VOCSN Patient Breathing Package Pediatric, Active, Oxygen, BlueRecalls & reported problems
Ventec Life Systems, Inc. · VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) · Class 2
All models of VOCSN Patient Breathing Package (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This is the Ventec VOCSN pediatric breathing circuit package, used with positive pressure ventilators. There are 25 reports, all malfunctions, mostly describing the spiral wrap or coil on the breathing circuit detaching or falling apart, sometimes straight out of the box, with no deaths and no injuries. One recall is open, started January 2024, because a manufacturing issue can let the bonded spiral wrap detach and cause blockages or leaks in the circuit.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 25
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 00855573007587), received between Dec 7, 2023 and Mar 12, 2024 — classified as 25 malfunctions.
The most frequently reported specific device problem is “Material Integrity Problem” (25 reports).
1 FDA recall record name this device's identifier, 1 still open.
4 other recalls cover devices cleared under the same FDA submission (K162877); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jan 9, 2024Class IOpen, Classified
The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the breathing circuit which may cause serious injury or death. The risk of injury or death is increased for ventilator dependent patients.
Z-1070-2024 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Integrity Problem25
Patient problems
- No Clinical Signs, Symptoms or Conditions25
Recent adverse event reports
The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 12, 2024Material Integrity Problem; Material Integrity Problem
VENTEC CONTACTED THE CUSTOMER TO ENSURE THE RECEIPT AND ACKNOWLEDGEMENT OF CUSTOMER RECALL NOTIFICATION Z-1070-2024 FOR THE VOCSN PATIENT BREATHING PACKAGE (PEDIATRIC, ACTIVE, OXYGEN, BLUE). THE CUSTOMER VERBALLY CONFIRMED RECEIPT OF THE CUSTOMER RECALL NOTIFICATION. DURING THAT CONVERSATION, THE CUSTOMER NOTIFIED VENTEC THAT THE PATIENT CIRCUIT HAD BEEN FOUND TO BE DEFECTIVE RIGHT OUT OF THE BOX, NOTING THAT THE BREATHING CIRCUIT PRODUCT COIL STRUCTURE HAD BECOME DELAMINATED FROM THE CIRCUIT TUBING. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18890681 ↗MalfunctionMar 12, 2024Material Integrity Problem; Material Integrity Problem
VENTEC CONTACTED THE CUSTOMER TO ENSURE THE RECEIPT AND ACKNOWLEDGEMENT OF CUSTOMER RECALL NOTIFICATION Z-1070-2024 FOR THE VOCSN PATIENT BREATHING PACKAGE (PEDIATRIC, ACTIVE, OXYGEN, BLUE). THE CUSTOMER VERBALLY CONFIRMED RECEIPT OF THE CUSTOMER RECALL NOTIFICATION. DURING THAT CONVERSATION, THE CUSTOMER NOTIFIED VENTEC THAT THE PATIENT CIRCUIT HAD BEEN FOUND TO BE DEFECTIVE RIGHT OUT OF THE BOX, NOTING THAT THE BREATHING CIRCUIT PRODUCT COIL STRUCTURE HAD BECOME DELAMINATED FROM THE CIRCUIT TUBING. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18890666 ↗MalfunctionMar 12, 2024Material Integrity Problem; Material Integrity Problem
VENTEC CONTACTED THE CUSTOMER TO ENSURE THE RECEIPT AND ACKNOWLEDGEMENT OF CUSTOMER RECALL NOTIFICATION Z-1070-2024 FOR THE VOCSN PATIENT BREATHING PACKAGE (PEDIATRIC, ACTIVE, OXYGEN, BLUE). THE CUSTOMER VERBALLY CONFIRMED RECEIPT OF THE CUSTOMER RECALL NOTIFICATION. DURING THAT CONVERSATION, THE CUSTOMER NOTIFIED VENTEC THAT THE PATIENT CIRCUIT HAD BEEN FOUND TO BE DEFECTIVE RIGHT OUT OF THE BOX, NOTING THAT THE BREATHING CIRCUIT PRODUCT COIL STRUCTURE HAD BECOME DELAMINATED FROM THE CIRCUIT TUBING. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18890641 ↗MalfunctionMar 12, 2024Material Integrity Problem; Material Integrity Problem
VENTEC CONTACTED THE CUSTOMER TO ENSURE THE RECEIPT AND ACKNOWLEDGEMENT OF CUSTOMER RECALL NOTIFICATION Z-1070-2024 FOR THE VOCSN PATIENT BREATHING PACKAGE (PEDIATRIC, ACTIVE, OXYGEN, BLUE). THE CUSTOMER VERBALLY CONFIRMED RECEIPT OF THE CUSTOMER RECALL NOTIFICATION. DURING THAT CONVERSATION, THE CUSTOMER NOTIFIED VENTEC THAT THE PATIENT CIRCUIT HAD BEEN FOUND TO BE DEFECTIVE RIGHT OUT OF THE BOX, NOTING THAT THE BREATHING CIRCUIT PRODUCT COIL STRUCTURE HAD BECOME DELAMINATED FROM THE CIRCUIT TUBING. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18890611 ↗MalfunctionMar 12, 2024Material Integrity Problem; Material Integrity Problem
VENTEC CONTACTED THE CUSTOMER TO ENSURE THE RECEIPT AND ACKNOWLEDGEMENT OF CUSTOMER RECALL NOTIFICATION Z-1070-2024 FOR THE VOCSN PATIENT BREATHING PACKAGE (PEDIATRIC, ACTIVE, OXYGEN, BLUE). THE CUSTOMER VERBALLY CONFIRMED RECEIPT OF THE CUSTOMER RECALL NOTIFICATION. DURING THAT CONVERSATION, THE CUSTOMER NOTIFIED VENTEC THAT THE PATIENT CIRCUIT HAD BEEN FOUND TO BE DEFECTIVE RIGHT OUT OF THE BOX, NOTING THAT THE BREATHING CIRCUIT PRODUCT COIL STRUCTURE HAD BECOME DELAMINATED FROM THE CIRCUIT TUBING. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18890534 ↗MalfunctionMar 12, 2024Material Integrity Problem; Material Integrity Problem
VENTEC CONTACTED THE CUSTOMER TO ENSURE THE RECEIPT AND ACKNOWLEDGEMENT OF CUSTOMER RECALL NOTIFICATION Z-1070-2024 FOR THE VOCSN PATIENT BREATHING PACKAGE (PEDIATRIC, ACTIVE, OXYGEN, BLUE). THE CUSTOMER VERBALLY CONFIRMED RECEIPT OF THE CUSTOMER RECALL NOTIFICATION. DURING THAT CONVERSATION, THE CUSTOMER NOTIFIED VENTEC THAT THE PATIENT CIRCUIT HAD BEEN FOUND TO BE DEFECTIVE RIGHT OUT OF THE BOX, NOTING THAT THE BREATHING CIRCUIT PRODUCT COIL STRUCTURE HAD BECOME DELAMINATED FROM THE CIRCUIT TUBING. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18890310 ↗MalfunctionMar 12, 2024Material Integrity Problem; Material Integrity Problem
VENTEC CONTACTED THE CUSTOMER TO ENSURE THE RECEIPT AND ACKNOWLEDGEMENT OF CUSTOMER RECALL NOTIFICATION Z-1070-2024 FOR THE VOCSN PATIENT BREATHING PACKAGE (PEDIATRIC, ACTIVE, OXYGEN, BLUE). THE CUSTOMER VERBALLY CONFIRMED RECEIPT OF THE CUSTOMER RECALL NOTIFICATION. DURING THAT CONVERSATION, THE CUSTOMER NOTIFIED VENTEC THAT THE PATIENT CIRCUIT HAD BEEN FOUND TO BE DEFECTIVE RIGHT OUT OF THE BOX, NOTING THAT THE BREATHING CIRCUIT PRODUCT COIL STRUCTURE HAD BECOME DELAMINATED FROM THE CIRCUIT TUBING. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18889278 ↗MalfunctionMar 12, 2024Material Integrity Problem; Material Integrity Problem
VENTEC CONTACTED THE CUSTOMER TO ENSURE THE RECEIPT AND ACKNOWLEDGEMENT OF CUSTOMER RECALL NOTIFICATION Z-1070-2024 FOR THE VOCSN PATIENT BREATHING PACKAGE (PEDIATRIC, ACTIVE, OXYGEN, BLUE). THE CUSTOMER VERBALLY CONFIRMED RECEIPT OF THE CUSTOMER RECALL NOTIFICATION. DURING THAT CONVERSATION, THE CUSTOMER NOTIFIED VENTEC THAT THE PATIENT CIRCUIT HAD BEEN FOUND TO BE DEFECTIVE RIGHT OUT OF THE BOX, NOTING THAT THE BREATHING CIRCUIT PRODUCT COIL STRUCTURE HAD BECOME DELAMINATED FROM THE CIRCUIT TUBING. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18889245 ↗MalfunctionMar 12, 2024Material Integrity Problem; Material Integrity Problem
VENTEC CONTACTED THE CUSTOMER TO ENSURE THE RECEIPT AND ACKNOWLEDGEMENT OF CUSTOMER RECALL NOTIFICATION Z-1070-2024 FOR THE VOCSN PATIENT BREATHING PACKAGE (PEDIATRIC, ACTIVE, OXYGEN, BLUE). THE CUSTOMER VERBALLY CONFIRMED RECEIPT OF THE CUSTOMER RECALL NOTIFICATION. DURING THAT CONVERSATION, THE CUSTOMER NOTIFIED VENTEC THAT THE PATIENT CIRCUIT HAD BEEN FOUND TO BE DEFECTIVE RIGHT OUT OF THE BOX, NOTING THAT THE BREATHING CIRCUIT PRODUCT COIL STRUCTURE HAD BECOME DELAMINATED FROM THE CIRCUIT TUBING. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18889073 ↗MalfunctionMar 12, 2024Material Integrity Problem; Material Integrity Problem
VENTEC CONTACTED THE CUSTOMER TO ENSURE THE RECEIPT AND ACKNOWLEDGEMENT OF CUSTOMER RECALL NOTIFICATION Z-1070-2024 FOR THE VOCSN PATIENT BREATHING PACKAGE (PEDIATRIC, ACTIVE, OXYGEN, BLUE). THE CUSTOMER VERBALLY CONFIRMED RECEIPT OF THE CUSTOMER RECALL NOTIFICATION. DURING THAT CONVERSATION, THE CUSTOMER NOTIFIED VENTEC THAT THE PATIENT CIRCUIT HAD BEEN FOUND TO BE DEFECTIVE RIGHT OUT OF THE BOX, NOTING THAT THE BREATHING CIRCUIT PRODUCT COIL STRUCTURE HAD BECOME DELAMINATED FROM THE CIRCUIT TUBING. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18889038 ↗MalfunctionDec 7, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED TO VENTEC THAT THE CLEAR FILM WALL OF A BREATHING CIRCUIT HAD DETACHED FROM THE BLUE POLYPROPYLENE SPIRAL PORTION OF THE PATIENT CIRCUIT (PEDIATRIC, ACTIVE, OXYGEN, BLUE) ASSEMBLY. THE REPORTED ISSUE WAS DISCOVERED DURING INCOMING INSPECTION. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT. SECTION D4 MODEL NUMBER, CATALOG NUMBER AND LOT CODE REFLECTS THE INFORMATION KNOWN ABOUT THE PATIENT CIRCUIT. THE UDI WAS UNAVAILABLE AT THE TIME OF THIS SUBMISSION, THEREFORE SECTION D4 UDI IS "UNKNOWN", AND SECTION H4 DATE OF MANUFACTURER WILL BE LEFT BLANK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18288876 ↗MalfunctionDec 7, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED TO VENTEC THAT THE CLEAR FILM WALL OF A BREATHING CIRCUIT HAD DETACHED FROM THE BLUE POLYPROPYLENE SPIRAL PORTION OF THE PATIENT CIRCUIT (PEDIATRIC, ACTIVE, OXYGEN, BLUE) ASSEMBLY. THE REPORTED ISSUE WAS DISCOVERED DURING INCOMING INSPECTION. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT. SECTION D4 MODEL NUMBER, CATALOG NUMBER AND LOT CODE REFLECTS THE INFORMATION KNOWN ABOUT THE PATIENT CIRCUIT. THE UDI WAS UNAVAILABLE AT THE TIME OF THIS SUBMISSION, THEREFORE SECTION D4 UDI IS "UNKNOWN", AND SECTION H4 DATE OF MANUFACTURER WILL BE LEFT BLANK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18288859 ↗MalfunctionDec 7, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED TO VENTEC THAT THE CLEAR FILM WALL OF A BREATHING CIRCUIT HAD DETACHED FROM THE BLUE POLYPROPYLENE SPIRAL PORTION OF THE PATIENT CIRCUIT (PEDIATRIC, ACTIVE, OXYGEN, BLUE) ASSEMBLY. THE REPORTED ISSUE WAS DISCOVERED DURING INCOMING INSPECTION. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT. SECTION D4 MODEL NUMBER, CATALOG NUMBER AND LOT CODE REFLECTS THE INFORMATION KNOWN ABOUT THE PATIENT CIRCUIT. THE UDI WAS UNAVAILABLE AT THE TIME OF THIS SUBMISSION, THEREFORE SECTION D4 UDI IS "UNKNOWN", AND SECTION H4 DATE OF MANUFACTURER WILL BE LEFT BLANK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18288799 ↗MalfunctionDec 7, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED TO VENTEC THAT THE CLEAR FILM WALL OF A BREATHING CIRCUIT HAD DETACHED FROM THE BLUE POLYPROPYLENE SPIRAL PORTION OF THE PATIENT CIRCUIT (PEDIATRIC, ACTIVE, OXYGEN, BLUE) ASSEMBLY. THE REPORTED ISSUE WAS DISCOVERED DURING INCOMING INSPECTION. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT. SECTION D4 MODEL NUMBER, CATALOG NUMBER AND LOT CODE REFLECTS THE INFORMATION KNOWN ABOUT THE PATIENT CIRCUIT. THE UDI WAS UNAVAILABLE AT THE TIME OF THIS SUBMISSION, THEREFORE SECTION D4 UDI IS "UNKNOWN", AND SECTION H4 DATE OF MANUFACTURER WILL BE LEFT BLANK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18288791 ↗MalfunctionDec 7, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED TO VENTEC THAT THE CLEAR FILM WALL OF A BREATHING CIRCUIT HAD DETACHED FROM THE BLUE POLYPROPYLENE SPIRAL PORTION OF THE PATIENT CIRCUIT (PEDIATRIC, ACTIVE, OXYGEN, BLUE) ASSEMBLY. THE REPORTED ISSUE WAS DISCOVERED DURING INCOMING INSPECTION. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT. SECTION D4 MODEL NUMBER, CATALOG NUMBER AND LOT CODE REFLECTS THE INFORMATION KNOWN ABOUT THE PATIENT CIRCUIT. THE UDI WAS UNAVAILABLE AT THE TIME OF THIS SUBMISSION, THEREFORE SECTION D4 UDI IS "UNKNOWN", AND SECTION H4 DATE OF MANUFACTURER WILL BE LEFT BLANK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18288775 ↗MalfunctionDec 7, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED TO VENTEC THAT THE CLEAR FILM WALL OF A BREATHING CIRCUIT HAD DETACHED FROM THE BLUE POLYPROPYLENE SPIRAL PORTION OF THE PATIENT CIRCUIT (PEDIATRIC, ACTIVE, OXYGEN, BLUE) ASSEMBLY. THE REPORTED ISSUE WAS DISCOVERED DURING INCOMING INSPECTION. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT. SECTION D4 MODEL NUMBER, CATALOG NUMBER AND LOT CODE REFLECTS THE INFORMATION KNOWN ABOUT THE PATIENT CIRCUIT. THE UDI WAS UNAVAILABLE AT THE TIME OF THIS SUBMISSION, THEREFORE SECTION D4 UDI IS "UNKNOWN", AND SECTION H4 DATE OF MANUFACTURER WILL BE LEFT BLANK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18288744 ↗MalfunctionDec 7, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED TO VENTEC THAT THE CLEAR FILM WALL OF A BREATHING CIRCUIT HAD DETACHED FROM THE BLUE POLYPROPYLENE SPIRAL PORTION OF THE PATIENT CIRCUIT (PEDIATRIC, ACTIVE, OXYGEN, BLUE) ASSEMBLY. THE REPORTED ISSUE WAS DISCOVERED DURING INCOMING INSPECTION. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT. SECTION D4 MODEL NUMBER, CATALOG NUMBER AND LOT CODE REFLECTS THE INFORMATION KNOWN ABOUT THE PATIENT CIRCUIT. THE UDI WAS UNAVAILABLE AT THE TIME OF THIS SUBMISSION, THEREFORE SECTION D4 UDI IS "UNKNOWN", AND SECTION H4 DATE OF MANUFACTURER WILL BE LEFT BLANK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18288739 ↗MalfunctionDec 7, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED TO VENTEC THAT THE CLEAR FILM WALL OF A BREATHING CIRCUIT HAD DETACHED FROM THE BLUE POLYPROPYLENE SPIRAL PORTION OF THE PATIENT CIRCUIT (PEDIATRIC, ACTIVE, OXYGEN, BLUE) ASSEMBLY. THE REPORTED ISSUE WAS DISCOVERED DURING INCOMING INSPECTION. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT. SECTION D4 MODEL NUMBER, CATALOG NUMBER AND LOT CODE REFLECTS THE INFORMATION KNOWN ABOUT THE PATIENT CIRCUIT. THE UDI WAS UNAVAILABLE AT THE TIME OF THIS SUBMISSION, THEREFORE SECTION D4 UDI IS "UNKNOWN", AND SECTION H4 DATE OF MANUFACTURER WILL BE LEFT BLANK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18288738 ↗MalfunctionDec 7, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED TO VENTEC THAT THE CLEAR FILM WALL OF A BREATHING CIRCUIT HAD DETACHED FROM THE BLUE POLYPROPYLENE SPIRAL PORTION OF THE PATIENT CIRCUIT (PEDIATRIC, ACTIVE, OXYGEN, BLUE) ASSEMBLY. THE REPORTED ISSUE WAS DISCOVERED DURING INCOMING INSPECTION. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT. SECTION D4 MODEL NUMBER, CATALOG NUMBER AND LOT CODE REFLECTS THE INFORMATION KNOWN ABOUT THE PATIENT CIRCUIT. THE UDI WAS UNAVAILABLE AT THE TIME OF THIS SUBMISSION, THEREFORE SECTION D4 UDI IS "UNKNOWN", AND SECTION H4 DATE OF MANUFACTURER WILL BE LEFT BLANK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18288736 ↗MalfunctionDec 7, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED TO VENTEC THAT THE CLEAR FILM WALL OF A BREATHING CIRCUIT HAD DETACHED FROM THE BLUE POLYPROPYLENE SPIRAL PORTION OF THE PATIENT CIRCUIT (PEDIATRIC, ACTIVE, OXYGEN, BLUE) ASSEMBLY. THE REPORTED ISSUE WAS DISCOVERED DURING INCOMING INSPECTION. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT. SECTION D4 MODEL NUMBER, CATALOG NUMBER AND LOT CODE REFLECTS THE INFORMATION KNOWN ABOUT THE PATIENT CIRCUIT. THE UDI WAS UNAVAILABLE AT THE TIME OF THIS SUBMISSION, THEREFORE SECTION D4 UDI IS "UNKNOWN", AND SECTION H4 DATE OF MANUFACTURER WILL BE LEFT BLANK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18288707 ↗MalfunctionDec 7, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED TO VENTEC THAT THE CLEAR FILM WALL OF A BREATHING CIRCUIT HAD DETACHED FROM THE BLUE POLYPROPYLENE SPIRAL PORTION OF THE PATIENT CIRCUIT (PEDIATRIC, ACTIVE, OXYGEN, BLUE) ASSEMBLY. THE REPORTED ISSUE WAS DISCOVERED DURING INCOMING INSPECTION. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT. SECTION D4 MODEL NUMBER, CATALOG NUMBER AND LOT CODE REFLECTS THE INFORMATION KNOWN ABOUT THE PATIENT CIRCUIT. THE UDI WAS UNAVAILABLE AT THE TIME OF THIS SUBMISSION, THEREFORE SECTION D4 UDI IS "UNKNOWN", AND SECTION H4 DATE OF MANUFACTURER WILL BE LEFT BLANK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18288648 ↗MalfunctionDec 7, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED TO VENTEC THAT THE CLEAR FILM WALL OF A BREATHING CIRCUIT HAD DETACHED FROM THE BLUE POLYPROPYLENE SPIRAL PORTION OF THE PATIENT CIRCUIT (PEDIATRIC, ACTIVE, OXYGEN, BLUE) ASSEMBLY. THE REPORTED ISSUE WAS DISCOVERED DURING INCOMING INSPECTION. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT. SECTION D4 MODEL NUMBER, CATALOG NUMBER AND LOT CODE REFLECTS THE INFORMATION KNOWN ABOUT THE PATIENT CIRCUIT. THE UDI WAS UNAVAILABLE AT THE TIME OF THIS SUBMISSION, THEREFORE SECTION D4 UDI IS "UNKNOWN", AND SECTION H4 DATE OF MANUFACTURER WILL BE LEFT BLANK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18288621 ↗MalfunctionDec 7, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED TO VENTEC THAT THE CLEAR FILM WALL OF A BREATHING CIRCUIT HAD DETACHED FROM THE BLUE POLYPROPYLENE SPIRAL PORTION OF THE PATIENT CIRCUIT (PEDIATRIC, ACTIVE, OXYGEN, BLUE) ASSEMBLY. THE REPORTED ISSUE WAS DISCOVERED DURING INCOMING INSPECTION. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT. SECTION D4 MODEL NUMBER, CATALOG NUMBER AND LOT CODE REFLECTS THE INFORMATION KNOWN ABOUT THE PATIENT CIRCUIT. THE UDI WAS UNAVAILABLE AT THE TIME OF THIS SUBMISSION, THEREFORE SECTION D4 UDI IS "UNKNOWN", AND SECTION H4 DATE OF MANUFACTURER WILL BE LEFT BLANK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18288615 ↗MalfunctionDec 7, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED TO VENTEC THAT THE CLEAR FILM WALL OF A BREATHING CIRCUIT HAD DETACHED FROM THE BLUE POLYPROPYLENE SPIRAL PORTION OF THE PATIENT CIRCUIT (PEDIATRIC, ACTIVE, OXYGEN, BLUE) ASSEMBLY. THE REPORTED ISSUE WAS DISCOVERED DURING INCOMING INSPECTION. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT. SECTION D4 MODEL NUMBER, CATALOG NUMBER AND LOT CODE REFLECTS THE INFORMATION KNOWN ABOUT THE PATIENT CIRCUIT. THE UDI WAS UNAVAILABLE AT THE TIME OF THIS SUBMISSION, THEREFORE SECTION D4 UDI IS "UNKNOWN", AND SECTION H4 DATE OF MANUFACTURER WILL BE LEFT BLANK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18288567 ↗MalfunctionDec 7, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED TO VENTEC THAT THE CLEAR FILM WALL OF A BREATHING CIRCUIT HAD DETACHED FROM THE BLUE POLYPROPYLENE SPIRAL PORTION OF THE PATIENT CIRCUIT (PEDIATRIC, ACTIVE, OXYGEN, BLUE) ASSEMBLY. THE REPORTED ISSUE WAS DISCOVERED DURING INCOMING INSPECTION. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT. SECTION D4 MODEL NUMBER, CATALOG NUMBER AND LOT CODE REFLECTS THE INFORMATION KNOWN ABOUT THE PATIENT CIRCUIT. THE UDI WAS UNAVAILABLE AT THE TIME OF THIS SUBMISSION, THEREFORE SECTION D4 UDI IS "UNKNOWN", AND SECTION H4 DATE OF MANUFACTURER WILL BE LEFT BLANK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18288548 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K162877. One submission often covers many models, so check the product and lot details in each record.
- Z-2525-2026 · May 8, 2026 · Open, ClassifiedVOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001
- Z-1942-2026 · Mar 25, 2026 · Open, ClassifiedVOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000
- Z-1196-2025 · Feb 3, 2025 · Open, ClassifiedVOCSN Multi-Function Ventilators: VOCSN+Pro (V+O+C+S+N+Pro, English), REF: PRT-00490-001; V+Pro (V+Pro, English), REF: PRT-01185-000, PRT-01185-002
- Z-1902-2021 · May 10, 2021 · Open, ClassifiedVentec Life Systems VOCSN Multi-Function Ventilator Model/Version/Catalog Number: V+O+C+S+N+Pro/English/PRT-00490-001; V+O+C+S+N+Pro/Japan/PRT-00490-100; V+C+Pro/English/PRT-00900
Other Device, Positive Pressure Breathing, Intermittent devices
- Percussionaire P5Percussionaire Corporation · Device, Positive Pressure Breathing, Intermittent4 recalls2 reports
- Volara System M08473Hill-Rom Services Pte. Ltd. · DEVICE, POSITIVE PRESSURE BREATHING, INTERMITTENT3 recalls0 reports
- Percussionaire F00001-CPercussionaire Corporation · Device, Positive Pressure Breathing, Intermittent2 recalls3 reports
- Percussionaire F00001-C-FRPercussionaire Corporation · Device, Positive Pressure Breathing, Intermittent2 recalls0 reports
- Percussionaire F00001-C-FRPercussionaire Corporation · Device, Positive Pressure Breathing, Intermittent2 recalls0 reports
- Volara System M08085Hill-Rom Services Pte. Ltd. · DEVICE, POSITIVE PRESSURE BREATHING, INTERMITTENT2 recalls0 reports
- Percussionaire F00001-CPercussionaire Corporation · Device, Positive Pressure Breathing, Intermittent2 recalls0 reports
- Percussionaire F00001-CPercussionaire Corporation · Device, Positive Pressure Breathing, Intermittent2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record