ProxiDiagnost N90Recalls & reported problems across all models

Philips Medical Systems DMC GmbH · 2 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

5 open recalls name at least one ProxiDiagnost N90 model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls
5

We found 2 models of ProxiDiagnost N90 by Philips Medical Systems DMC GmbH in the FDA's device registry (GUDID). Together they are named in 7 adverse event reports in the FDA's MAUDE database, received between Dec 17, 2020 and Apr 22, 2026 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

5 FDA recall records name at least one of these models by its device identifier, 5 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

ProxiDiagnost N90 recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Oct 1, 2025Open, Classified

    Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual movement) due to potential manufacturing issues.

    Z-0461-2026 · Root cause: Device Design · Recall details · FDA record ↗

  2. Dec 15, 2023Open, Classified

    Issue 1: Mixed Images: The Radio Fluoroscopy (RF) viewer may also display a previous patient s radiography (RAD) images when starting the next patient scan while the previous patient image export is still processing. If the issue occurs, there is a potential for incorrect patient data to be displayed in the image. Issue 2: Detector Access point: Security vulnerability specific to the Wireless Portable Detector configuration items in Philips Support Connect (PSC). Due to this vulnerability it is possible, with physical access to the system and knowledge of specific settings, to modify and export data to removable media (example: USB)

    Z-1072-2024 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

  3. Sep 21, 2023Open, Classified

    Potential for units suspended on the ceiling to fail and fall.

    Z-0280-2024 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗

  4. Oct 21, 2022Open, Classified

    Potential table stop due to a broken tabletop cable: On the ProxiDiagnost N90 table there is a cable located under the tabletop that is covered by a plastic sleeve which can crack and damage the cable. If the tabletop cable breaks, the user may not be able to initiate table tilting movements or braking

    Z-1477-2023 · Root cause: Component design/selection · names 2 of these models · Recall details · FDA record ↗

  5. Oct 21, 2021Open, Classified

    If the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell smoke emitting from the cabinet, or hear a popping sound coming from the cabinet.

    Z-1356-2023 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

Most reported ProxiDiagnost N90 problems

Device problems

  • Detachment of Device or Device Component1
  • Device Fell1
  • Electrical /Electronic Property Problem1
  • Flare or Flash1
  • Mechanical Jam1
  • Misassembled During Installation1
  • Radiation Output Problem1
  • Smoking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Muscle/Tendon Damage1
  • No Patient Involvement1

ProxiDiagnost N90 models

2 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent ProxiDiagnost N90 adverse event reports

The 7 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Philips Medical Systems DMC GmbH products

All Philips Medical Systems DMC GmbH devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page