Revolution recallZ-1301-2025 · GE Medical Systems, LLC · initiated Feb 18, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Medical Systems, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
FDA root cause category: Process change control
What the company is doing
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/18/2025 via letter delivered using a traceable means. The notice explained the issue, safety issue, and requested the following actions be taken: "You can continue to use your system. Please monitor the floor near the base of the gantry for any accumulation of blue fluid. If blue fluid is observed, wear nitrile or rubber gloves, clean up the fluid by absorbing it with a noncombustible absorbent material, and dispose of the absorbent material in a suitable container for disposal. Please contact your local GE HealthCare service representative if any blue fluid is observed. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please complete and return the attached acknowledgement form to [email protected]."
Recall details
- Recall number
- Z-1301-2025
- Status
- Open, Classified
- Initiated by the company
- Feb 18, 2025
- Posted by the FDA
- Mar 7, 2025
- Recalling firm
- GE Medical Systems, LLC
- Quantity
- 1 unit
- Product code
- JAK
- FDA submissions
- K213715
- Distribution
- Worldwide distribution.
Product description
GE Healthcare Revolution Apex Select, System, X-ray, Tomography, Computed
Code information (models, lots, serial numbers)
System ID Number 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Revolution Apex EliteGE Medical Systems, L.L.C. · System, x-ray, tomography, computed8 recalls0 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K213715 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Revolution Apex SelectGE Hangwei Medical Systems Company Ltd. · System, x-ray, tomography, computed2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96370).
- Z-1297-2025 · Open, ClassifiedGE Healthcare Revolution Apex, System, X-ray, Tomography, Computed
- Z-1298-2025 · Open, ClassifiedGE Healthcare Revolution Apex Elite, System, X-ray, Tomography, Computed
- Z-1299-2025 · Open, ClassifiedGE Healthcare Revolution Apex Expert, System, X-ray, Tomography, Computed
- Z-1300-2025 · Open, ClassifiedGE Healthcare Revolution Apex Plus, System, X-ray, Tomography, Computed
- Z-1302-2025 · Open, ClassifiedGE Healthcare Revolution CT, System, X-ray, Tomography, Computed
- Z-1303-2025 · Open, ClassifiedGE Healthcare Revolution CT ES, System, X-ray, Tomography, Computed
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page