Ventec Life Systems, Inc. · 17 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

4 open recalls name at least one VOCSN model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
2,922
Deaths reported
7
Injuries reported
81
Recalls
4

We found 17 models of VOCSN by Ventec Life Systems, Inc. in the FDA's device registry (GUDID). Together they are named in 2,922 adverse event reports in the FDA's MAUDE database, received between Apr 16, 2018 and Sep 30, 2026 — classified as 7 deaths, 81 injuries, 2,834 malfunctions.

The most frequently reported specific device problem is “Output Problem” (2,121 reports).

4 FDA recall records name at least one of these models by its device identifier, 4 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 82 reportsDec 2023: 49 reportsJan 2024: 97 reportsFeb 2024: 94 reportsMar 2024: 80 reportsApr 2024: 63 reportsMay 2024: 64 reportsJun 2024: 47 reportsJul 2024: 48 reportsAug 2024: 46 reportsSep 2024: 65 reportsOct 2024: 115 reportsNov 2024: 128 reportsDec 2024: 36 reportsJan 2025: 103 reportsFeb 2025: 6 reportsMar 2025: 118 reportsApr 2025: 73 reportsMay 2025: 75 reportsJun 2025: 108 reportsJul 2025: 105 reportsAug 2025: 72 reportsSep 2025: 36 reportsOct 2025: 32 reportsNov 2025: 139 reportsDec 2025: 36 reportsJan 2026: 27 reportsFeb 2026: 50 reportsMar 2026: 85 reportsApr 2026: 68 reportsMay 2026: 57 reportsJun 2026: 21 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 8 reportsOct 2026: 0 reports202420252026139
Last 36 months. The most recent month may be incomplete.

VOCSN recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. May 8, 2026Open, Classified

    Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition.

    Z-2525-2026 · Root cause: Vendor change control · Recall details · FDA record ↗

  2. Mar 25, 2026Open, Classified

    Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to reduced below intended levels of fraction of inspired oxygen delivery prior to and during ventilation and this oxygen-enriched environment may increase fire risk

    Z-1942-2026 · Root cause: Process change control · Recall details · FDA record ↗

  3. Feb 3, 2025Open, Classified

    Multi-Function Ventilators were serviced using incorrect parts which have the potential to cause unexpected shutdown, or when using an active circuit could result in inaccurate tidal volume monitor, not triggering on patient efforts, less inspiratory volume, less inspiratory pressure, and less PEEP delivered.

    Z-1196-2025 · Root cause: Component change control · names 3 of these models · Recall details · FDA record ↗

  4. May 10, 2021Open, Classified

    Due to a component failure, ventilators may unexpectedly shut down or have a loss of blower function while in patient use. The shutdown potentially prevents the ventilator from providing ventilation therapy.

    Z-1902-2021 · Root cause: Device Design · names 6 of these models · Recall details · FDA record ↗

Most reported VOCSN problems

Device problems

  • Output Problem2,121
  • Unexpected Shutdown513
  • Connection Problem165
  • Display or Visual Feedback Problem119
  • Device Alarm System68
  • Inaccurate Delivery34
  • No Display/Image22
  • Use of Device Problem19

Patient problems

  • No Clinical Signs, Symptoms or Conditions2,810
  • No Known Impact Or Consequence To Patient29
  • Low Oxygen Saturation23
  • Insufficient Information17
  • Discomfort14
  • No Patient Involvement11
  • No Consequences Or Impact To Patient7
  • Distress4

VOCSN models

17 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent VOCSN adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.