VOCSN Patient Breathing PackageRecalls & reported problems across all models
Ventec Life Systems, Inc. · 5 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names at least one VOCSN Patient Breathing Package model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 49
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 1
We found 5 models of VOCSN Patient Breathing Package by Ventec Life Systems, Inc. in the FDA's device registry (GUDID). Together they are named in 49 adverse event reports in the FDA's MAUDE database, received between Dec 7, 2023 and Oct 15, 2025 — classified as 49 malfunctions.
The most frequently reported specific device problem is “Material Integrity Problem” (28 reports).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
VOCSN Patient Breathing Package recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Jan 9, 2024Class IOpen, Classified
The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the breathing circuit which may cause serious injury or death. The risk of injury or death is increased for ventilator dependent patients.
Z-1070-2024 · Root cause: Process control · Recall details · FDA record ↗
Most reported VOCSN Patient Breathing Package problems
Device problems
- Material Integrity Problem28
- Material Separation14
- Connection Problem7
- Output Problem6
Patient problems
- No Clinical Signs, Symptoms or Conditions49
VOCSN Patient Breathing Package models
5 models, most reported first. Each model has its own page with its reports, recalls and device details.
- VOCSN Patient Breathing Package Pediatric, Active, Oxygen, BlueDevice, Positive Pressure Breathing, Intermittent1 recall25 reports
- VOCSN Patient Breathing Package Adult, Active, Heated, Oxygen, BlueHeater, Breathing System W/Wo Controller (Not Humidifier Or Nebulizer19 reports
- VOCSN Patient Breathing Package Adult, Passive, Oxygen, BlueDevice, Positive Pressure Breathing, Intermittent3 reports
- VOCSN Patient Breathing Package Pediatric, Active, Heated, Oxygen, BlueHeater, Breathing System W/Wo Controller (Not Humidifier Or Nebulizer1 report
- VOCSN Patient Breathing Package Adult, Active, Heated, BlueContinuous, Ventilator, Home Use1 report
Recent VOCSN Patient Breathing Package adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionOct 15, 2025Material Separation; Output Problem; Material Separation; Output Problem
IT WAS REPORTED TO VENTEC THAT A PATIENT CIRCUIT (PEDIATRIC, ACTIVE, HEATED, BLUE) BROKE AT THE EXHALATION VALVE. THE REPORTER ALSO STATED THAT THE DEVICE HAD DISPLAYED A PATIENT CIRCUIT DISCONNECT ALARM, WHICH MAY HAVE BEEN DUE TO THE BROKEN CIRCUIT. THERE WAS PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT, HOWEVER, THERE WERE NO REPORTS OF PATIENT HARM AS A RESULT OF THE REPORTED ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23287560 ↗MalfunctionAug 15, 2025Material Separation; Material Separation
IT WAS REPORTED TO VENTEC THAT A PATIENT CIRCUIT (ADULT, ACTIVE, HEATED, BLUE) HAD BROKEN DURING USE. THERE WAS PATIENT USE ASSOCIATED WITH THE REPORTED EVENT. THE PATIENT WAS PLACED ON THEIR BACK UP VENTILATOR. THE REPORTER ADVISED THAT THERE WAS NO PATIENT HARM AS A RESULT OF THE REPORTED ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22821755 ↗MalfunctionSep 12, 2024Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED TO VENTEC THAT THE CLEAR FILM WALL OF A BREATHING CIRCUIT HAD DETACHED FROM THE BLUE POLYPROPYLENE SPIRAL PORTION OF THE PATIENT CIRCUIT (ADULT, PASSIVE, OXYGEN, BLUE) ASSEMBLY. THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20208342 ↗MalfunctionSep 12, 2024Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED TO VENTEC THAT THE CLEAR FILM WALL OF A BREATHING CIRCUIT HAD DETACHED FROM THE BLUE POLYPROPYLENE SPIRAL PORTION OF THE PATIENT CIRCUIT (ADULT, PASSIVE, OXYGEN, BLUE) ASSEMBLY. THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20208038 ↗MalfunctionAug 7, 2024Material Separation; Output Problem; Material Separation; Output Problem
IT WAS REPORTED TO VENTEC THAT THE O2 DELIVERY TUBE PORTION OF THE PATIENT CIRCUIT (ADULT, ACTIVE, HEATED, OXYGEN, BLUE) HAD SEPARATED FROM THE MAIN TUBING ASSEMBLY. THERE WAS PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT. STAFF AT THE FACILITY ADVISED VENTEC THAT THEY INITIALLY OBSERVED THAT THE VENTILATOR'S EXHALED TIDAL VOLUME (VTE) LEVELS HAD DROPPED BY 40-60%, AS WELL AS CHANGES IN OTHER VENTILATOR PARAMETERS (HIGH LEAK, LOW TIDAL VOLUMES, ETC.). THEY FURTHER ADVISED THAT O2 DELIVERY WAS ALSO IMPACTED FOR PATIENTS WHO WERE USING THE INTERNAL CONCENTRATOR. STAFF THEN OBSERVED THAT THE PATIENT CIRCUIT WAS BROKEN AND CHANGED IT OUT TO RESOLVE THE ISSUE. THERE WERE NO REPORTS OF PATIENT HARM AS A RESULT OF THE REPORTED ISSUE. NO FURTHER DETAILS ABOUT THE PATIENT OR THE EVENT WERE PROVIDED. VENTEC HAS MADE MULTIPLE ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION ABOUT THE PATIENT, BUT NO RESPONSE HAS BEEN RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19934640 ↗MalfunctionAug 7, 2024Material Separation; Output Problem; Material Separation; Output Problem
IT WAS REPORTED TO VENTEC THAT THE O2 DELIVERY TUBE PORTION OF THE PATIENT CIRCUIT (ADULT, ACTIVE, HEATED, OXYGEN, BLUE) HAD SEPARATED FROM THE MAIN TUBING ASSEMBLY. THERE WAS PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT. STAFF AT THE FACILITY ADVISED VENTEC THAT THEY INITIALLY OBSERVED THAT THE VENTILATOR'S EXHALED TIDAL VOLUME (VTE) LEVELS HAD DROPPED BY 40-60%, AS WELL AS CHANGES IN OTHER VENTILATOR PARAMETERS (HIGH LEAK, LOW TIDAL VOLUMES, ETC.). THEY FURTHER ADVISED THAT O2 DELIVERY WAS ALSO IMPACTED FOR PATIENTS WHO WERE USING THE INTERNAL CONCENTRATOR. STAFF THEN OBSERVED THAT THE PATIENT CIRCUIT WAS BROKEN AND CHANGED IT OUT TO RESOLVE THE ISSUE. THERE WERE NO REPORTS OF PATIENT HARM AS A RESULT OF THE REPORTED ISSUE. NO FURTHER DETAILS ABOUT THE PATIENT OR THE EVENT WERE PROVIDED. VENTEC HAS MADE MULTIPLE ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION ABOUT THE PATIENT, BUT NO RESPONSE HAS BEEN RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19934578 ↗MalfunctionAug 7, 2024Material Separation; Output Problem; Material Separation; Output Problem
IT WAS REPORTED TO VENTEC THAT THE O2 DELIVERY TUBE PORTION OF THE PATIENT CIRCUIT (ADULT, ACTIVE, HEATED, OXYGEN, BLUE) HAD SEPARATED FROM THE MAIN TUBING ASSEMBLY. THERE WAS PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT. STAFF AT THE FACILITY ADVISED VENTEC THAT THEY INITIALLY OBSERVED THAT THE VENTILATOR'S EXHALED TIDAL VOLUME (VTE) LEVELS HAD DROPPED BY 40-60%, AS WELL AS CHANGES IN OTHER VENTILATOR PARAMETERS (HIGH LEAK, LOW TIDAL VOLUMES, ETC.). THEY FURTHER ADVISED THAT O2 DELIVERY WAS ALSO IMPACTED FOR PATIENTS WHO WERE USING THE INTERNAL CONCENTRATOR. STAFF THEN OBSERVED THAT THE PATIENT CIRCUIT WAS BROKEN AND CHANGED IT OUT TO RESOLVE THE ISSUE. THERE WERE NO REPORTS OF PATIENT HARM AS A RESULT OF THE REPORTED ISSUE. NO FURTHER DETAILS ABOUT THE PATIENT OR THE EVENT WERE PROVIDED. VENTEC HAS MADE MULTIPLE ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION ABOUT THE PATIENT, BUT NO RESPONSE HAS BEEN RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19934478 ↗MalfunctionAug 7, 2024Material Separation; Output Problem; Material Separation; Output Problem
IT WAS REPORTED TO VENTEC THAT THE O2 DELIVERY TUBE PORTION OF THE PATIENT CIRCUIT (ADULT, ACTIVE, HEATED, OXYGEN, BLUE) HAD SEPARATED FROM THE MAIN TUBING ASSEMBLY. THERE WAS PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT. STAFF AT THE FACILITY ADVISED VENTEC THAT THEY INITIALLY OBSERVED THAT THE VENTILATOR'S EXHALED TIDAL VOLUME (VTE) LEVELS HAD DROPPED BY 40-60%, AS WELL AS CHANGES IN OTHER VENTILATOR PARAMETERS (HIGH LEAK, LOW TIDAL VOLUMES, ETC.). THEY FURTHER ADVISED THAT O2 DELIVERY WAS ALSO IMPACTED FOR PATIENTS WHO WERE USING THE INTERNAL CONCENTRATOR. STAFF THEN OBSERVED THAT THE PATIENT CIRCUIT WAS BROKEN AND CHANGED IT OUT TO RESOLVE THE ISSUE. THERE WERE NO REPORTS OF PATIENT HARM AS A RESULT OF THE REPORTED ISSUE. NO FURTHER DETAILS ABOUT THE PATIENT OR THE EVENT WERE PROVIDED. VENTEC HAS MADE MULTIPLE ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION ABOUT THE PATIENT, BUT NO RESPONSE HAS BEEN RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19934212 ↗MalfunctionAug 7, 2024Material Separation; Output Problem; Material Separation; Output Problem
IT WAS REPORTED TO VENTEC THAT THE O2 DELIVERY TUBE PORTION OF THE PATIENT CIRCUIT (ADULT, ACTIVE, HEATED, OXYGEN, BLUE) HAD SEPARATED FROM THE MAIN TUBING ASSEMBLY. THERE WAS PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT. STAFF AT THE FACILITY ADVISED VENTEC THAT THEY INITIALLY OBSERVED THAT THE VENTILATOR'S EXHALED TIDAL VOLUME (VTE) LEVELS HAD DROPPED BY 40-60%, AS WELL AS CHANGES IN OTHER VENTILATOR PARAMETERS (HIGH LEAK, LOW TIDAL VOLUMES, ETC.). THEY FURTHER ADVISED THAT O2 DELIVERY WAS ALSO IMPACTED FOR PATIENTS WHO WERE USING THE INTERNAL CONCENTRATOR. STAFF THEN OBSERVED THAT THE PATIENT CIRCUIT WAS BROKEN AND CHANGED IT OUT TO RESOLVE THE ISSUE. THERE WERE NO REPORTS OF PATIENT HARM AS A RESULT OF THE REPORTED ISSUE. NO FURTHER DETAILS ABOUT THE PATIENT OR THE EVENT WERE PROVIDED. VENTEC HAS MADE MULTIPLE ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION ABOUT THE PATIENT, BUT NO RESPONSE HAS BEEN RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19934134 ↗MalfunctionAug 7, 2024Material Separation; Material Separation
IT WAS REPORTED TO VENTEC THAT THE O2 DELIVERY TUBE PORTION OF THE PATIENT CIRCUIT (ADULT, ACTIVE, HEATED, OXYGEN, BLUE) HAD SEPARATED FROM THE MAIN TUBING ASSEMBLY. THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19931528 ↗MalfunctionAug 7, 2024Material Separation; Material Separation
IT WAS REPORTED TO VENTEC THAT THE O2 DELIVERY TUBE PORTION OF THE PATIENT CIRCUIT (ADULT, ACTIVE, HEATED, OXYGEN, BLUE) HAD SEPARATED FROM THE MAIN TUBING ASSEMBLY. THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19931342 ↗MalfunctionAug 7, 2024Material Separation; Material Separation
IT WAS REPORTED TO VENTEC THAT THE O2 DELIVERY TUBE PORTION OF THE PATIENT CIRCUIT (ADULT, ACTIVE, HEATED, OXYGEN, BLUE) HAD SEPARATED FROM THE MAIN TUBING ASSEMBLY. THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19931314 ↗MalfunctionAug 7, 2024Material Separation; Material Separation
IT WAS REPORTED TO VENTEC THAT THE O2 DELIVERY TUBE PORTION OF THE PATIENT CIRCUIT (ADULT, ACTIVE, HEATED, OXYGEN, BLUE) HAD SEPARATED FROM THE MAIN TUBING ASSEMBLY. THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19931109 ↗MalfunctionAug 7, 2024Material Separation; Material Separation
IT WAS REPORTED TO VENTEC THAT THE O2 DELIVERY TUBE PORTION OF THE PATIENT CIRCUIT (ADULT, ACTIVE, HEATED, OXYGEN, BLUE) HAD SEPARATED FROM THE MAIN TUBING ASSEMBLY. THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19930893 ↗MalfunctionAug 7, 2024Material Separation; Material Separation
IT WAS REPORTED TO VENTEC THAT THE O2 DELIVERY TUBE PORTION OF THE PATIENT CIRCUIT (ADULT, ACTIVE, HEATED, OXYGEN, BLUE) HAD SEPARATED FROM THE MAIN TUBING ASSEMBLY. THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19930626 ↗
Other Ventec Life Systems, Inc. products
All Ventec Life Systems, Inc. devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.