Skip to content

Revolution Apex PlusRecalls & reported problems

GE Medical Systems, L.L.C. · Revolution Apex Plus - Designated System Medical Device · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00195278502179), received between Jan 13, 2026 and Jan 13, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

17 other recalls cover devices cleared under the same FDA submission (K191777); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 26, 2026Open, Classified

    GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.

    Z-2152-2026 · Root cause: Software design · FDA record ↗

  2. Feb 18, 2025Open, Classified

    GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.

    Z-1302-2025 · Root cause: Process change control · FDA record ↗

Most reported problems

Device problems

  • Improper or Incorrect Procedure or Method1

Patient problems

  • Unspecified Tissue Injury1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 13, 2026Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT, DURING INSTALLATION OF THE SYSTEM, A FIELD ENGINEER'S THUMB GOT PINCHED IN THE TABLE SPRING, CAUSING A WOUND THAT WAS TREATED WITH SURGERY.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 24054776 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K191777. One submission often covers many models, so check the product and lot details in each record.

  • Z-2150-2026 · Mar 26, 2026 · Open, Classified
    Revolution CT, x-ray computed tomography, Model Numbers 5590000 and 5590000-6
  • Z-2153-2026 · Mar 26, 2026 · Open, Classified
    Revolution CT ES, X-ray Computed Tomography, Model Number 5590000-15
  • Z-2154-2026 · Mar 26, 2026 · Open, Classified
    Revolution Apex, X-ray Computed Tomography, Model Number 5590000-20
  • Z-1300-2025 · Feb 18, 2025 · Open, Classified
    GE Healthcare Revolution Apex Plus, System, X-ray, Tomography, Computed
  • Z-1303-2025 · Feb 18, 2025 · Open, Classified
    GE Healthcare Revolution CT ES, System, X-ray, Tomography, Computed
  • Z-1297-2025 · Feb 18, 2025 · Open, Classified
    GE Healthcare Revolution Apex, System, X-ray, Tomography, Computed
  • Z-0816-2022 · Jun 30, 2021 · Open, Classified
    Revolution CT, Revolution CT ES
  • Z-2021-2021 · Jun 30, 2021 · Open, Classified
    The system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission projection data from the same axial plane taken at different a
  • Z-0815-2022 · Jun 30, 2021 · Open, Classified
    Revolution Apex, Revolution CT with Apex Edition
  • Z-2022-2021 · Jun 30, 2021 · Open, Classified
    The system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission projection data from the same axial plane taken at different a
  • Z-0242-2022 · Jun 30, 2021 · Open, Classified
    Computed Tomography Systems Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition
  • Z-2004-2021 · Mar 19, 2021 · Completed
    Revolution CT
  • Z-1130-2021 · Dec 17, 2020 · Terminated
    System, X-Ray, Tomography, Computed: Revolution CT, Revolution CT ES.
  • Z-1129-2021 · Dec 17, 2020 · Terminated
    System, X-Ray, Tomography, Computed: Revolution Apex, Revolution CT with Apex Edition
  • Z-0695-2021 · Nov 23, 2020 · Terminated
    Revolution CT with Apex Edition CT System - Product Usage: intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications.

…and 2 more.

Other System, x-ray, tomography, computed devices

All System, x-ray, tomography, computed devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page