Skip to content

SABERX .014 PTA Dilatation Catheter 47512015MRecalls & reported problems

Cordis US Corp. · PTA, RX, 1.25 x 15, 150cm · Class 2

All models of SABERX .014 PTA Dilatation Catheter (16)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 20705032089811), received between May 15, 2025 and Mar 18, 2026 — classified as 1 injury, 7 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel8
  • Difficult to Remove1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 18, 2026Burst Container or Vessel

    AS REPORTED, THE BALLOON ON A 1.25MM X 15MM SABERX .014 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED ON ITS FIRST INFLATION TO 8 ATMOSPHERES (ATM). THE DEVICE WAS REMOVED FROM THE PATIENT, AND A NON-CORDIS DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THIS WAS DURING A PROCEDURE TO TREAT A HIGHLY CALCIFIED ANTERIOR TIBIAL ARTERY (ATA) OCCLUSION VIA AN IPSILATERAL ANTEGRADE APPROACH. THE TARGET VESSEL WAS NOT TORTUOUS AND EXHIBITED 100% STENOSIS. THE SABERX .014 BALLOON CATHETER WAS BEING USED FOR PRE-DILATION. THE DEVICE WAS STORED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU), AND THERE WAS NO DIFFICULTY REMOVING THE STERILE PACKAGING COMPONENTS OR THE PROTECTIVE BALLOON COVER. THE DEVICE MAINTAINED NEGATIVE PRESSURE DURING PREPARATION, AND A 2:1 CONTRAST TO SALINE RATIO WAS USED. THE DEVICE WAS NOT PUT INTO AN ACUTE BEND, WAS ABLE TO CROSS THE LESION, DID NOT KINK DURING THE PROCEDURE, AND WAS REMOVED EASILY FROM THE PATIENT. THE PRODUCT WAS REMOVED INTACT IN ONE PIECE. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24627774 ↗
  • MalfunctionJan 16, 2026Burst Container or Vessel

    AS REPORTED, THE BALLOON ON A 1.25MM X 15MM SABERX .014 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED AT APPROXIMATELY 7 ATMOSPHERES (ATM) DURING DILATION OF THE DORSALIS PEDIS ARTERY. ANOTHER 1.25MM X 15MM SABERX .014 PTA BALLOON CATHETER WAS USED; HOWEVER, THE BALLOON RUPTURED AGAIN AT APPROXIMATELY 13 ATMOSPHERES (ATM). A NON-CORDIS 1.5MM BALLOON CATHETER WAS THEN USED AS A REPLACEMENT AND WAS ABLE TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THIS WAS DURING A PROCEDURE TO TREAT STENOSIS OF THE LEFT SUPERFICIAL FEMORAL ARTERY (SFA), THE ANTERIOR TIBIAL ARTERY, AND THE DORSALIS PEDIS ARTERY. THE TARGET VESSEL WAS CHARACTERIZED BY MODERATE TO SEVERE CALCIFICATION, MODERATE TORTUOSITY, AND 100% STENOSIS. AN UNKNOWN 6F GUIDE SHEATH WAS INSERTED FROM THE LEFT GROIN, AND AN UNKNOWN .014 GUIDEWIRE WAS USED TO CROSS THE LESION. THE DEVICES WERE STORED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU), AND THERE WAS NO DIFFICULTY REMOVING ANY STERILE PACKAGING COMPONENTS OR PROTECTIVE BALLOON COVERS. THE DEVICES MAINTAINED NEGATIVE PRESSURE DURING PREPARATION. BOTH SABERX .014 PTA BALLOON CATHETERS WERE ABLE TO CROSS THE LESION PRIOR TO

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24093839 ↗
  • MalfunctionJan 16, 2026Burst Container or Vessel

    AS REPORTED, THE BALLOON ON A 1.25MM X 15MM SABERX .014 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED AT APPROXIMATELY 7 ATMOSPHERES (ATM) DURING DILATION OF THE DORSALIS PEDIS ARTERY. ANOTHER 1.25MM X 15MM SABERX .014 PTA BALLOON CATHETER WAS USED; HOWEVER, THE BALLOON RUPTURED AGAIN AT APPROXIMATELY 13 ATMOSPHERES (ATM). A NON-CORDIS 1.5MM BALLOON CATHETER WAS THEN USED AS A REPLACEMENT AND WAS ABLE TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THIS WAS DURING A PROCEDURE TO TREAT STENOSIS OF THE LEFT SUPERFICIAL FEMORAL ARTERY (SFA), THE ANTERIOR TIBIAL ARTERY, AND THE DORSALIS PEDIS ARTERY. THE TARGET VESSEL WAS CHARACTERIZED BY MODERATE TO SEVERE CALCIFICATION, MODERATE TORTUOSITY, AND 100% STENOSIS. AN UNKNOWN 6F GUIDE SHEATH WAS INSERTED FROM THE LEFT GROIN, AND AN UNKNOWN .014 GUIDEWIRE WAS USED TO CROSS THE LESION. THE DEVICES WERE STORED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU), AND THERE WAS NO DIFFICULTY REMOVING ANY STERILE PACKAGING COMPONENTS OR PROTECTIVE BALLOON COVERS. THE DEVICES MAINTAINED NEGATIVE PRESSURE DURING PREPARATION. BOTH SABERX .014 PTA BALLOON CATHETERS WERE ABLE TO CROSS THE LESION PRIOR TO

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24093789 ↗
  • MalfunctionDec 23, 2025Burst Container or Vessel; Burst Container or Vessel

    AS REPORTED, DURING INFLATION OF A 1.25MM X 15MM SABERX.014 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER, CALCIFICATION WAS ENCOUNTERED IN THE SUPERFICIAL FEMORAL ARTERY (SFA) LESION AND THE BALLOON RUPTURED. THE DEICE WAS REMOVED AND A 1.5MM X 40M NON-CORDIS BALLOON CATHETER WAS USED TO CONTINUE THE PROCEDURE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THE DEVICE WAS STORED AND PREPPED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU), WITH NO DIFFICULTY ENCOUNTERED IN REMOVING STERILE PACKAGING COMPONENTS OR THE PROTECTIVE BALLOON COVER. NEGATIVE PRESSURE WAS MAINTAINED DURING PREPARATION. THE TARGET VESSEL EXHIBITED NO TORTUOSITY AND 100% STENOSIS. THE DEVICE WAS NOT PLACED IN AN ACUTE BEND DURING THE PROCEDURE AND WAS UNABLE TO CROSS THE LESION. THE DEVICE DID NOT KINK. IT WAS REMOVED EASILY AND INTACT IN ONE PIECE FROM THE PATIENT. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23881345 ↗
  • MalfunctionDec 11, 2025Burst Container or Vessel; Burst Container or Vessel

    DEVICE WAS DISCARDED AT SITE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23785606 ↗
  • InjuryAug 22, 2025Burst Container or Vessel; Difficult to Remove; Burst Container or Vessel; Difficult to Remove

    AS REPORTED, THE BALLOON OF A 1.25MM X 15MM SABERX .014 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED 6 ATMOSPHERES (ATM) DURING ITS FIRST INFLATION. WHEN ATTEMPTING TO REMOVE THE DEVICE FROM THE VESSEL, THE DEVICE WAS STUCK AND THE BALLOON PORTION STRETCHED MAKING REMOVAL IMPOSSIBLE. THE LUER HUB OF THE SABERX PTA BALLOON CATHETER WAS REMOVED AND A 5F NON-CORDIS CATHETER WAS PLACED OVER THE SABERX PTA BALLOON CATHETER, AND BOTH DEVICES WERE SUCCESSFULLY REMOVED TOGETHER. AFTER REMOVAL OF THE SABERX .014 BALLOON CATHETER, A NON-CORDIS BALLOON CATHETER WAS USED TO PERFORM PLAIN OLD BALLOON ANGIOPLASTY (POBA), AND THE PROCEDURE WAS COMPLETED. THE SABERX BALLOON CATHETER WAS BEING USED FOR PRE-DILATION OF AN ANTERIOR TIBIAL ARTERY (ATA) OCCLUSION WITH SEVERE CALCIFICATION, NO TORTUOSITY AND 100% STENOSIS. THE LESION WAS ACCESS VIA AN IPSILATERAL ANTEGRADE PUNCTURE. THE DEVICE WAS STORED AND PREPPED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DIFFICULTY REMOVING ANY OF THE STERILE PACKAGING COMPONENTS. THE BALLOON MAINTAINED NEGATIVE PRESSURE DURING PREPARATION. THE CONTRAST TO SALINE RATIO WAS 50%. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22871904 ↗
  • MalfunctionAug 4, 2025Burst Container or Vessel; Burst Container or Vessel

    AS REPORTED, A 1.25MM X 15MM SABER .014 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER MATERIAL RUPTURED DURING INFLATION AT THE LESION. THE BALLOON WAS INFLATED TO ABOUT 8 ATM WHEN IT RUPTURED. A 2.5X200 .014 SABERX BALLOON CATHETER WAS USED AS A REPLACEMENT. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THE DEVICE WAS STORED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DIFFICULTY REMOVING ANY OF THE STERILE PACKAGING COMPONENTS. THE DEVICE MAINTAINED NEGATIVE PRESSURE DURING PREPARATION. THE CONTRAST TO SALINE RATIO WAS 1:1. THE TARGET VESSEL WAS THE ANTERIOR TIBIAL ARTERY (ATA). THE TARGET SITE VESSEL WAS SEVERELY CALCIFIED, NOT TORTUOUS, AND HAD 100% STENOSIS. THE DEVICE WILL BE RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22703785 ↗
  • MalfunctionMay 15, 2025Burst Container or Vessel; Burst Container or Vessel

    AS REPORTED, THE BALLOON OF A 150CM SABERX014 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED AND WAS REMOVED. A NON-CORDIS DEVICE WAS USED AS REPLACEMENT TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE WAS INFLATED TO 6 ATMOSPHERES (ATM) BEFORE THE ANOMALY WAS NOTED, AND THE GAUGE OF THE INDEFLATOR INDICATED A LOSS OF PRESSURE ABOUT 3 SECONDS AFTER PRESSURE WAS MAINTAINED. THE LESION WAS THE SUPERFICIAL FEMORAL ARTERY AND FEMORAL POPLITEAL WHICH HAD SEVERE CALCIFICATION AND OCCLUSION. A CONTRALATERAL APPROACH WAS MADE. AFTER AN UNKNOWN GUIDEWIRE CROSSED, AN INTRAVASCULAR ULTRASOUND (IVUS) COULD NOT CROSS AND THE COMPLAINT DEVICE WAS USED; IT CROSSED AND INFLATED; HOWEVER, RUPTURED. THE INTENDED PROCEDURE WAS AN ENDOVASCULAR THROMBECTOMY (EVT). THE PRODUCT WAS STORED AND HANDLED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). NO DEVICE DAMAGE WAS NOTICED PRIOR TO OPENING THE PACKAGE. THERE WAS NO DIFFICULTY REMOVING THE DEVICE FROM THE STERILE PACKAGING OR THE PROTECTIVE BALLOON COVER, AND NO DIFFICULTY REMOVING ANY OF THE STERILE PACKAGING COMPONENTS. A CONTRALATERAL APPROACH WAS USED. THE TARGET SITE VESSEL CHARACTERISTICS INCLUDED S

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22031904 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page