Skip to content

RevolutionRecalls & reported problems across all models

GE Medical Systems, L.L.C. · 7 models in the FDA device registry

Data updated Oct 18, 2026 from official FDA records · Editorial policy

14 open recalls name at least one Revolution model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls
14

We found 7 models of Revolution by GE Medical Systems, L.L.C. in the FDA's device registry (GUDID). Together they are named in 1 adverse event report in the FDA's MAUDE database, received between Jan 13, 2026 and Jan 13, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).

14 FDA recall records name at least one of these models by its device identifier, 14 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Revolution recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Mar 26, 2026Class IIOpen, Classified

    GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.

    Z-2154-2026 · Root cause: Software design · Recall details · FDA record ↗

  2. Mar 26, 2026Class IIOpen, Classified

    GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.

    Z-2153-2026 · Root cause: Software design · Recall details · FDA record ↗

  3. Mar 26, 2026Class IIOpen, Classified

    GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.

    Z-2152-2026 · Root cause: Software design · Recall details · FDA record ↗

  4. Mar 26, 2026Class IIOpen, Classified

    GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.

    Z-2151-2026 · Root cause: Software design · Recall details · FDA record ↗

  5. Mar 26, 2026Class IIOpen, Classified

    GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.

    Z-2150-2026 · Root cause: Software design · Recall details · FDA record ↗

  6. Feb 18, 2025Class IIOpen, Classified

    GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.

    Z-1303-2025 · Root cause: Process change control · names 4 of these models · Recall details · FDA record ↗

  7. Feb 18, 2025Class IIOpen, Classified

    GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.

    Z-1302-2025 · Root cause: Process change control · names 5 of these models · Recall details · FDA record ↗

  8. Feb 18, 2025Class IIOpen, Classified

    GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.

    Z-1301-2025 · Root cause: Process change control · Recall details · FDA record ↗

  9. Feb 18, 2025Class IIOpen, Classified

    GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.

    Z-1300-2025 · Root cause: Process change control · names 3 of these models · Recall details · FDA record ↗

  10. Feb 18, 2025Class IIOpen, Classified

    GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.

    Z-1299-2025 · Root cause: Process change control · names 4 of these models · Recall details · FDA record ↗

  11. Feb 18, 2025Class IIOpen, Classified

    GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.

    Z-1298-2025 · Root cause: Process change control · Recall details · FDA record ↗

  12. Feb 18, 2025Class IIOpen, Classified

    GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.

    Z-1297-2025 · Root cause: Process change control · names 4 of these models · Recall details · FDA record ↗

  13. Nov 5, 2024Class IIOpen, Classified

    GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.

    Z-0615-2025 · Root cause: Component design/selection · Recall details · FDA record ↗

  14. Jun 2, 2022Class IIOpen, Classified

    As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place resulting in injury to service personnel.

    Z-1511-2022 · Root cause: Error in labeling · names 2 of these models · Recall details · FDA record ↗

Most reported Revolution problems

Device problems

  • Improper or Incorrect Procedure or Method1

Patient problems

  • Unspecified Tissue Injury1

Revolution models

7 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Revolution adverse event reports

The 1 most recent reports across all models, in the reporter's own words as published by the FDA.

Other GE Medical Systems, L.L.C. products

All GE Medical Systems, L.L.C. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page