Skip to content

ProxiDiagnost N90 706110Recalls & reported problems

Philips Medical Systems DMC GmbH · The ProxiDiagnost N90 is a multi-functional general Radiography/ Fluoroscopy (R/F) system and includes • Footrest • Hand Grips Radiation Protection Accessories • Flexible Radiation Protection Apron • Front Radiation Protection Apron Additional Accessories (Optional) • Monitor Trolley • Monitor Ceiling Suspension • Parking Frame for Accessories • Shoulder Support • Side bar • Compression Belt • Adjustable Lateral Cassette Holder • Leg Supports • Infusion Bottle Holder • Arm Support for Catheterization • Ankle Clamps • Overhead Hand Grip • Adult Headrest • Mattress • Rotatable Stool for Footrest • Pediatric Micturition Set • Stretch Grip for Wall Stand • Bar Code Scanner • Patient Support • Stitching Ruler Accessories for the SkyPlate Detector (Optional) • Mobile Detector Holder • Detector Holder Patient Bed • Portable Panel Protector • Detector Handle • WPD Bags • Grids for SkyPlate Detector large · Class 2

All models of ProxiDiagnost N90 (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

5 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
5

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00884838103221), received between Jun 21, 2024 and Jun 21, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Fell” (1 report).

5 FDA recall records name this device's identifier, 5 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 1, 2025Class IIOpen, Classified

    Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual movement) due to potential manufacturing issues.

    Z-0461-2026 · Root cause: Device Design · Recall details · FDA record ↗

  2. Dec 15, 2023Class IIOpen, Classified

    Issue 1: Mixed Images: The Radio Fluoroscopy (RF) viewer may also display a previous patient s radiography (RAD) images when starting the next patient scan while the previous patient image export is still processing. If the issue occurs, there is a potential for incorrect patient data to be displayed in the image. Issue 2: Detector Access point: Security vulnerability specific to the Wireless Portable Detector configuration items in Philips Support Connect (PSC). Due to this vulnerability it is possible, with physical access to the system and knowledge of specific settings, to modify and export data to removable media (example: USB)

    Z-1072-2024 · Root cause: Software design · Recall details · FDA record ↗

  3. Sep 21, 2023Class IIOpen, Classified

    Potential for units suspended on the ceiling to fail and fall.

    Z-0280-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  4. Oct 21, 2022Class IIOpen, Classified

    Potential table stop due to a broken tabletop cable: On the ProxiDiagnost N90 table there is a cable located under the tabletop that is covered by a plastic sleeve which can crack and damage the cable. If the tabletop cable breaks, the user may not be able to initiate table tilting movements or braking

    Z-1477-2023 · Root cause: Component design/selection · Recall details · FDA record ↗

  5. Oct 21, 2021Class IIOpen, Classified

    If the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell smoke emitting from the cabinet, or hear a popping sound coming from the cabinet.

    Z-1356-2023 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Fell1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 21, 2024Use of Device Problem; Device Fell; Use of Device Problem; Device Fell

    IT WAS REPORTED THAT THERE WAS DAMAGE TO THE DETECTOR (SKYPLATE) WITH ARTIFACTS. IT WAS UNABLE TO CONNECT PORTABLE SKYPLATE DETECTOR TO THE SYSTEM. THERE WAS NO HARM TO PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19583462 ↗

Other System, X-Ray, Fluoroscopic, Image-Intensified devices

All System, X-Ray, Fluoroscopic, Image-Intensified devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page