ProxiDiagnost N90 706100Recalls & reported problems

Philips Medical Systems DMC GmbH · Image-intensified fluoroscopic x-ray system · Class 2

All models of ProxiDiagnost N90 (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
4

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00884838085619), received between Dec 17, 2020 and Apr 22, 2026 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

4 FDA recall records name this device's identifier, 4 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 15, 2023Class IIOpen, Classified

    Issue 1: Mixed Images: The Radio Fluoroscopy (RF) viewer may also display a previous patient s radiography (RAD) images when starting the next patient scan while the previous patient image export is still processing. If the issue occurs, there is a potential for incorrect patient data to be displayed in the image. Issue 2: Detector Access point: Security vulnerability specific to the Wireless Portable Detector configuration items in Philips Support Connect (PSC). Due to this vulnerability it is possible, with physical access to the system and knowledge of specific settings, to modify and export data to removable media (example: USB)

    Z-1072-2024 · Root cause: Software design · Recall details · FDA record ↗

  2. Sep 21, 2023Class IIOpen, Classified

    Potential for units suspended on the ceiling to fail and fall.

    Z-0280-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. Oct 21, 2022Class IIOpen, Classified

    Potential table stop due to a broken tabletop cable: On the ProxiDiagnost N90 table there is a cable located under the tabletop that is covered by a plastic sleeve which can crack and damage the cable. If the tabletop cable breaks, the user may not be able to initiate table tilting movements or braking

    Z-1477-2023 · Root cause: Component design/selection · Recall details · FDA record ↗

  4. Oct 21, 2021Class IIOpen, Classified

    If the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell smoke emitting from the cabinet, or hear a popping sound coming from the cabinet.

    Z-1356-2023 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Electrical /Electronic Property Problem1
  • Flare or Flash1
  • Mechanical Jam1
  • Misassembled During Installation1
  • Radiation Output Problem1
  • Smoking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Muscle/Tendon Damage1
  • No Patient Involvement1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 22, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE PIECE OF THE FOOTBOARD WAS MISSING, RESULTING IN EXPOSED PARTS AND A POTENTIAL SAFETY CONCERN. THE DEVICE WAS OUTSIDE OF USE AND THERE WAS NO HARM TO THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24956411 ↗
  • InjuryFeb 17, 2026Mechanical Jam

    IT WAS REPORTED THAT THE WALKING MECHANISMS LOCK, AND WHEN MOVING FROM ONE SIDE OF THE ROOM TO ANOTHER, THE BUTTON CAN BE PUSHED TO MOVE FROM SIDE TO SIDE. WHEN PULLING IT ACROSS THE ROOM, IT JERKED THE STAFF'S ARM, CAUSING SHOULDER DAMAGE. THE DEVICE WAS IN CLINICAL USE AND THERE WAS NO PATIENT HARM.

    Patient problems: Muscle/Tendon Damage

    MAUDE report 24373816 ↗
  • MalfunctionJun 18, 2025Electrical /Electronic Property Problem; Smoking; Electrical /Electronic Property Problem; Smoking

    IT WAS REPORTED THAT ELECTRICAL PART IN M-CABINET BURNED UP. THE USE OF DEVICE WAS UNKNOWN AND THERE WAS NO PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22269657 ↗
  • MalfunctionMar 22, 2023Misassembled During Installation; Misassembled During Installation

    DURING INSTALLATION OF THE SYSTEM, IT WAS FOUND THAT A COUPLE OF THE RAIL CONNECTION BLOCKS WERE MIS-ALIGNED IN THE CEILING UNISTRUT. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16591773 ↗
  • MalfunctionFeb 7, 2023Flare or Flash; Flare or Flash

    WHILE THE SYSTEM WAS BEING USED FOR A FLUOROSCOPY STUDY, THE MAINS CABINET STARTED MAKING LOUD POPPING SOUNDS ALONG WITH SPARKING, SMOKING, AND AN ELECTRICAL FIRE INSIDE OF THE CABINET. NOBODY WAS INJURED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16313564 ↗
  • MalfunctionDec 17, 2020Radiation Output Problem; Radiation Output Problem

    DURING ANNUAL PHYSICS TESTING, THE STABILIZED DOSE IS HIGH DURING CONTINUOUS FLUOROSCOPY AND PULSED FLUOROSCOPY FOR 20 AND 30 FRAMES PER SECOND.

    Patient problems: No Patient Involvement

    MAUDE report 11027848 ↗

Other System, x-ray, fluoroscopic, image-intensified devices

All System, x-ray, fluoroscopic, image-intensified devices →