Skip to content

Plum 2079225Recalls & reported problems

ICU Medical, Inc. · Plum A+ Version 13.43 w/MedNet · Class 2

All models of Plum (26)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI M3352079204251), received between Aug 29, 2017 and Aug 20, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Infusion or Flow Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Infusion or Flow Problem2
  • Device Sensing Problem1
  • Excess Flow or Over-Infusion1
  • Inaccurate Flow Rate1
  • Smoking1

Patient problems

  • No Consequences Or Impact To Patient2
  • No Clinical Signs, Symptoms or Conditions1
  • No Patient Involvement1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 20, 2024Inaccurate Flow Rate; Infusion or Flow Problem; Inaccurate Flow Rate; Infusion or Flow Problem

    THE COMPLAINT/EVENT INVOLVED A PLUM A+ (11.3 VERSION) NEW RESTART. IT WAS REPORTED THAT A POTASSIUM CHLORIDE DRIP WAS INSTALLED (NOT LABELED ON THE PUMP). A TOTAL VOLUME OF 500 ML WAS INSTALLED AT 40 CC/H, AND IT WAS REPORTED THAT IN 01 HOUR THE TOTAL VOLUME OF 500 ML WAS INFUSED. THIS OCCURED BETWEEN 9:00 P.M. AND 12:00 A.M. THE INFIRMARY REPORTED THAT THE INFUSION RATE WAS DISPLAYED AT 47 CC/H. THE MESSAGE "CASSETTE ERROR" WAS DISPLAYED, BUT SUBSEQUENTLY OTHER MEDICATIONS CONTINUED TO BE INFUSED, FOR WHICH NO MESSAGE WAS DISPLAYED. THERE WAS NO CLINICAL ALTERATION IN THE PATIENT AFTER THE EVENT AND UNSPECIFIED PERTINENT MEASURES WERE TAKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20006413 ↗
  • MalfunctionMay 5, 2020Infusion or Flow Problem; Infusion or Flow Problem

    THIS IS A GENERAL COMPLAINT OF UNSPECIFIED PLUM A+ PUMPS THAT ON UNKNOWN DATES, WERE REPORTED TO HAVE ISSUES WITH AIR GETTING INTO THE CHAMBER AND NOT ALARMING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10025947 ↗
  • MalfunctionDec 11, 2019Excess Flow or Over-Infusion; Device Sensing Problem; Excess Flow or Over-Infusion; Device Sensing Problem

    THE EVENT INVOLVED A PLUM A+ PUMP THAT WAS REPORTED TO BE OVER INFUSING AND DID NOT AUDIBLY ALARM FOR AIR IN LINE. THE DEVICE WAS REPORTED TO DELIVER 50 MLS FASTER THAN IT SHOULD HAVE. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9449724 ↗
  • MalfunctionAug 29, 2017Smoking; Smoking

    THE CUSTOMER REPORTED THE DEVICE HAD A SMOKE PHENOMENON. TESTING AND INVESTIGATION CONFIRMED THE SMOKE PHENOMENON AND FOUND THAT THE POWER SUPPLY WAS BURNT AND REQUIRED REPLACEMENT. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Patient Involvement

    MAUDE report 6827673 ↗

Other Pump, infusion devices

All Pump, infusion devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page