE report 50954564 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Fluid/Blood Leak; Material Twisted/Bent; Unintended Movement; Insufficient Device Problem Information

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Hyperglycemia

    MAUDE report 50953987 ↗
  • InjurySep 30, 2026Insufficient Device Problem Information

    It was reported that the patient went to the emergency room for 6 hours on (b)(6), 2026 due to hypoglycemia. The patient's blood glucose levels declined to 52 mg/dL while wearing the Pod between 4 and 24 hours. Symptoms reported include loss of consciousness. The patient was treated with ¿sugar bags¿. The Pod was removed in the emergency room. The patient alleged that the Pod was receiving incorrect blood glucose readings, which led to the Pod delivering more insulin than was needed.

    Patient problems: Hypoglycemia; Loss of consciousness

    MAUDE report 50953682 ↗
  • MalfunctionSep 30, 2026Failure to Fire

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50953307 ↗
  • MalfunctionSep 30, 2026Failure to Fire

    Patient problems: Hyperglycemia

    MAUDE report 50952779 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Fluid/Blood Leak; Unintended Movement

    Patient problems: Hyperglycemia

    MAUDE report 50952565 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Fluid/Blood Leak; Unintended Movement

    Patient problems: Hyperglycemia

    MAUDE report 50952440 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Unintended Movement

    Patient problems: Hyperglycemia

    MAUDE report 50952421 ↗
  • InjurySep 30, 2026Fluid/Blood Leak; Insufficient Device Problem Information

    The patient¿s mother reported that a skin irritation causing redness at the Pod site on the patient¿s leg had occurred while wearing the Pod longer than 48 hours. The patient¿s blood glucose level rose over 400 mg/dL. The patient mother also stated they could smell insulin, which indicates leaking during wear. As treatment, the patient was prescribed an unspecified cream.

    Patient problems: Skin Inflammation/ Irritation

    MAUDE report 50951927 ↗
  • InjurySep 30, 2026Fluid/Blood Leak

    It was reported that the patient had been hospitalized with Diabetic Ketoacidosis (DKA) on (b)(6) 2026. The patient's blood glucose levels rose over 600 mg/dL while wearing the Pod between 4 and 24 hours on the leg. The patient stated the Pod was leaking while they were asleep, leaving 0.5 units of insulin in their Pod. The patient was taken to the hospital via ambulance. Symptoms reported include thirst, shaking, headache and abdominal pain. The patient also mentioned they were septic. The patient was treated with intravenous fluids, insulin drip, unspecified pain and nausea medication, electrolytes, magnesium, phosphates, potassium and unspecified antibiotics for kidney infection. The Pod was removed and discarded. The patient was released the four days later on (b)(6) 2026. A new Pod was applied after the patient left the hospital.

    Patient problems: Abdominal Cramps; Elevated Ketones/ Diabetic Ketoacidosis; Headache; Hyperglycemia; Sepsis

    MAUDE report 50951764 ↗
  • InjurySep 30, 2026Loss of or Failure to Bond; Unintended Movement

    It was reported that the patient went to the emergency room for hyperglycemia on (b)(6)2026. The patient's blood glucose levels reached around 400 mg/dL while wearing the Pod between 1 and 4 hours. The Pod reportedly fell off of the infusion site (abdomen) while sleeping, and the patient's blood glucose levels did not drop despite applying a new Pod and administering boluses. Symptoms reported include pain throughout their body. The patient was treated with IV fluids, pain medication, sodium chloride, Famotidine and Ondansetron. The patient was released after about an hour and a half.

    Patient problems: Hyperglycemia; Pain

    MAUDE report 50948372 ↗
  • MalfunctionSep 30, 2026Thermal Decomposition of Device

    Investigation of the returned device confirmed corrosion with undetermined cause. During the fill process, the Pod began emitting a continuous alarm. Upon inspection, the interior of the Pod appeared brown and damaged. The complaint was originally coded as Pod - Corroded/Discolored and POD Hazard Alarm/Alert/Advisory - During Operation.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946816 ↗
  • MalfunctionSep 29, 2026Insufficient Device Problem Information

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Hyperglycemia; Scar Tissue; Skin Inflammation/ Irritation

    MAUDE report 50941332 ↗
  • InjurySep 29, 2026Insufficient Device Problem Information

    It was reported that the patient sought medical assistance with hyperglycemia and diabetic ketoacidosis (DKA) on (b)(6), 2026. The patient's blood glucose levels reached over 350 mg/dL while wearing the Pod between 36 and 48 hours. Symptoms reported include loss of consciousness, diaphoresis, and not being able to communicate properly. The patient had a CT scan, x-ray, and other general tests performed. Blood glucose was treated with insulin. The Pod was removed at the hospital and discarded. The patient was released the same day. The patient alleged that there had been a difference between sensor values in the Omnipod 5/Dexcom app versus real time values when checked manually.

    Patient problems: Alteration in Body Temperature; Elevated Ketones/ Diabetic Ketoacidosis; Hyperglycemia; Loss of consciousness; Speech Disorder

    MAUDE report 50941274 ↗
  • InjurySep 29, 2026Excess Flow or Over-Infusion

    It was reported that the patient activated a Pod at 3:15 PM and later that evening went to the emergency room with hypoglycemia on (b)(6) 2026. The patient's blood glucose levels declined to 27 mg/dL while wearing the Pod between 4 and 24 hours. Symptoms reported include loss of consciousness. The patient was treated with intravenous glucose. However, blood glucose remained critically low for several hours, until 1 hour after hospital staff removed the Pod. Blood glucose stabilized at 233 mg/dL. The patient was released 8 hours later on the following day ((b)(6) 2026). The Pod was removed in the emergency room and discarded. The patient alleged that they paused insulin delivery, but was still receiving insulin. It was also reported that the patient did not use the Omnipod app, but rather a handheld controller with Lispro insulin. During the call with Product Support, it was advised that while the Pod is in manual mode, the Pod cannot automatically suspend insulin delivery when blood glucose levels are running low (unlike automated mode). The Pod is not available for return.

    Patient problems: Hypoglycemia; Loss of consciousness

    MAUDE report 50941054 ↗
  • MalfunctionSep 29, 2026Difficult to Insert

    Patient problems: Hyperglycemia

    MAUDE report 50940940 ↗
  • InjurySep 29, 2026Insufficient Device Problem Information

    It was reported that the patient went to ER on Sep 2nd, 2026 for Diabetic Ketoacidosis (DKA) because of the POD malfunctioned as it was not delivering enough insulin. The patient's blood glucose (BG) level rose over 600 mg/dL while wearing the Pod between 36 and 48 hours on infusion site (abdomen). As treatment, intravenous insulin and fluids were given. The patient was discharged on same day.

    Patient problems: Hyperglycemia

    MAUDE report 50940926 ↗
  • MalfunctionSep 29, 2026Loss of or Failure to Bond; Unintended Movement

    Patient problems: Hyperglycemia

    MAUDE report 50940911 ↗

FDA clearance and classification

The FDA regulates product code QFG (alternate controller enabled insulin infusion pump) under 21 CFR 880.5730 as a Class II device. 21 CFR 880.5730 ↗

Devices of this type normally need 510(k) clearance before sale.

FDA definition of this device type: An alternate controller enabled insulin infusion pump (ACE insulin pump) is a device intended for the infusion of insulin into a patient. The ACE insulin pump may include basal and bolus drug delivery at set or variable rates. ACE insulin pumps are designed to reliably and securely communicate with external devices, such as automated insulin dosing systems, to allow insulin delivery commands to be received, executed, and confirmed. ACE insulin pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of insulin delivery.

Common questions about the Omnipod 5 Pod POD-OMNI-I1-6720

Has the Omnipod 5 Pod POD-OMNI-I1-6720 been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-2603-2026 (Class I), began on May 20, 2026: External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Omnipod 5 Pod POD-OMNI-I1-6720?

The FDA's MAUDE database lists 60,668 adverse event reports for this device, received between Mar 7, 2024 and Sep 30, 2026. In those reports, the most common specific device problems are “Unintended Movement” (30,476), “Loss of or Failure to Bond” (28,215) and “Material Twisted/Bent” (10,986); the most common patient problems are “Hyperglycemia” (49,122), “Erythema” (1,859) and “Hemorrhage/Blood Loss/Bleeding” (1,560). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Omnipod 5 Pod POD-OMNI-I1-6720?

11 reports are classified as deaths, 4,386 as injuries and 56,267 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Omnipod 5 Pod POD-OMNI-I1-6720 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the Omnipod 5 Pod POD-OMNI-I1-6720, and is it still sold?

The FDA's device database lists Insulet Corporation as the company responsible for it (brand Omnipod 5 Pod, model POD-OMNI-I1-6720). Its record shows it as in commercial distribution.