GE Healthcare Revolution Apex, System, X-ray, Tomography… recallZ-1297-2025 · GE Medical Systems, LLC · initiated Feb 18, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Medical Systems, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.

FDA root cause category: Process change control

What the company is doing

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/18/2025 via letter delivered using a traceable means. The notice explained the issue, safety issue, and requested the following actions be taken: "You can continue to use your system. Please monitor the floor near the base of the gantry for any accumulation of blue fluid. If blue fluid is observed, wear nitrile or rubber gloves, clean up the fluid by absorbing it with a noncombustible absorbent material, and dispose of the absorbent material in a suitable container for disposal. Please contact your local GE HealthCare service representative if any blue fluid is observed. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please complete and return the attached acknowledgement form to [email protected]."

Recall details

Recall number
Z-1297-2025
Status
Open, Classified
Initiated by the company
Feb 18, 2025
Posted by the FDA
Mar 7, 2025
Recalling firm
GE Medical Systems, LLC
Quantity
56 units
Product code
JAK
FDA submissions
K163213, K191777, K213715
Distribution
Worldwide distribution.

Product description

GE Healthcare Revolution Apex, System, X-ray, Tomography, Computed

Code information (models, lots, serial numbers)

Show the full code information (6k characters)

System ID Number: 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN; System ID Number: 650493REV1, UDI To be provided, Mfg. Lot/Serial Number REVVX1900035CN; System ID Number: 630266CTRV, UDI (01)00195278460271(11)240800(21)REV2X2400093CN, Mfg. Lot/Serial Number REV2X2400093CN; System ID Number: 773702REVCT6ED, UDI (01)00195278460271(11)231000(21)REV2X2300071CN, Mfg. Lot/Serial Number REV2X2300071CN; System ID Number: 312996CTRV2, UDI (01)00195278460271(11)241000(21)REV2X2400126CN, Mfg. Lot/Serial Number REV2X2400126CN; System ID Number: 1000000018, UDI To be provided, Mfg. Lot/Serial Number REV2X2300074CN; System ID Number: 617726Y6REV2, UDI (01)00840682118552(11)170900(21)REVVX1700090CN, Mfg. Lot/Serial Number REVVX1700090CN; System ID Number: 617726Y6REV1, UDI (01)00840682118552(11)160700(21)REVVX1600018CN, Mfg. Lot/Serial Number REVVX1600018CN; System ID Number: 336890APEX, UDI (01)00195278460271(11)240900(21)REV2X2400116CN, Mfg. Lot/Serial Number REV2X2400116CN; System ID Number: 704801APEX, UDI (01)00195278460271(11)241000(21)REV2X2400139CN, Mfg. Lot/Serial Number REV2X2400139CN; System ID Number: 336832APEX, UDI (01)00195278460271(11)241000(21)REV2X2400129CN, Mfg. Lot/Serial Number REV2X2400129CN; System ID Number: 980993CT1, UDI (01)00840682146616(11)210700(21)REV2A2100029CN, Mfg. Lot/Serial Number REV2A2100029CN; System ID Number: 910715REV1, UDI (01)00195278460271(11)241200(21)REV2X2400155CN, Mfg. Lot/Serial Number REV2X2400155CN; System ID Number: 603650APEX1, UDI (01)00195278460271(11)240600(21)REV2X2400068CN, Mfg. Lot/Serial Number REV2X2400068CN; System ID Number: 603526APEX, UDI (01)00195278469700(11)230100(21)REV242200004CN, Mfg. Lot/Serial Number REV242200004CN; System ID Number: 973971REV161, UDI (01)00195278460271(11)241200(21)REV2X2400158CN, Mfg. Lot/Serial Number REV2X2400158CN; System ID Number: 914787APEX, UDI (01)00195278460271(11)231000(21)REV2X2300081CN, Mfg. Lot/Serial Number REV2X2300081CN; System ID Number: 718963EDCT, UDI (01)00195278460271(11)240600(21)REV2X2400070CN, Mfg. Lot/Serial Number REV2X2400070CN; System ID Number: 315473APEX1, UDI (01)00195278460271(11)241100(21)REV2X2400152CN, Mfg. Lot/Serial Number REV2X2400152CN; System ID Number: 614RMHAPEX1, UDI (01)00195278460271(11)240300(21)REV2X2400026CN, Mfg. Lot/Serial Number REV2X2400026CN; System ID Number: 724357APEX, UDI (01)00195278460271(11)230800(21)REV2X2300046CN, Mfg. Lot/Serial Number REV2X2300046CN; System ID Number: 843797REVCT2, UDI (01)00840682123471(11)211100(21)REVV82100041CN, Mfg. Lot/Serial Number REVV82100041CN; System ID Number: 806775APEX, UDI (01)00195278460271(11)231000(21)REV2X2300075CN, Mfg. Lot/Serial Number REV2X2300075CN; System ID Number: 703776APEX, UDI To be provided, Mfg. Lot/Serial Number REV2X2400167CN; System ID Number: 414805REVCT, UDI Not Applicable, Mfg. Lot/Serial Number 00000441066CN8; System ID Number: 262577REVCT, UDI (01)00195278460271(11)240500(21)REV2X2400055CN, Mfg. Lot/Serial Number REV2X2400055CN; System ID Number: 832CT000059, UDI (01)00840682146616(11)220100(21)REV2A2200006CN, Mfg. Lot/Serial Number REV2A2200006CN; System ID Number: 833CT000155, UDI (01)00195278460271(11)240500(21)REV2X2400053CN, Mfg. Lot/Serial Number REV2X2400053CN; System ID Number: AH7084CT02, UDI (01)00840682146616(11)220100(21)REV2A2200007CN, Mfg. Lot/Serial Number REV2A2200007CN; System ID Number: AH1923CT01, UDI (01)00195278460271(11)240300(21)REV2X2400024CN, Mfg. Lot/Serial Number REV2X2400024CN; System ID Number: BG4549CT04, UDI (01)00195278460271(11)230600(21)REV2X2300026CN, Mfg. Lot/Serial Number REV2X2300026CN; System ID Number: E416249APEXCTE, UDI (01)00195278460271(11)231100(21)REV2X2300087CN, Mfg. Lot/Serial Number REV2X2300087CN; System ID Number: 082421140968, UDI To be provided, Mfg. Lot/Serial Number Not Available; System ID Number: 082421140960, UDI (01)00195278633149(11)240500(21)CBFCG2400006HM, Mfg. Lot/Serial Number CBFCG2400006HM; System ID Number: 082421140959, UDI (01)00195278633149(11)240500(21)CBFCG2400004HM, Mfg. Lot/Serial Number CBFCG2400004HM; System ID Number: DK1044CT12, UDI (01)00840682146616(11)200200(21)REV2A2000001CN, Mfg. Lot/Serial Number REV2A2000001CN; System ID Number: DK1547CT01, UDI (01)00840682123471(11)210100(21)REVV82100002CN, Mfg. Lot/Serial Number REVV82100002CN; System ID Number: DK1051CT04, UDI To be provided, Mfg. Lot/Serial Number 00000442412CN3; System ID Number: CT52261701, UDI To be provided, Mfg. Lot/Serial Number REV2X2300050CN; System ID Number: M40230104, UDI (01)00840682118552(11)160600(21)REVVX1600012CN, Mfg. Lot/Serial Number REVVX1600012CN; System ID Number: REV020, UDI To be provided, Mfg. Lot/Serial Number 00000442334CN9; System ID Number: CT00486, UDI (01)00195278460271(11)231000(21)REV2X2300079CN, Mfg. Lot/Serial Number REV2X2300079CN; System ID Number: KRCT54972221, UDI To be provided, Mfg. Lot/Serial Number Not Available; System ID Number: KRCT54893311, UDI To be provided, Mfg. Lot/Serial Number Not Available; System ID Number: 0850211084, UDI To be provided, Mfg. Lot/Serial Number REV2X2300016CN; System ID Number: NO1068CT04, UDI (01)00840682118552(11)171000(21)REVVX1700106CN, Mfg. Lot/Serial Number REVVX1700106CN; System ID Number: P00652CT01, UDI (01)00195278460271(11)240200(21)REV2X2400017CN, Mfg. Lot/Serial Number REV2X2400017CN; System ID Number: PL1672CT03, UDI (01)00840682146616(11)210600(21)REV2A2100019CN, Mfg. Lot/Serial Number REV2A2100019CN; System ID Number: SA1420CT01, UDI To be provided, Mfg. Lot/Serial Number REV2X2400137CN; System ID Number: SA2973CT01, UDI To be provided, Mfg. Lot/Serial Number REV2X23000030CN; System ID Number: CT53972381, UDI To be provided, Mfg. Lot/Serial Number REV2X2400121CN; System ID Number: A201CT07, UDI (01)00195278460271(11)240200(21)REV2X2400014CN, Mfg. Lot/Serial Number REV2X2400014CN; System ID Number: AC001CT05, UDI (01)00840682118552(11)181000(21)REVVX1800101CN, Mfg. Lot/Serial Number REVVX1800101CN; System ID Number: 541010CT02, UDI (01)00840682118552(11)190900(21)REVVX1900106CN, Mfg. Lot/Serial Number REVVX1900106CN; System ID Number: THCTX24524601, UDI (01)00195278460271(11)241000(21)REV2X2400131CN, Mfg. Lot/Serial Number REV2X2400131CN; System ID Number: AE1025CT03, UDI (01)00840682118552(11)210300(21)REVVX2100010CN, Mfg. Lot/Serial Number REVVX2100010CN.

Devices named in this recall

5 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

2 devices cleared under K163213, K191777, K213715 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 96370).