Omnipod 5 Pod POD-BLE-H1-520Recalls & reported problems
Insulet Corporation · Omnipod 5 Pod · Class 2
All models of Omnipod 5 (8) · Omnipod 5 Pod (3)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This is an Insulet Omnipod 5 pod. Its reports mostly describe needles or cannulas that fail to insert, pods falling off or leaking, and pods that fail to connect. There are 71,566 reports, including 80 deaths and 9,144 injuries. One recall, started in May 2026, is open: a cannula damaged during manufacturing can let insulin leak around the pod. Reports dropped to 6,931 in the last 12 months, from 15,650 before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 71,491
- Deaths reported
- 74
- Injuries reported
- 9,398
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 71,491 adverse event reports that list this exact device (DI 10385083000114), received between Mar 8, 2022 and Sep 30, 2026 — classified as 74 deaths, 9,398 injuries, 62,019 malfunctions.
The most frequently reported specific device problem is “Unintended Movement” (32,747 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- May 20, 2026Class IOpen, Classified
External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.
Z-2603-2026 · Root cause: Process design · Recall details · FDA record ↗
Most reported problems
Device problems
- Unintended Movement32,747
- Loss of or Failure to Bond29,011
- Material Twisted/Bent12,573
- Insufficient Device Problem Information8,385
- Failure to Fire7,831
- Fluid/Blood Leak6,230
- Difficult to Insert4,940
- Activation, Positioning or Separation Problem3,260
Patient problems
- Hyperglycemia53,830
- No Clinical Signs, Symptoms or Conditions13,380
- Elevated Ketones/ Diabetic Ketoacidosis2,884
- Vomiting1,817
- Erythema1,544
- Hypoglycemia1,540
- Pain1,502
- Skin Inflammation/ Irritation1,372
Recent adverse event reports
The 25 most recent of 71,491 reports, in the reporter's own words as published by the FDA.
InjurySep 30, 2026Insufficient Device Problem Information
It was reported by the patient¿s legal guardian that, while traveling in an unspecified country, the patient developed a skin infection at the Pod site on the back. The duration of Pod wear was not specified. The legal guardian stated that the patient was prescribed an antibiotic cream, described as a 0.5% gel similar in consistency to Vaseline, which reportedly resolved the infection after ten days of treatment. The legal guardian declined to provide additional information. No further details are available, including the name of the healthcare facility where the prescription was issued.
Patient problems: Skin Infection
MAUDE report 50955161 ↗InjurySep 30, 2026Insufficient Device Problem Information
It was reported by the patient¿s legal guardian that, while traveling in an unspecified country, the patient developed a skin infection at the Pod site on the leg. The duration of Pod wear was not specified. The legal guardian stated that the patient was prescribed an antibiotic cream, described as a 0.5% gel similar in consistency to Vaseline, which reportedly resolved the infection after ten days of treatment. The legal guardian declined to provide additional information. No further details are available, including the name of the healthcare facility where the prescription was issued.
Patient problems: Skin Infection
MAUDE report 50955159 ↗InjurySep 30, 2026Insufficient Device Problem Information
It was reported that the patient experienced a skin infection at the Pod site on the leg, which developed after approximately 5-24 hours of Pod wear. Reportedly, the site became red, swollen, and painful, prompting the patient to seek medical attention. On (b)(6), 2026, the patient consulted a healthcare professional and was prescribed antibiotic treatment (Amoxicillin). The medical visit lasted approximately 30 minutes. The Pod involved was discarded and is unavailable for investigation.
Patient problems: Erythema; Pain; Skin Infection; Swelling/ Edema
MAUDE report 50955100 ↗InjurySep 30, 2026Insufficient Device Problem Information
It was reported that the patient was hospitalized with Hyperglycemia on an unspecified date. The customer reported a three-day hospital stay, but the discharge date was unavailable. The patient's blood glucose levels were not reported. The duration of pod wear was not provided.\ Symptoms experienced prior to seeking medical attention were reported as a sensor discrepancy; however, further details were unavailable due to the patient's inability to recall specific information. The patient stated that the diagnosis received was diabetes. Treatment provided by healthcare professionals reportedly included administration of large amounts of insulin.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954680 ↗InjurySep 30, 2026Insufficient Device Problem Information
It was reported that the patient developed scarring at the Pod¿s insertion site. The patient¿s blood glucose rose by seventeen points; while wearing the Pod for an unknown duration; however, no blood glucose values were reported. The patient declined to provide additional information regarding medical treatment, or patient outcomes.
Patient problems: Scar Tissue
MAUDE report 50953647 ↗InjurySep 30, 2026Insufficient Device Problem Information
The patient reported that the Pods were scarring their skin with its adhesive. No impact to the patient's glucose levels were reported. The patient declined to provide additional information regarding this incident.
Patient problems: Scar Tissue
MAUDE report 50953562 ↗InjurySep 30, 2026Insufficient Device Problem Information
Caregiver reported that a skin irritation causing scar tissue occurred while wearing the Pod for an unspecified duration of use at the Pod site. The caregiver mentioned the scar tissue is due to wearing the Pod repeatedly. The patient¿s blood glucose level rose over 400 mg/dL. As treatment, the patient applied a new Pod.
Patient problems: Scar Tissue; Skin Inflammation/ Irritation
MAUDE report 50952166 ↗InjurySep 29, 2026Insufficient Device Problem Information
It was reported that the patient went to the hospital for Diabetic Ketoacidosis (DKA) because of the POD malfunctioned as it was not delivering enough insulin. The patient's blood glucose (BG) level was unknown. There were no further information available and multiple good-faith efforts to obtain additional event details were unsuccessful.
Patient problems: Elevated Ketones/ Diabetic Ketoacidosis
MAUDE report 50940915 ↗InjurySep 29, 2026Insufficient Device Problem Information
It was reported that the patient went to the hospital for Diabetic Ketoacidosis (DKA) because of the POD malfunctioned as it was not delivering enough insulin. The patient's blood glucose (BG) level was unknown. There were no further information available and multiple good-faith efforts to obtain additional event details were unsuccessful.
Patient problems: Elevated Ketones/ Diabetic Ketoacidosis
MAUDE report 50940912 ↗InjurySep 29, 2026Insufficient Device Problem Information
It was reported that a skin irritation causing itchiness, redness, pain, and burning sensation had occurred while wearing the Pod for an unspecified period. Incident occurred about 2 weeks prior, but the exact date could not be provided. Length of visit was between 2 PM and 7 PM. They were discharged the same day. The patient reported that when removing the Pod, it also removed some of their skin. They normally use oil, Vaseline, and warm water to remove pods. However with this particular Pod, the patient just took it off. As treatment, the patient was prescribed medication for the pain along with silver nitrate cream.
Patient problems: Burning Sensation; Erythema; Itching Sensation; Pain
MAUDE report 50940498 ↗InjurySep 29, 2026Insufficient Device Problem Information
It was reported that the patient was hospitalized with hypoglycemia on an unspecified date. The patient's blood glucose levels declined to 1.8 mmol/L (32 mg/dL) while wearing the Pod. The patient's diabetes specialist nurse provided additional information that the patient reportedly consumed 20 grams of carbohydrates. No alcohol consumption was involved and no correction bolus was administered in relation to the event. No further information was provided. Additional information is being solicited, a supplemental report will be submitted if received.
Patient problems: Hypoglycemia
MAUDE report 50939770 ↗InjurySep 29, 2026Loss of or Failure to Bond; Unintended Movement
It was reported by the legal guardian that the patient had been hospitalized with Diabetic Ketoacidosis (DKA) on (b)(6) 2026 at The Great Western Hospital. The patient's blood glucose levels reached 22 mmol/L (396 mg/dL) with ketones of 3.7 while wearing the Pod between 1 and 4 hours. The LG reported that the Pod fell off of the infusion site (Arm), causing the cannula to dislodge. Symptoms reported include nausea, vomiting, loss of appetite and sweating. The patient was treated with intravenous fluids and insulin via pen. The patient was released the following day ((b)(6) 2026).
Patient problems: Alteration in Body Temperature; Decreased Appetite; Elevated Ketones/ Diabetic Ketoacidosis; Hyperglycemia; Nausea; Vomiting
MAUDE report 50938663 ↗InjurySep 29, 2026Insufficient Device Problem Information
It was reported that the patient was admitted to the hospital on (b)(6), 2026, with Diabetic Ketoacidosis (DKA). The patient¿s blood glucose rose to 20 mmol/L (360 mg/dL) while wearing the Pod for an unspecified duration. The patient reported feeling unwell on the night of (b)(6), 2026, and was transported by ambulance to Kingston General Hospital. The patient received antibiotic treatment and was discharged the same day. A few hours following discharge, the patient's condition worsened and they were transported back to the hospital by ambulance. Upon evaluation, the patient was diagnosed with DKA and hyperglycemia. Medical management included intravenous insulin therapy and antibiotics. The patient remained hospitalized until (b)(6), 2026, when they were discharged. The Pod was discarded by the hospital.
Patient problems: Elevated Ketones/ Diabetic Ketoacidosis; Hyperglycemia; Malaise
MAUDE report 50938324 ↗InjurySep 29, 2026Insufficient Device Problem Information
It was reported that the patient was hospitalized for hyperglycemia on (b)(6) 2026. The patient¿s blood glucose levels reached greater than 400 mg/dL while wearing the Pod for an unknown amount of time. Symptoms reported include vomiting, elevated blood glucose levels and a rapid heartbeat. No further information regarding Pod is available. Treatment details were not provided. The patient was discharged from the intensive care unit (ICU) after 24 hours.
Patient problems: Hyperglycemia; Vomiting
MAUDE report 50933853 ↗InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem
The legal guardian disclosed that the patient had been hospitalized on the preceding Friday, (b)(6) 2026, due to elevated ketones and progression toward diabetic ketoacidosis (DKA). The issue that was experienced that led to hospitalization was not disclosed. Additional attempts are being made to request further details including symptoms, treatment, and the length of hospitalization. If further information is provided, a supplemental report will be submitted.
Patient problems: Elevated Ketones/ Diabetic Ketoacidosis; Hyperglycemia
MAUDE report 50931269 ↗InjurySep 28, 2026Insufficient Device Problem Information
It was reported that the patient developed an infection, attributed to the Pod adhesive, at the Pod¿s infusion site (arm) while wearing the Pod for 24 hours. The patient did a 2-hour visit with their healthcare professional (HCP) to have the infected Pod site examined. The HCP reviewed a picture of the affected area but did not provide a specific diagnosis. The patient prescribed an unspecified 5-day course of antibiotics. The Pod was not worn during the visit as it already had been replaced.
Patient problems: Skin Infection
MAUDE report 50926366 ↗InjurySep 28, 2026Insufficient Device Problem Information
It was reported that the patient had been hospitalized with Diabetic Ketoacidosis (DKA) on (b)(6) 2026. The patient¿s blood glucose levels reached greater than 250 mg/dL while wearing the Pod for an unknown amount of time. The patient administered multiple corrective boluses, which were ineffective. Symptoms reported include nausea, pain and vomiting. The patient was treated with intravenous insulin drip and pain medication (Zofran, trazodone). The patient was discharged on (b)(6) 2026.
Patient problems: Elevated Ketones/ Diabetic Ketoacidosis; Hyperglycemia; Nausea; Pain; Vomiting
MAUDE report 50916999 ↗InjurySep 28, 2026Incorrect, Inadequate or Imprecise Result or Readings
It was reported that the patient went to the emergency room due to hyperglycemia. The patient's blood glucose levels reached 588 mg/dL while wearing the Pod for an unspecified period. No other symptoms were reported. The patient alleged that the previous evening, a Dexcom G7 sensor displayed a low blood glucose reading while a fingerstick readings showed 588 mg/dL. No treatment details were provided.
Patient problems: Hyperglycemia; Malaise
MAUDE report 50910516 ↗InjurySep 28, 2026Insufficient Device Problem Information
It was reported that the patient developed a staph infection at the Pod¿s insertion site while wearing the device between 24 and 36 hours. The patient reported the infusion site to have purulent drainage and swelling. The patient went to the emergency room where the area was incised and drained of pus. The Pod was discarded and is not available for return.
Patient problems: Bacterial Infection; Erythema; Pain; Purulent Discharge
MAUDE report 50910509 ↗InjurySep 28, 2026Insufficient Device Problem Information
It was reported that a skin irritation causing swelling, redness, and burning sensation had occurred while wearing the Pod for an unspecified period. The patient consulted their healthcare provider and was prescribed antibiotics. The Pod was discarded and is not available for return. No blood glucose readings were reported.
Patient problems: Burning Sensation; Erythema; Itching Sensation; Swelling/ Edema; Unspecified Infection
MAUDE report 50910506 ↗InjurySep 27, 2026Insufficient Device Problem Information
It was reported that the patient went to the emergency room due to severe hypoglycemia in early(b)(6)2026. The patient¿s blood glucose levels declined to 40 mg/dL while wearing the Pod for an unspecified period. Symptoms reported include palpitations, dizziness, sweating, difficulty walking, falling, dry lips, and nervousness. The patient was treated with ¿something sweet¿. The patient was released the same day. The Pod was discarded and is not available for return.
Patient problems: Dizziness; Dry Mouth; Dyskinesia; Emotional Changes; Hypoglycemia
MAUDE report 50909896 ↗InjurySep 27, 2026Insufficient Device Problem Information
It was reported that the patient had been hospitalized for 5 days with hypoglycemia and gastroparesis between (b)(6) 2026. No blood glucose readings were reported. Duration of Pod usage was not provided. Symptoms reported include diaphoresis, vomiting, and diarrhea. The patient was treated with intravenous fluid and manual injections.
Patient problems: Alteration in Body Temperature; Diarrhea; Vomiting
MAUDE report 50909590 ↗InjurySep 26, 2026Insufficient Device Problem Information
It was reported that the patient developed a blood infection that affected their brain and resulted in hospitalization. The patient stated that their memory was impacted. Therefore, event details were ¿fuzzy¿. The patient was still able to confirm that a Pod was worn at the time of incident and it was worn for less than 1 hour. No blood glucose readings were reported. No additional event details are available.
Patient problems: Cognitive Changes; Unspecified Infection
MAUDE report 50906908 ↗InjurySep 25, 2026Unintended Movement; Insufficient Device Problem Information
It was reported that the patient removed the Pod because of a new spray from a healthcare professional. While the purpose of the spray could not be confirmed from the available information, it may indicate management of a skin-related condition. Additionally, it was noted the pod fell off. No blood glucose values were provided. As treatment, a new pod was placed. Follow up has been conducted and is pending for further information. No further information is available at this time.
Patient problems: Skin Inflammation/ Irritation
MAUDE report 50905557 ↗InjurySep 25, 2026Insufficient Device Problem Information
It was reported that the patient had been hospitalized with Diabetic Ketoacidosis (DKA) on (b)(6) 2026. The patient's blood glucose levels reached 1000 mg/dL while wearing the Pod on the arm between 4 and 24 hours. The patient stated that on (b)(6) 2026, while wearing a Pod, they experienced severe hyperglycemia. According to the patient, they were asleep when their husband checked their glucose level and observed that it was reading \"high.\" The patient's husband attempted to wake patient; however, patient was reportedly unconscious and unresponsive.\Symptoms reported include loss of consciousness, unresponsiveness, and extremely elevated blood glucose levels. The patient stated that they were diagnosed with ketoacidosis (DKA) and pneumonia. The patient was informed by medical staff that it was related to a mechanical complication from the pump. Treatment reportedly included Novolog insulin administered intravenously (IV) and antibiotics. The patient reported that they remained hospitalized for approximately six days and was discharged on (b)(6) 2026. The pod was removed and discarded at the hospital.
Patient problems: Elevated Ketones/ Diabetic Ketoacidosis; Hyperglycemia; Loss of consciousness
MAUDE report 50905435 ↗
FDA clearance and classification
The FDA regulates product code QFG (alternate controller enabled insulin infusion pump) under 21 CFR 880.5730 as a Class II device. 21 CFR 880.5730 ↗
Devices of this type normally need 510(k) clearance before sale.
FDA definition of this device type: An alternate controller enabled insulin infusion pump (ACE insulin pump) is a device intended for the infusion of insulin into a patient. The ACE insulin pump may include basal and bolus drug delivery at set or variable rates. ACE insulin pumps are designed to reliably and securely communicate with external devices, such as automated insulin dosing systems, to allow insulin delivery commands to be received, executed, and confirmed. ACE insulin pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of insulin delivery.
Common questions about the Omnipod 5 Pod POD-BLE-H1-520
Has the Omnipod 5 Pod POD-BLE-H1-520 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2603-2026 (Class I), began on May 20, 2026: External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Omnipod 5 Pod POD-BLE-H1-520?
The FDA's MAUDE database lists 71,491 adverse event reports for this device, received between Mar 8, 2022 and Sep 30, 2026. In those reports, the most common specific device problems are “Unintended Movement” (32,747), “Loss of or Failure to Bond” (29,011) and “Material Twisted/Bent” (12,573); the most common patient problems are “Hyperglycemia” (53,830), “Elevated Ketones/ Diabetic Ketoacidosis” (2,884) and “Vomiting” (1,817). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Omnipod 5 Pod POD-BLE-H1-520?
74 reports are classified as deaths, 9,398 as injuries and 62,019 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Omnipod 5 Pod POD-BLE-H1-520 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the Omnipod 5 Pod POD-BLE-H1-520, and is it still sold?
The FDA's device database lists Insulet Corporation as the company responsible for it (brand Omnipod 5 Pod, model POD-BLE-H1-520). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Alternate Controller Enabled Insulin Infusion Pump devices
- t:slim X2 Insulin Pump with Control-IQ 1006402Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump2 recalls18 deaths159,198 reports
- t:slim X2 Insulin Pump with Control-IQ 1006408Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump2 recalls12 deaths92,505 reports
- t:slim X2 Insulin Pump with Control-IQ 1009637Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall4 deaths83,767 reports
- Omnipod 5 Pod POD-OMNI-I1-6720Insulet Corporation · Alternate Controller Enabled Insulin Infusion Pump2 recalls11 deaths60,668 reports
- t:slim X2 Insulin Pump with Control-IQ 1009643Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall7 deaths51,050 reports
- t:slim X2 Insulin Pump with Control-IQ Technology 1013198Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall8 deaths47,498 reports
- t:slim X2 Insulin Pump with Control-IQ 1006404Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall18 deaths43,714 reports
- t:slim X2 Insulin Pump with Control-IQ 1005011Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall3 deaths35,598 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page