Skip to content

SABERX .014 PTA Dilatation Catheter 47530300MRecalls & reported problems

Cordis US Corp. · PTA, RX, 3.0 x 300, 150cm · Class 2

All models of SABERX .014 PTA Dilatation Catheter (16)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20705032088753), received between Apr 28, 2025 and Mar 18, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel1
  • Decrease in Pressure1
  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 18, 2026Burst Container or Vessel

    AS REPORTED, THE 3.0 X 300 SABERX PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON PASSED THROUGH THE NON-CORDIS FIXED CATHETER (FC). HOWEVER, THE BALLOON RUPTURED BEFORE REACHING NOMINAL PRESSURE/SIZE. AFTERWARDS, A 3.0 X 200 SABERX BALLOON WAS USED AND THE PROCEDURE WAS COMPLETED WITHOUT PATIENT INJURY. ADDITIONAL INFORMATION WAS REQUESTED BUT NOT PROVIDED. THE DEVICE WILL BE RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24632396 ↗
  • MalfunctionApr 28, 2025Decrease in Pressure; Material Rupture; Decrease in Pressure; Material Rupture

    AS REPORTED, A 3.0 X 300, 150CM, .014 SABER PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED AT 4 ATMOSPHERES. THE DEVICE WAS REPLACED WITH A NEW DEVICE, WHICH INFLATED AT ITS NOMINAL PRESSURE. THE LESION WAS THE SUPERFICIAL FEMORAL ARTERY, WHERE A CONTRALATERAL APPROACH WAS MADE AND A 6F NON-CORDIS SHEATH WAS USED. A NON- CORDIS GUIDEWIRE CROSSED AND AFTER THAT THE SABER WAS USED. THE DEVICE IS NOT AVAILABLE FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21923403 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page