VOCSN Patient Breathing Package recallZ-1070-2024 · Ventec Life Systems, Inc. · initiated Jan 9, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ventec Life Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the breathing circuit which may cause serious injury or death. The risk of injury or death is increased for ventilator dependent patients.
FDA root cause category: Process control
What the company is doing
On 12/22/2023, the firm sent an URGENT: MEDICAL DEVICE RECALL NOTIFICATION via Direct overnight mail, or FedEx, or email, or hand delivered, to customers informing them that Ventec Life Systems is recalling a specific lot (listed below) of the VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) since the bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Customer are instructed to: 1. Operate the ventilator as defined in the Instructions for Use and promptly respond to all alarms activated by the ventilator. 2. Stop use of the affected Patient Breathing Package and replace it with a non-affected product. 3. Inspect current stock and quarantine/segregate any unused affected products to prevent their use. For questions/assistance - contact Ventec Life Systems directly at 844-MYVOCSN (844-698-6276), Monday through Friday between 8:00 a.m. - 5:00 p.m. PDT.
Recall details
- Recall number
- Z-1070-2024
- Status
- Open, Classified
- Initiated by the company
- Jan 9, 2024
- Posted by the FDA
- Feb 23, 2024
- Recalling firm
- Ventec Life Systems, Inc.
- Quantity
- 250 breathing circuits
- Product code
- CBK
- FDA submissions
- K162877
- Distribution
- U.S.: CA, CO, FL, HI, IN, MO, NC, and WA O.U.S.: Japan
Product description
VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN Unified Respiratory System which provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.
Code information (models, lots, serial numbers)
Box Label - Part Numbers: PRT-01013-000 Pouch Label - Part Number: PRT-00802-001 UDI-DI Code: 00855573007594 Lot Number: 220922 Pack Slip Lot Number: 257337
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- VOCSN Patient Breathing Package Pediatric, Active, Oxygen, BlueVentec Life Systems, Inc. · Device, Positive Pressure Breathing, Intermittent1 recall25 reports
Product lines
Other devices under the same FDA submission
22 devices cleared under K162877 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- VOCSN V+Pro, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit1 recall570 reports
- VOCSN V+Pro Emergency, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit558 reports
- VOCSN V+O+C+S+N+Pro, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit3 recalls366 reports
- VOCSN V+Pro, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit2 recalls2 deaths325 reports
- VOCSN V+O+C+S+N, EnglishVentec Life Systems, Inc. · Ventilator, Continuous, Facility Use1 recall286 reports
- VOCSN V*Home, EnglishVentec Life Systems, Inc. · Heater, Breathing System W/Wo Controller (Not Humidifier Or Nebulizer4 deaths318 reports
- VOCSN V+C+S+N+Pro, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit185 reports
- VOCSN VOCSN, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit1 death96 reports
- VOCSN V+C+Pro, JapaneseVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit1 recall33 reports
- VOCSN V+O+C+S+N+Pro, JapaneseVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit1 recall32 reports
- VOCSN V+C, EnglishVentec Life Systems, Inc. · Ventilator, Continuous, Facility Use1 recall27 reports
- VOCSN V+C+Pro, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit1 recall25 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.