SABERX .014 PTA Dilatation CatheterRecalls & reported problems across all models
Cordis US Corp. · 16 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 34
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls
- 0
We found 16 models of SABERX .014 PTA Dilatation Catheter by Cordis US Corp. in the FDA's device registry (GUDID). Together they are named in 34 adverse event reports in the FDA's MAUDE database, received between Sep 30, 2024 and Apr 8, 2026 — classified as 3 injuries, 31 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (30 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported SABERX .014 PTA Dilatation Catheter problems
Device problems
- Burst Container or Vessel30
- Detachment of Device or Device Component6
- Difficult to Remove2
- Material Rupture2
- Decrease in Pressure1
- Deflation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions34
SABERX .014 PTA Dilatation Catheter models
16 models, most reported first. Each model has its own page with its reports, recalls and device details.
- SABERX .014 PTA Dilatation Catheter 47512015MCatheter, angioplasty, peripheral, transluminal8 reports
- SABERX .014 PTA Dilatation Catheter 47525300MCatheter, angioplasty, peripheral, transluminal7 reports
- SABERX .014 PTA Dilatation Catheter 47520300MCatheter, angioplasty, peripheral, transluminal3 reports
- SABERX .014 PTA Dilatation Catheter 47560200MCatheter, angioplasty, peripheral, transluminal2 reports
- SABERX .014 PTA Dilatation Catheter 47530300LCatheter, angioplasty, peripheral, transluminal2 reports
- SABERX .014 PTA Dilatation Catheter 47530300MCatheter, angioplasty, peripheral, transluminal2 reports
- SABERX .014 PTA Dilatation Catheter 47540150MCatheter, angioplasty, peripheral, transluminal1 report
- SABERX .014 PTA Dilatation Catheter 47560150MCatheter, angioplasty, peripheral, transluminal1 report
- SABERX .014 PTA Dilatation Catheter 47550200MCatheter, angioplasty, peripheral, transluminal1 report
- SABERX .014 PTA Dilatation Catheter 47530040MCatheter, angioplasty, peripheral, transluminal1 report
- SABERX .014 PTA Dilatation Catheter 47525200MCatheter, angioplasty, peripheral, transluminal1 report
- SABERX .014 PTA Dilatation Catheter 47520250LCatheter, angioplasty, peripheral, transluminal1 report
- SABERX .014 PTA Dilatation Catheter 47525150MCatheter, angioplasty, peripheral, transluminal1 report
- SABERX .014 PTA Dilatation Catheter 47535300MCatheter, angioplasty, peripheral, transluminal1 report
- SABERX .014 PTA Dilatation Catheter 47520150MCatheter, angioplasty, peripheral, transluminal1 report
- SABERX .014 PTA Dilatation Catheter 47530200MCatheter, angioplasty, peripheral, transluminal1 report
Recent SABERX .014 PTA Dilatation Catheter adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionApr 8, 2026Burst Container or Vessel
AS REPORTED, A 3MM X 40MM SABERX .014 RUPTURED DURING THE FIRST INFLATION AT 6 ATMOSPHERES (ATM). THE PROCEDURE WAS COMPLETED WITHOUT ISSUE USING AN UNKNOWN REPLACEMENT PRODUCT. THERE WAS NO REPORTED INJURY TO THE PATIENT. THE PRODUCT WAS STORED PROPERLY ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DIFFICULTY REMOVING THE PRODUCT FROM THE HOOP. THERE WAS NO DIFFICULTY REMOVING THE PROTECTIVE BALLOON COVER. THERE WAS NO DIFFICULTY REMOVING THE STYLET OR ANY OF THE STERILE PACKAGING COMPONENTS. THERE WERE NO KINKS OR OTHER DAMAGES NOTED PRIOR TO INSERTING THE PRODUCT INTO THE PATIENT. THE DEVICE WAS PREPPED PER THE IFU. THE DEVICE WAS PREPPED NORMALLY. THE DEVICE WAS USED FOR AN ENDOVASCULAR THROMBECTOMY (EVT). THERE WAS SEVERE VESSEL CALCIFICATION WITHOUT TORTUOSITY AND 100% STENOSIS. THERE WAS NO RESISTANCE OR FRICTION WILE INSERTING THE BALLOON THROUGH THE THROUGH THE ROTATING HEMOSTATIC VALVE, THE GUIDE CATHETER OR ADVANCING THE BALLOON CATHETER THROUGH THE VESSEL. THERE WAS NO DIFFICULTY CROSSING THE LESION WITH THE BALLOON CATHETER AND THE CATHETER WAS NEVER IN AN ACUTE BEND. THERE WAS NO BALLOON CATHETER KINK DURING USE. THE CONTRAST TO SALINE RATIO WAS 2:1. THE P
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24821749 ↗MalfunctionMar 18, 2026Burst Container or Vessel
AS REPORTED, THE 3.0 X 300 SABERX PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON PASSED THROUGH THE NON-CORDIS FIXED CATHETER (FC). HOWEVER, THE BALLOON RUPTURED BEFORE REACHING NOMINAL PRESSURE/SIZE. AFTERWARDS, A 3.0 X 200 SABERX BALLOON WAS USED AND THE PROCEDURE WAS COMPLETED WITHOUT PATIENT INJURY. ADDITIONAL INFORMATION WAS REQUESTED BUT NOT PROVIDED. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24632396 ↗MalfunctionMar 18, 2026Burst Container or Vessel
AS REPORTED, THE BALLOON ON A 1.25MM X 15MM SABERX .014 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED ON ITS FIRST INFLATION TO 8 ATMOSPHERES (ATM). THE DEVICE WAS REMOVED FROM THE PATIENT, AND A NON-CORDIS DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THIS WAS DURING A PROCEDURE TO TREAT A HIGHLY CALCIFIED ANTERIOR TIBIAL ARTERY (ATA) OCCLUSION VIA AN IPSILATERAL ANTEGRADE APPROACH. THE TARGET VESSEL WAS NOT TORTUOUS AND EXHIBITED 100% STENOSIS. THE SABERX .014 BALLOON CATHETER WAS BEING USED FOR PRE-DILATION. THE DEVICE WAS STORED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU), AND THERE WAS NO DIFFICULTY REMOVING THE STERILE PACKAGING COMPONENTS OR THE PROTECTIVE BALLOON COVER. THE DEVICE MAINTAINED NEGATIVE PRESSURE DURING PREPARATION, AND A 2:1 CONTRAST TO SALINE RATIO WAS USED. THE DEVICE WAS NOT PUT INTO AN ACUTE BEND, WAS ABLE TO CROSS THE LESION, DID NOT KINK DURING THE PROCEDURE, AND WAS REMOVED EASILY FROM THE PATIENT. THE PRODUCT WAS REMOVED INTACT IN ONE PIECE. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24627774 ↗MalfunctionMar 16, 2026Burst Container or Vessel
AS REPORTED, THE BALLOON ON A 2.5MM X 300MM SABERX .014 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED BEFORE NOMINAL DILATION. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THIS WAS DURING BELOW KNEE (BK) TREATMENT. THERE WAS SEVERE CALCIFICATION. ADDITIONAL INFORMATION WAS REQUESTED BUT WAS NOT PROVIDED. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24607849 ↗MalfunctionJan 16, 2026Burst Container or Vessel
AS REPORTED, THE BALLOON ON A 1.25MM X 15MM SABERX .014 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED AT APPROXIMATELY 7 ATMOSPHERES (ATM) DURING DILATION OF THE DORSALIS PEDIS ARTERY. ANOTHER 1.25MM X 15MM SABERX .014 PTA BALLOON CATHETER WAS USED; HOWEVER, THE BALLOON RUPTURED AGAIN AT APPROXIMATELY 13 ATMOSPHERES (ATM). A NON-CORDIS 1.5MM BALLOON CATHETER WAS THEN USED AS A REPLACEMENT AND WAS ABLE TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THIS WAS DURING A PROCEDURE TO TREAT STENOSIS OF THE LEFT SUPERFICIAL FEMORAL ARTERY (SFA), THE ANTERIOR TIBIAL ARTERY, AND THE DORSALIS PEDIS ARTERY. THE TARGET VESSEL WAS CHARACTERIZED BY MODERATE TO SEVERE CALCIFICATION, MODERATE TORTUOSITY, AND 100% STENOSIS. AN UNKNOWN 6F GUIDE SHEATH WAS INSERTED FROM THE LEFT GROIN, AND AN UNKNOWN .014 GUIDEWIRE WAS USED TO CROSS THE LESION. THE DEVICES WERE STORED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU), AND THERE WAS NO DIFFICULTY REMOVING ANY STERILE PACKAGING COMPONENTS OR PROTECTIVE BALLOON COVERS. THE DEVICES MAINTAINED NEGATIVE PRESSURE DURING PREPARATION. BOTH SABERX .014 PTA BALLOON CATHETERS WERE ABLE TO CROSS THE LESION PRIOR TO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24093839 ↗MalfunctionJan 16, 2026Burst Container or Vessel
AS REPORTED, THE BALLOON ON A 1.25MM X 15MM SABERX .014 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED AT APPROXIMATELY 7 ATMOSPHERES (ATM) DURING DILATION OF THE DORSALIS PEDIS ARTERY. ANOTHER 1.25MM X 15MM SABERX .014 PTA BALLOON CATHETER WAS USED; HOWEVER, THE BALLOON RUPTURED AGAIN AT APPROXIMATELY 13 ATMOSPHERES (ATM). A NON-CORDIS 1.5MM BALLOON CATHETER WAS THEN USED AS A REPLACEMENT AND WAS ABLE TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THIS WAS DURING A PROCEDURE TO TREAT STENOSIS OF THE LEFT SUPERFICIAL FEMORAL ARTERY (SFA), THE ANTERIOR TIBIAL ARTERY, AND THE DORSALIS PEDIS ARTERY. THE TARGET VESSEL WAS CHARACTERIZED BY MODERATE TO SEVERE CALCIFICATION, MODERATE TORTUOSITY, AND 100% STENOSIS. AN UNKNOWN 6F GUIDE SHEATH WAS INSERTED FROM THE LEFT GROIN, AND AN UNKNOWN .014 GUIDEWIRE WAS USED TO CROSS THE LESION. THE DEVICES WERE STORED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU), AND THERE WAS NO DIFFICULTY REMOVING ANY STERILE PACKAGING COMPONENTS OR PROTECTIVE BALLOON COVERS. THE DEVICES MAINTAINED NEGATIVE PRESSURE DURING PREPARATION. BOTH SABERX .014 PTA BALLOON CATHETERS WERE ABLE TO CROSS THE LESION PRIOR TO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24093789 ↗MalfunctionJan 12, 2026Burst Container or Vessel
AS REPORTED, A 3.5MM X 300MM 150CM SABER X .014 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) DILATATION CATHETER RUPTURED DURING THE FIRST INFLATION AT 6 ATMOSPHERES IN A SAME-SIDE BELOW-THE-KNEE CASE. A NON-CORDIS PTA BALLOON WAS USED AS A REPLACEMENT, AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO REPORTED PATIENT INJURY. ADDITIONAL EVENT DETAILS WERE REQUESTED; HOWEVER, COULD NOT BE OBTAINED. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION AS ANTICIPATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24032704 ↗MalfunctionJan 9, 2026Material Rupture
AS REPORTED, A 4 MM X 15 CM SABERX .014 150 CM PTA BALLOON CATHETER RUPTURED DURING INFLATION AFTER EXCEEDING NOMINAL PRESSURE. AFTER THE BALLOON RUPTURE, DILATION WAS PERFORMED WITH A NEW UNKNOWN 4 MM X 4 CM BALLOON, FOLLOWED BY AN UNKNOWN DRUG-COATED BALLOON, AND THE PROCEDURE WAS COMPLETED. THERE WAS NO REPORTED PATIENT INJURY. THE PROCEDURE INVOLVED TREATMENT OF LEFT SUPERFICIAL FEMORAL ARTERY (SFA) TO BELOW-KNEE ARTERY (BKA) STENOSIS VIA RIGHT GROIN ACCESS, INSERTION OF AN UNKNOWN 6F GUIDING SHEATH, LESION CROSSING WITH AN UNKNOWN 0.014 GUIDEWIRE, AND PRE-DILATION USING AN UNKNOWN 1.25 MM X 1.5 CM BALLOON FOLLOWED BY AN UNKNOWN 3 MM X 4 CM BALLOON. IT WAS UNCLEAR WHETHER THE CAUSE OF THE BALLOON RUPTURE WAS LESION CALCIFICATION OR PRODUCT QUALITY. THE PRODUCT WAS STORED PROPERLY PER THE INSTRUCTIONS FOR USE (IFU) AND THERE WERE NO DIFFICULTIES REMOVING IT FROM THE HOOP, THE PROTECTIVE BALLOON COVER, THE STYLET, OR ANY STERILE PACKAGING COMPONENTS. NO KINKS OR DAMAGES WERE NOTED PRIOR TO INSERTION. THE DEVICE WAS PREPPED ACCORDING TO THE IFU AND MAINTAINED NEGATIVE PRESSURE NORMALLY. THE INTENDED PROCEDURE WAS PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) USING A CONTRALATERAL AP
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24026126 ↗MalfunctionJan 8, 2026Burst Container or Vessel
AS REPORTED, THE BALLOON OF A SABERX .014 PTA, 2.5 X 200, 150CM, RUPTURED BEFORE REACHING NOMINAL PRESSURE. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE WAS USED IN A BELOW-KNEE (BK) CASE. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24008125 ↗MalfunctionDec 30, 2025Burst Container or Vessel; Burst Container or Vessel
AS REPORTED, THE BALLOON ON A 2.5MM X 300MM SABERX .014 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED BEFORE NOMINAL EXPANSION. A NON-CORDIS DEVICE WAS USED AS A REPLACEMENT TO PERFORM PRE-DILATION. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THIS WAS DURING A PROCEDURE TO TREAT A POPLITEAL ¿ BELOW KNEE (BK) VESSEL WITH CALCIFIED LESIONS. ADDITIONAL INFORMATION WAS REQUESTED BUT WAS NOT PROVIDED. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23922330 ↗MalfunctionDec 23, 2025Burst Container or Vessel; Burst Container or Vessel
AS REPORTED, DURING INFLATION OF A 1.25MM X 15MM SABERX.014 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER, CALCIFICATION WAS ENCOUNTERED IN THE SUPERFICIAL FEMORAL ARTERY (SFA) LESION AND THE BALLOON RUPTURED. THE DEICE WAS REMOVED AND A 1.5MM X 40M NON-CORDIS BALLOON CATHETER WAS USED TO CONTINUE THE PROCEDURE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THE DEVICE WAS STORED AND PREPPED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU), WITH NO DIFFICULTY ENCOUNTERED IN REMOVING STERILE PACKAGING COMPONENTS OR THE PROTECTIVE BALLOON COVER. NEGATIVE PRESSURE WAS MAINTAINED DURING PREPARATION. THE TARGET VESSEL EXHIBITED NO TORTUOSITY AND 100% STENOSIS. THE DEVICE WAS NOT PLACED IN AN ACUTE BEND DURING THE PROCEDURE AND WAS UNABLE TO CROSS THE LESION. THE DEVICE DID NOT KINK. IT WAS REMOVED EASILY AND INTACT IN ONE PIECE FROM THE PATIENT. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23881345 ↗MalfunctionDec 11, 2025Burst Container or Vessel; Burst Container or Vessel
DEVICE WAS DISCARDED AT SITE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23785606 ↗MalfunctionNov 4, 2025Burst Container or Vessel; Burst Container or Vessel
AS REPORTED, THE BALLOON ON A 6MM X 200MM SABERX 014 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED AT 4 ATMOSPHERES (ATM). A NON-CORDIS DEVICE WAS USED, AND EXPANSION WAS POSSIBLE AT NOMINAL PRESSURE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THIS WAS DURING A PROCEDURE TO TREAT A SUPERFICIAL FEMORAL ARTERY (SFA). ADDITIONAL INFORMATION WAS REQUESTED BUT WAS NOT PROVIDED. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23464241 ↗MalfunctionOct 22, 2025Burst Container or Vessel; Burst Container or Vessel
AS REPORTED, AFTER PASSAGE OF AN UNKNOWN 0.014 WIRE, A 2.5 X 300 .014 SABERX PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON WAS USED. DURING THE FIRST DILATION AT 8 ATM, THE BALLOON RUPTURED. A REPLACEMENT SABERX014 OF THE SAME SIZE WAS THEN USED, AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO REPORTED INJURY TO THE PATIENT. THIS WAS DURING A BELOW KNEE (BK) STENOSIS CASE USING AN IPSILATERAL APPROACH. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION. ADDITIONAL INFORMATION COULD NOT BE OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23360926 ↗MalfunctionOct 22, 2025Burst Container or Vessel; Burst Container or Vessel
AS REPORTED, THE BALLOON ON A 3MM X 200MM SABERX .014 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED AT 8 ATMOSPHERES (ATM) DURING DILATION. A 3 X 180 NON-CORDIS BALLOON CATHETER WAS USED AS A REPLACEMENT. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THIS WAS DURING A PROCEDURE TO TREAT A BELOW KNEE (BK) LESION WITH SEVERE CALCIFICATION, NO TORTUOSITY, AND 100% STENOSIS. THE DEVICE WAS STORED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DIFFICULTY REMOVING ANY OF THE STERILE PACKAGING COMPONENTS OR THE PROTECTIVE BALLOON COVER. THE DEVICE MAINTAINED NEGATIVE PRESSURE DURING PREPARATION. THE DEVICE WAS NOT PUT INTO AN ACUTE BEND AT ANY POINT DURING THE PROCEDURE AND WAS ABLE TO CROSS THE LESION WITHOUT KINKING. THE DEVICE WAS REMOVED EASILY AND INTACT FROM THE PATIENT. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23356853 ↗
Other Cordis US Corp. products
- MYNXGRIP2 models · 2,276 reports · 2 recalls
- Cordis ExoSeal Vascular Closure Device3 models · 1,279 reports
- MYNX CONTROL2 models · 534 reports
- POWERFLEX56 models · 306 reports
- SABER51 models · 270 reports
- Cordis S.M.A.R.T. Control Vascular Stent System30 models · 99 reports · 2 recalls
- INFINITI20 models · 107 reports
- AVIATOR19 models · 75 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page